| | Class 1 Device Recall HeartMate Touch Communication System |  |
| Date Initiated by Firm | January 03, 2024 |
| Date Posted | February 02, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0842-2024 |
| Recall Event ID |
93697 |
| PMA Number | P160054 |
| Product Classification |
Ventricular (assist) bypass - Product Code DSQ
|
| Product | 1. HeartMate Touch Communication System, REF HMT1100
2. HeartMate Touch Communication System, Rental REF HMT1100-R
3. HeartMate Touch Communication System, HMT1150
The HeartMate Touch Application is loaded onto a tablet and will provide a user interface with state-of-the-art hardware/software platforms for supporting LVAD patients. Along with being able to program and monitor the LVAS, the HeartMate Touch Application introduces features that enables users to setup pairing and connection with the existing HeartMate 3 Left Ventricular Assist System (LVAS) via the HeartMate Touch Wireless Adapter and generate reports from the log files that are downloaded from the HM3 System Controller. |
| Code Information |
HeartMate Touch Communication System Model Number: HMT1100
UDI-DI Code: 05415067032058
Serial Numbers:
DMPV40AAHP83
DMPVN1DFHP83
DMPVQ2JSHP51
DMPVQ374HP51
DMPVR6EXHP51
DMPVW1KDHP51
DMPW90USHP83
DMPWD118HP83
DMPWG096HP83
DMPWH036HP83
DMPWL02KHP83
DMPWL04QHP83
DMPWL05PHP83
DMPWL0AMHP83
DMPWL0T1HP83
DMPWL0XHHP83
DMPWL111HP83
DMPWL13CHP83
DMPWL14XHP83
DMPWL19RHP83
DMPWL1B7HP83
DMPWL1BPHP83
DMPWL1CCHP83
DMPWL1DVHP83
DMPWL21LHP83
DMPWL243HP83
DMPWL24PHP83
DMPWL29YHP83
DMPWL2BAHP83
DMPWL2DAHP83
DMPWM02VHP83
DMPWM0B5HP83
DMPWN01MHP83
DMPWN041HP83
DMPWN0A0HP83
DMPWN0AVHP83
DMPWN0XRJ28L
DMPWR0AKHP83
DMPWR0AZHP83
DMPWR0BSHP83
DMPWR0C5HP83
DMPWR0CCHP83
DMPWR0DWHP83
DMPWR10BHP83
DMPWR11AHP83
DMPWR124HP83
DMPWR160HP83
DMPWR16ZHP83
DMPWR17CHP83
DMPWR17SHP83
DMPWR19BHP83
DMPWR19JHP83
DMPWR1D2HP83
DMPWR1DTHP83
DMPWR1F6HP83
DMPWR1MZHP83
DMPWT00PHP83
DMPWT02RHP83
DMPWT0CBHP83
DMPWT0NQHP83
DMPWT0PJHP83
DMPWT0URHP83
DMPWT0YHHP83
