| Class 2 Device Recall Transferrin reagent (Product ID: OSR6152) | |
Date Initiated by Firm | September 06, 2023 |
Date Posted | November 16, 2023 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-0309-2024 |
Recall Event ID |
93206 |
510(K)Number | K963139 |
Product Classification |
Transferrin, rhodamine, antigen, antiserum, control - Product Code DDD
|
Product | AU/DxC AU Chemistry Transferrin REF: OSR6152
Immune complexes formed in solution scatter light in proportion to their size, shape and concentration. Turbidimeters measure the reduction of incident light due to reflection, absorption, or scatter. In the procedure, the measurement of the decrease in light transmitted (increase in absorbance) through particles suspended in solution as a result of complexes formed during the antigen-antibody reaction, is the basis of this assay. |
Code Information |
UDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024 |
Recalling Firm/ Manufacturer |
Beckman Coulter Biomedical Division Lismeehan, O'Callaghan's Mills County Clare Ireland
|
For Additional Information Contact | (0)65 683 1100 |
Manufacturer Reason for Recall | The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | On 9/6/23, recall notices were mailed and emailed to customers who were asked to do the following:
1) Discontinue use of affected transferrin.
2) Contact firm for replacement lots.
3) No retrospective review of past patient results is required but this may be carried out at the discretion of the Laboratory Director.
4) Share this information with laboratory staff and retain the notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, provide them a copy of this letter.
5) Complete and return the response form via email to Email: regaffairsfax@beckman.com
Questions can be directed to the Customer Support Center via website: http://www.beckmancoulter.com or phone: 1-800-854-3633 in the United States. |
Quantity in Commerce | 2,754 |
Distribution | Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = DDD
|
|
|
|