Class 2 Device Recall Custom Sterile and NonSterile Procedural Kits containing Turkuaz Ultrasound Gel

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Custom Sterile and NonSterile Procedural Kits containing Turkuaz Ultrasound Gel see related information
Date Initiated by Firm May 15, 2023
Date Posted August 03, 2023
Recall Status1 Open3, Classified
Recall Number Z-2335-2023
Recall Event ID 92399
Product Classification Hysterectomy kit - Product Code OJF
Product Custom Sterile and Non-Sterile Procedural Kits containing Turkuaz Ultrasound Gel. Labeled as the following:

a. CUSTOM GYN, Model Number: DYNJ63257C;

b. GYN EXP LAPAROTOMY, Model Number: DYNJ906578C;

c. INTRAUTERINE TRNSFUSION -LF, Model Number: DYNJ9048708I;

d. PERI NEURO BLOCK PACK-LF, Model Number: PHS762887001B;

e. PICC TRAY, Model Number: DYNJ32988;

f. SLING PACK-LF, Model Number: DYNJ55361, DYNJ55361A;

g. TRANS VAGINAL TRAY (KW01JF), Model Number: OBQ056E;

h. TVS4700 CUSTOM PACK, Model Number: TVS4700;
Code Information All lot numbers for the following models and corresponding UPCs distributed from Medline between 04/01/2020 and 04/28/2023: a. Model Number: DYNJ63257C, UPC Number: 019348993830. b. Model Number: DYNJ906578C, UPC Number: 019532702355. c. Model Number: DYNJ9048708I, UPC Number: 019532707273. d. Model Number: PHS762887001B, UPC Number: 088994284931. e. Model Number: DYNJ32988, UPC Number: 088438924278. f. Model Number: DYNJ55361, UPC Number: 088994254669; Model Number: DYNJ55361A, UPC Number: 019532717352. g. Model Number: OBQ056E, UPC Number: 088827758937. h. Model Number: TVS4700, UPC Number: 081698202339.
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information Contact Karin Johnson
886-359-1704
Manufacturer Reason
for Recall
Potential for the labeled sterile ultrasound gel component within packs to not meet sterility specifications.
FDA Determined
Cause 2
Nonconforming Material/Component
Action Medline notified customers beginning on 05/15/2023, via mail and email with letter. Customers with Medline Probe Cover Kits were instructed to check inventory for affected kits, destroy any affected kits, and complete and return the response form. Distributors were instructed to notify their customers and have them document and destroy affected kits. Customers with the non-sterile and sterile procedural kits were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits. Distributors were instructed to notify customers of the recall and have customers return any affected product for return to Medline. Medline identified additional units of Medline Probe Cover Kits and notified customers on 06/22/2023. Customers were instructed to quarantine kits, complete and return the response form and, once updated labels are received, over label and discard the affected ultrasound gel product within the kits.
Quantity in Commerce 6,340 kits
Distribution Worldwide distribution - US Nationwide and the countries of Canada, China, Costa Rica, India, Lebanon, Panama, Qatar, Saudi Arabia, Singapore, United Arab Emirates.
Total Product Life Cycle TPLC Device Report