DMPWT118HP83
DMPWV09DHP83
DMPWV0BHHP51
DMPWV0C5HP83
DMPWV0C6HP83
DMPWV0CCHP83
DMPWV0CRHP83
DMPWV0DRHP83
DMPWV0DVHP83
DMPWV0E8HP83
DMPWV0EDHP83
DMPWV0FFHP83
DMPWV0FJHP83
DMPWV0FTHP83
DMPWV0GJHP83
DMPWV0GXHP83
DMPWV0H4HP83
DMPWV0H9HP83
DMPWV0J6HP83
DMPWV0JEHP83
DMPWV0JNHP83
DMPWV0K2HP83
DMPWV0K9HP83
DMPWV0LDHP83
DMPWW004HP83
DMPWW00KHP83
DMPWW00LHP83
DMPWW012HP83
DMPWW019HP83
DMPWW01BHP83
DMPWW04JHP83
DMPWW06PHP83
DMPWW095HP83
DMPWW0DKHP83
DMPWW0E9HP83
DMPWW0F1HP83
DMPWW0FGHP83
DMPWW0G7HP83
DMPWW0GAHP83
DMPWW0GBHP83
DMPWW0GLHP83
DMPWW0GWHP83
DMPWW0H0HP83
DMPWW0H1HP83
DMPWW0H7HP83
DMPWW0HLHP83
DMPWW0J7HP83
DMPWW0JGHP83
DMPWW0JZHP83
DMPWW0KDHP83
DMPWW0KKHP83
DMPWW0KWHP83
DMPWW0LCHP83
DMPWW0MCHP83
DMPWW0MKHP83
DMPWW0PWHP83
DMPWW0Q9J28L
DMPWW0R5HP83
DMPWW0RGHP83
DMPWW0S5HP83
DMPX10UAHP51
DMPX10UDHP51
DMPX10ZMHP51
DMPX11Z2HP51
DMPX127EHP51
DMPX12GXHP51
DMPX12J0HP51
DMPX12JJHP51
DMPX12MKHP51
DMPX12RPHP51
DMPX12RRHP51
DMPX12S2HP51
DMPX12UPHP51
DMPX143YHP51
DMPX1AYLHP50
DMPX20FNHP83
DMPX20FPHP83
DMPX20G0HP83
DMPX20VNHP83
DMPX20VVHP83
DMPX20W2HP50
DMPX20ZRHP51
DMPX2141HP51
DMPX21A6HP83
DMPX21C8HP83
DMPX21LXHP83
DMPX21QAHP83
DMPX21TLHP51
DMPX228LHP50
DMPX22A2HP50
DMPX22B3HP50
DMPX22J4HP50
DMPX22WXHP51
DMPX2398HP51
DMPX3031HP83
DMPX304KHP83
DMPX305FHP83
DMPX30DEHP83
DMPX30E6HP83
DMPX30EDHP83
DMPX30EFHP83
DMPX30ELHP83
DMPX30FKHP83
DMPX30FPHP83
DMPX30FTHP83
DMPX30FWHP83
DMPX30G8HP83
DMPX30H8HP83
DMPX30H9HP83
DMPX30HBHP83
DMPX30HEHP83
DMPX30HGHP83
DMPX30J0HP83
DMPX30JHHP83
DMPX30JKHP83
DMPX30JNHP83
DMPX30JVHP83
DMPX30KCHP83
DMPX30KSHP83
DMPX30LXHP83
DMPX30M5HP83
DMPX30MHHP83
DMPX30MXHP83
DMPX30NCHP83
DMPX30NVHP83
DMPX30NXHP83
DMPX30P1HP83
DMPX30PFHP83
DMPX30PVHP83
DMPX30Q3HP83
DMPX30Q4HP83
DMPX30QQHP83
DMPX30RHHP83
DMPX30S3HP83
DMPX30SEHP83
DMPX30TMHP83
DMPX30TZHP83
DMPX30U7HP83
DMPX30USHP83
DMPX30UUHP83
DMPX30UYHP83
DMPX30VNHP83
DMPX30VRHP83
DMPX30WJHP83
DMPX30XXHP83
DMPX30YPHP83
DMPX30YSHP83
DMPX30Z3HP83
DMPX310WHP83
DMPX310YHP83
DMPX311MHP83
DMPX311SHP83
DMPX311VHP83
DMPX3126HP83
DMPX3127HP83
DMPX312GHP83
DMPX312QHP83
DMPX313CHP83
DMPX313MHP83
DMPX313NHP83
DMPX313UHP83
DMPX313YHP83
DMPX3148HP83
DMPX314LHP83
DMPX314NHP83
DMPX314SHP83
DMPX314WHP83
DMPX3152HP83
DMPX3158HP83
DMPX3159HP83
DMPX315BHP83
DMPX315KHP83
DMPX315QHP83
DMPX315XHP83
DMPX315YHP83
DMPX3169HP83
DMPX316HHP83
DMPX316MHP83
DMPX316XHP83
DMPX3172HP83
DMPX3177HP83
DMPX3179HP83
DMPX317MHP83
DMPX3180HP83
DMPX3186HP83
DMPX318CHP83
DMPX318FHP83
DMPX318QHP83
DMPX318UHP83
DMPX318YHP83
DMPX3195HP51
DMPX319BHP83
DMPX319CHP83
DMPX319VHP83
DMPX31AJHP83
DMPX31B4HP83
DMPX31CDHP83
DMPX31CWHP83
DMPX31D7HP51
DMPX31D8HP83
DMPX31DDHP83
DMPX31DVHP83
DMPX31EZHP83
DMPX31F7HP83
DMPX31F9HP83
DMPX31FDHP83
DMPX31FJHP83
DMPX31FKHP83
DMPX31GNHP83
DMPX31H4HP83
DMPX31X0HP51
DMPX31XZHP51
DMPX3238HP51
DMPX3459HP51
DMPX404VHP83
DMPX405UHP51
DMPX40AXHP83
DMPX41GLHP51
DMPX46T2HP50
DMPX470MHP50
DMPX50N3HP51
DMPX50P0HP50
DMPX50R3HP50
DMPX51E4HP51
DMPX51QLHP50
DMPX51RRHP50
DMPX51SVHP50
DMPX51V3HP50
DMPX51VKHP50
DMPX51ZAHP50
DMPX52NFHP51
DMPX52NPHP50
DMPX5310HP50
DMPX5357HP50
DMPX53J1HP50
DMPX53KZHP50
DMPX53WEHP50
DMPX53ZGHP50
DMPX544QHP50
DMPX54BQHP50
DMPX54DBHP50
DMPX54DWHP50
DMPX54EWHP50
DMPX54FQHP50
DMPX54FXHP50
DMPX54HGHP50
DMPX54KDHP50
DMPX54SRHP50
DMPX54VDHP50
DMPX54W9HP50
DMPX54WCHP50
DMPX5506HP50
DMPX5573HP50
DMPX557QHP50
DMPX5596HP50
DMPX559VHP50
DMPX559XHP50
DMPX55ASHP50
DMPX55D7HP50
DMPX55EGHP50
DMPX55FSHP50
DMPX55G7HP50
DMPX55L2HP50
DMPX55M2HP50
DMPX55MDHP50
DMPX55NWHP50
DMPX55P2HP50
DMPX55PTHP50
DMPX55Q2HP50
DMPX55QHHP50
DMPX55QZHP50
DMPX55VGHP50
DMPX55ZXHP50
DMPX563ZHP50
DMPX56FWHP50
DMPX56HLHP50
DMPX56J3HP50
DMPX56KRHP50
DMPX56LCHP50
DMPX56MEHP50
DMPX61WQJ28L
DMPX61Y1J28L
DMPX7115HP51
DMPX7174HP51
DMPX71DKHP51
DMPX71LAHP51
DMPX71LEHP51
DMPX71TXHP51
DMPX71V5HP51
DMPX71X0HP51
DMPX725HHP51
DMPX7282HP51
DMPX728LHP51
DMPX7299HP51
DMPX732UHP50
DMPX73GGHP50
DMPX74UNHP50
DMPX74VKHP50
DMPX74WCHP50
DMPX74WHHP50
DMPX74YCHP50
DMPX753LHP50
DMPX7595HP50
DMPX76DTHP50
DMPX76E6HP50
DMPX76ESHP50
DMPX76EWHP50
DMPX76FAHP50
DMPX76GGHP50
DMPX76GQHP50
DMPX76HLHP50
DMPX76HMHP50
DMPX76HWHP50
DMPX76JKHP50
DMPX76LKHP50
DMPX76NHHP50
DMPX76NZHP50
DMPX76QVHP50
DMPX76RCHP50
DMPX76RHHP50
DMPX76SYHP50
DMPX77DWHP50
DMPX77FSHP50
DMPX77LZHP50
DMPX77M2HP50
DMPX77NAHP50
DMPX77PUHP50
DMPX77SCHP50
DMPX77YNHP50
DMPX77Z9HP50
DMPX77ZWHP50
DMPX780AHP50
DMPX780EHP50
DMPX782GHP50
DMPX782WHP50
DMPX7839HP50
DMPX783BHP50
DMPX783VHP50
DMPX784NHP50
DMPX785JHP50
DMPX78A8HP50
DMPX78AEHP50
DMPX78CGHP50
DMPX78DUHP50
DMPX78G2HP50
DMPX78J5HP50
DMPX78J8HP50
DMPX78JQHP50
DMPX78NNHP50
DMPX78P5HP50
DMPX78P6HP50
DMPX78Q4HP50
DMPX78QHHP50
DMPX78QNHP50
DMPX78QQHP50
DMPX78RSHP50
DMPX78SHHP50
DMPX78STHP50
DMPX7935HP50
DMPX795PHP50
DMPX79K9HP50
DMPX7BFUHP50
DMPX80MYHP51
DMPX80Q4HP51
DMPX813ZHP51
DMPX822THP51
DMPX837VHP50
DMPX83FSHP50
DMPX903DHP83
DMPX9058HP83
DMPX905WHP83
DMPX965QHP50
DMPX96Z6HP50
DMPXC09PHP50
DMPXC0GZHP51
DMPXC0LEHP51
DMPXC18QHP51
DMPXC1Z0HP50
DMPXC359HP50
DMPXC4E7HP50
DMPXC4MXHP50
DMPXD0HEHP51
DMPXD13LHP51
DMPXD17GHP51
DMPXD1ASHP51
DMPXD1DVHP51
DMPXD1EDHP51
DMPXD1FAHP51
DMPXD1G9HP51
DMPXD1HKHP51
DMPXD1LMHP51
DMPXD1LWHP51
DMPXD1N9HP51
DMPXD1NLHP51
DMPXD1NMHP51
DMPXD1P1HP51
DMPXD1QRHP50
DMPXD1QSHP51
DMPXD1R2HP51
DMPXD1SHHP51
DMPXD1T9HP51
DMPXD21WJ28L
DMPXD586HP50
DMPXF019HP83
DMPXF01RHP83
DMPXF04LHP83
DMPXF05SHP83
DMPXF07UHP83
DMPXF0ABHP83
DMPXF0B2HP83
DMPXF0J0HP83
DMPXF0QWHP83
DMPXF353HP51
DMPXG076HP83
DMPXG0L1HP83
DMPXG0Q7HP83
DMPXG0QJHP83
DMPXG0QPHP83
DMPXG0R7HP83
DMPXG0RKHP83
DMPXG0Z1HP51
DMPXG1PTHP51
DMPXG1QNHP51
DMPXG2DZHP51
DMPXH02HHP51
DMPXH09YHP51
DMPXH0AEHP51
DMPXJ009HP83
DMPXJ07PHP83
DMPXJ0BRHP83
DMPXJ0BXHP83
DMPXJ0C0HP83
DMPXJ0C2HP83
DMPXJ0CMHP83
DMPXJ0CSHP83
DMPXJ0CVHP83
DMPXJ0ERHP83
DMPXJ0FXHP83
DMPXJ0G5HP83
DMPXJ0G8HP83
DMPXJ0HNHP83
DMPXJ0HRHP83
DMPXJ0J8HP83
DMPXJ0MHHP83
DMPXJ0P8HP83
DMPXK001HP83
DMPXK00THP83
DMPXK00WHP83
DMPXK010HP83
DMPXK01RHP83
DMPXK01SHP83
DMPXK01ZHP83
DMPXK02UHP83
DMPXK03EHP83
DMPXK03GHP83
DMPXK03VHP83
DMPXK03WHP83
DMPXK03ZHP83
DMPXK040HP83
DMPXK048HP83
DMPXK049HP83
DMPXK04QHP83
DMPXK04RHP83
DMPXK04XHP83
DMPXK04ZHP83
DMPXK050HP83
DMPXK051HP83
DMPXK053HP83
DMPXK05FHP83
DMPXK05GHP83
DMPXK05SHP83
DMPXK05VHP83
DMPXK05WHP83
DMPXK05YHP83
DMPXK061HP83
DMPXK062HP83
DMPXK065HP83
DMPXK068HP83
DMPXK06JHP83
DMPXK06KHP83
DMPXK079HP83
DMPXK07DHP83
DMPXK083HP83
DMPXK08HHP83
DMPXK08KHP83
DMPXK08NHP83
DMPXK092HP83
DMPXK094HP83
DMPXK099HP83
DMPXK09DHP83
DMPXK09KHP83
DMPXK0A2HP83
DMPXK0ADHP83
DMPXK0AJHP83
DMPXK0AVHP83
DMPXK0B2HP83
DMPXK0B3J28L
DMPXK0B8HP83
DMPXK0BEHP83
DMPXK0BRHP83
DMPXK0C4HP83
DMPXK0CBHP83
DMPXK0CEHP83
DMPXK0CFHP83
DMPXK0CJHP83
DMPXK0CSHP83
DMPXK0CUHP83
DMPXK0CZHP83
DMPXK0D0HP83
DMPXK0D6HP83
DMPXK0D9HP83
DMPXK0DBHP83
DMPXK0DJHP83
DMPXK0DSHP83
DMPXK0DVHP83
DMPXK0DXHP83
DMPXK0DYHP83
DMPXK0DZHP83
DMPXK0E0HP83
DMPXK0E2HP83
DMPXK0ECHP83
DMPXK0EDHP83
DMPXK0EEHP83
DMPXK0EFHP83
DMPXK0EMHP83
DMPXK0ENHP83
DMPXK0EPHP83
DMPXK0EWHP83
DMPXK0F7HP83
DMPXK0FGHP83
DMPXK0FKHP83
DMPXK0FTHP83
DMPXK0FWHP83
DMPXK0G6HP83
DMPXK0GHHP83
DMPXK0GTHP83
DMPXK0GWHP83
DMPXK0H5HP83
DMPXK0H6HP83
DMPXK0H7HP83
DMPXK0HCHP83
DMPXK0HDHP83
DMPXK0HEHP83
DMPXK0HQHP83
DMPXK0J3HP83
DMPXK0J4HP83
DMPXK0J7HP83
DMPXK0JAHP83
DMPXK0JCHP83
DMPXK0JGHP83
DMPXK0JMHP83
DMPXK0JPHP83
DMPXK0JQHP83
DMPXK0JRHP83
DMPXK0JVHP83
DMPXK0JZHP83
DMPXK0K0HP83
DMPXK0KSHP83
DMPXK0KTHP83
DMPXK0KUHP83
DMPXK0KYHP83
DMPXK0KZHP83
DMPXK0L1HP83
DMPXK0L8HP83
DMPXK0LBHP83
DMPXK0LHHP83
DMPXK0LKHP83
DMPXK0LPHP83
DMPXK0LTHP83
DMPXK0LUHP83
DMPXK0M5HP83
DMPXK0M7HP83
DMPXK0M9HP83
DMPXK0MCHP83
DMPXK0MHHP83
DMPXK0MPHP83
DMPXK0MRHP83
DMPXK0MSHP83
DMPXK0MXHP83
DMPXK0MZHP83
DMPXK0N7HP83
DMPXK0NKHP83
DMPXK0NPHP83
DMPXK0NVHP83
DMPXK0NYHP83
DMPXK0P1HP83
DMPXK0P2HP83
DMPXK0P3HP83
DMPXK0P7HP83
DMPXK0PAHP83
DMPXK0PDHP83
DMPXK0PLHP83
DMPXK0PRHP83
DMPXK0Q1HP83
DMPXK0Q3HP83
DMPXK0Q4HP83
DMPXK0Q6HP83
DMPXK0Q8HP83
DMPXK0Q9HP83
DMPXK0QAHP83
DMPXK0QCHP83
DMPXK0QEHP83
DMPXK0QHHP83
DMPXK0QJHP83
DMPXK0QLHP83
DMPXK0QPHP83
DMPXK0QRHP83
DMPXK0QTHP83
DMPXK0QWHP83
DMPXK0QXHP83
DMPXK0QZHP83
DMPXK0R3HP83
DMPXK0R7HP83
DMPXK0R8HP83
DMPXK0RAHP83
DMPXK0RBHP83
DMPXK0RCHP83
DMPXK0RDHP83
DMPXK0RGHP83
DMPXK0RJHP83
DMPXK0RMHP83
DMPXK0RNHP83
DMPXK0RRHP83
DMPXK0RSHP83
DMPXK0S0HP83
DMPXK0S2HP83
DMPXK0S4HP83
DMPXK0S5HP83
DMPXK0S6HP83
DMPXK0S8HP83
DMPXK0SAHP83
DMPXK0SMHP83
DMPXK0SRHP83
DMPXK0SVHP83
DMPXK0SZHP83
DMPXK0T0HP83
DMPXK0T1HP83
DMPXK0TAHP83
DMPXK0TBHP83
DMPXK0TCHP83
DMPXK0TFHP83
DMPXK0TRHP83
DMPXK0TVHP83
DMPXK0TXHP83
DMPXK0TZHP83
DMPXK0U0HP83
DMPXK0U2HP83
DMPXK0UMHP83
DMPXK0UNHP83
DMPXK0V6HP83
DMPXK0VCHP83
DMPXK0VDHP83
DMPXK0VFHP83
DMPXK0VJHP83
DMPXK0VKHP83
DMPXK0VPHP83
DMPXK0VQHP83
DMPXK0VRHP83
DMPXK0VVHP83
DMPXK0VYHP83
DMPXK0W2HP83
DMPXK0W6HP83
DMPXK0W7HP83
DMPXK0WDHP83
DMPXK0WGHP83
DMPXK0WJHP83
DMPXK0WLHP83
DMPXK0WNHP83
DMPXK0X5HP83
DMPXK0X6HP83
DMPXK0XAHP83
DMPXK0XCHP83
DMPXK0XHHP83
DMPXK0XNHP83
DMPXK0XTHP83
DMPXK0Y0HP83
DMPXK0Y1HP83
DMPXK0Y4HP83
DMPXK0Y5HP83
DMPXK0Y6HP83
DMPXK0YJHP83
DMPXK0Z2HP83
DMPXK0Z8HP83
DMPXK0ZCHP83
DMPXK0ZFHP83
DMPXK0ZJHP83
DMPXK0ZQHP83
DMPXK104HP83
DMPXK105HP83
DMPXK107HP83
DMPXK108HP83
DMPXK109HP83
DMPXK10BHP83
DMPXK10DHP83
DMPXK10KHP83
DMPXK10LHP83
DMPXK10MHP83
DMPXK10PHP83
DMPXK10QHP83
DMPXK10UHP83
DMPXK10ZHP83
DMPXK110HP83
DMPXK112HP83
DMPXK114HP83
DMPXK118HP83
DMPXK119HP83
DMPXK11DHP83
DMPXK11LHP83
DMPXK11NHP83
DMPXK11UHP83
DMPXK11YHP83
DMPXK120HP83
DMPXK121HP83
DMPXK125HP83
DMPXK12BHP83
DMPXK12HHP83
DMPXK12JHP83
DMPXK12MHP83
DMPXK12PHP83
DMPXK12QHP83
DMPXK130HP83
DMPXK133HP83
DMPXK134HP83
DMPXK135HP83
DMPXK136HP83
DMPXK139HP83
DMPXK13LHP83
DMPXK13XHP83
DMPXK140HP83
DMPXK145HP83
DMPXK14AHP83
DMPXK14CHP83
DMPXK14FHP83
DMPXK14GHP83
DMPXK14KHP83
DMPXK14RHP83
DMPXK14UHP83
DMPXK14YHP83
DMPXK154HP83
DMPXK158HP83
DMPXK159HP83
DMPXK15LHP83
DMPXK15UHP83
DMPXK16DHP83
DMPXK16EHP83
DMPXK16LHP83
DMPXK16YHP83
DMPXK170HP83
DMPXK172HP83
DMPXK173HP83
DMPXK17BHP83
DMPXK17CHP83
DMPXK17DHP83
DMPXK17FHP83
DMPXK17NHP83
DMPXK17RHP83
DMPXK17SHP83
DMPXK17THP83
DMPXK17UHP83
DMPXK17VHP83
DMPXK18CHP83
DMPXK18GHP83
DMPXK18QHP83
DMPXK18UHP83
DMPXK191HP83
DMPXK19EHP83
DMPXK19GHP83
DMPXK19NHP83
DMPXK19WHP83
DMPXK19YHP83
DMPXK1ACHP83
DMPXK1AHHP83
DMPXK1AMHP83
DMPXK1AWHP83
DMPXK1B1HP83
DMPXK1B2HP83
DMPXK1B3HP83
DMPXK1B6HP83
DMPXK1BAHP83
DMPXK1BFHP83
DMPXK1BRHP83
DMPXK1BVHP83
DMPXK1CYHP83
DMPXK1D4HP83
DMPXK1D8HP83
DMPXK1DDHP83
DMPXK1DFHP83
DMPXK1DVHP83
DMPXK1DWHP83
DMPXK1DXHP83
DMPXK1E8HP83
DMPXK1EGHP83
DMPXK1F1HP83
DMPXK1F3HP83
DMPXK1F9HP83
DMPXK1FGHP83
DMPXK1FKHP83
DMPXK1FRHP83
DMPXK1JXHP83
DMPXK1K8HP83
DMPXK1K9HP83
DMPXK1KBHP83
DMPXK1KLHP83
DMPXK1KNHP83
DMPXK1L5HP83
DMPXK1MDHP83
DMPXK1MMHP83
DMPXK1N8HP83
DMPXK1NSHP83
DMPXK1NZHP83
DMPXK1P0HP83
DMPXK1P1HP83
DMPXK1P6HP83
DMPXK1PAHP83
DMPXK1PDHP83
DMPXK1PHHP83
DMPXK1PKHP83
DMPXK1PQHP83
DMPXK1PSHP83
DMPXK1PWHP83
DMPXK1PXHP83
DMPXK1PYHP83
DMPXK1Q2HP83
DMPXK1Q8HP83
DMPXK1QLHP83
DMPXK1QPHP83
DMPXK1QWHP83
DMPXK1QXHP83
DMPXK1R1HP83
DMPXK1RCHP83
DMPXK1RHHP83
DMPXK1RQHP83
DMPXK1RUHP83
DMPXK1RWHP83
DMPXK1SQHP83
DMPXK1SVHP83
DMPXK1T6HP83
DMPXK1TBHP83
DMPXK1TQHP83
DMPXK1TXHP83
DMPXK1U5HP83
DMPXK1UJHP83
DMPXK1V2HP83
DMPXK1V7HP83
DMPXK1V8HP83
DMPXK1VYHP83
DMPXK1W3HP83
DMPXK1W4HP83
DMPXK1W7HP83
DMPXK1WBHP83
DMPXK1WHHP83
DMPXK1WMHP83
DMPXK1WQHP83
DMPXK1WUHP83
DMPXK1WVHP83
DMPXK1XLHP83
DMPXK1Y3HP83
DMPXK1Y8HP83
DMPXK1YFHP83
DMPXK1YHHP83
DMPXK1YWHP83
DMPXK249HP83
DMPXK27UHP83
DMPXK297HP83
DMPXK2ADHP83
DMPXL002HP83
DMPXL006HP83
DMPXL00KHP83
DMPXL00PHP83
DMPXL00RHP83
DMPXL00SHP83
DMPXL00VHP83
DMPXL00WHP83
DMPXL00ZHP83
DMPXL019HP83
DMPXL01VHP83
DMPXL01YHP83
DMPXL02RHP83
DMPXL02YHP83
DMPXL032HP83
DMPXL03EHP83
DMPXL03KHP83
DMPXL04DHP83
DMPXL04KHP83
DMPXL04ZHP83
DMPXL050HP83
DMPXL05CHP83
DMPXL05GHP83
DMPXL05XHP83
DMPXL063HP83
DMPXL064HP83
DMPXL066HP83
DMPXL06YHP83
DMPXL073HP83
DMPXL075HP83
DMPXL07AHP83
DMPXL07XHP83
DMPXL081HP83
DMPXL086HP83
DMPXL087HP83
DMPXL08AHP83
DMPXL08EHP83
DMPXL08FHP83
DMPXL08MHP83
DMPXL08THP83
DMPXL08YHP83
DMPXL08ZHP83
DMPXL092HP83
DMPXL09CHP83
DMPXL09KHP83
DMPXL09YHP83
DMPXL0A1HP83
DMPXL0A5HP83
DMPXL0A6HP83
DMPXL0ABHP83
DMPXL0AEHP83
DMPXL0AHHP83
DMPXL0AMHP83
DMPXL0AQHP83
DMPXL0ASHP83
DMPXL0AVHP83
DMPXL0AXHP83
DMPXL0B0HP83
DMPXL0B6HP83
DMPXL0B9HP83
DMPXL0BZHP83
DMPXL0CDHP83
DMPXL0CGHP83
DMPXL0CSHP83
DMPXL0CZHP83
DMPXL0D3HP83
DMPXL0D9HP83
DMPXL0DCHP83
DMPXL0DGHP83
DMPXL0DRHP83
DMPXL0DXHP83
DMPXL0EDHP83
DMPXM04GHP51
DMPXM04RHP51
DMPXM0JCJ28L
DMPXM0MPHP51
DMPXM1KSHP50
DMPXN011HP51
DMPXN01RHP51
DMPXN02MHP51
DMPXN02QHP51
DMPXN03MHP51
DMPXN03ZHP51
DMPXN04YHP51
DMPXN054HP51
DMPXN059HP51
DMPXN05AHP51
DMPXN06FHP51
DMPXN06JHP51
DMPXN070HP83
DMPXN07PHP51
DMPXN07THP51
DMPXN08MHP83
DMPXN09DHP51
DMPXN09THP51
DMPXN0C7HP51
DMPXN0F3HP51
DMPXN0MCHP51
DMPXN11RHP50
DMPXN131HP50
DMPXN16ZHP50
DMPXN19EHP50
DMPXN1QWHP50
DMPXN1RDHP50
DMPXN23LHP50
DMPXN25EHP50
DMPXN294HP50
DMPXN29RHP50
DMPXN2AXHP50
DMPXN2CUHP50
DMPXN2DHHP50
DMPXN2DMHP50
DMPXN2DXHP50
DMPXN2ECHP50
DMPXN2EFHP50
DMPXN2TUHP50
DMPXN2U8HP50
DMPXN2UUHP50
DMPXN2YGHP50
DMPXN332HP50
DMPXN33UHP50
DMPXN34RHP50
DMPXN350HP50
DMPXN35EHP50
DMPXN388HP50
DMPXN3A8HP50
DMPXN3C9HP50
DMPXN3CYHP50
DMPXN3FAHP50
DMPXN3FSHP50
DMPXN3JCHP50
DMPXN3LJHP50
DMPXN3LNHP50
DMPXN3LSHP50
DMPXN3NJHP50
DMPXN3STHP50
DMPXN407HP50
DMPXN4AZHP50
DMPXN4DCHP50
DMPXP0AFHP83
DMPXP0BGHP83
DMPXP0CCHP83
DMPXP0CQHP83
DMPXP0D2HP83
DMPXP0DGHP83
DMPXP0DPHP83
DMPXP0DQHP83
DMPXP0DVHP83
DMPXP0E1HP83
DMPXP0ERHP83
DMPXP0F7HP83
DMPXP0FLHP83
DMPXP0FMHP83
DMPXP0FNHP83
DMPXP0G5HP83
DMPXP0P9HP51
DMPXP0Q3HP51
DMPXP0QHHP51
DMPXP0Z6HP51
DMPXP237HP50
DMPXQ003HP83
DMPXQ004HP83
DMPXQ005HP83
DMPXQ00BHP83
DMPXQ00CHP83
DMPXQ00FHP83
DMPXQ00HHP83
DMPXQ00JHP83
DMPXQ00KHP83
DMPXQ00PHP83
DMPXQ00THP83
DMPXQ00UHP83
DMPXQ010HP83
DMPXQ015HP83
DMPXQ016HP83
DMPXQ017HP83
DMPXQ01CHP83
DMPXQ01FHP83
DMPXQ01GHP83
DMPXQ01LHP83
DMPXQ01MHP83
DMPXQ01PHP83
DMPXQ01THP83
DMPXQ01WHP83
DMPXQ01YHP83
DMPXQ020HP83
DMPXQ021HP83
DMPXQ025HP83
DMPXQ026HP83
DMPXQ028HP83
DMPXQ029HP83
DMPXQ02CHP83
DMPXQ02HHP83
DMPXQ02THP83
DMPXQ02XHP83
DMPXQ034HP83
DMPXQ038HP83
DMPXQ03CHP83
DMPXQ03DHP83
DMPXQ03FHP83
DMPXQ03MHP83
DMPXQ03NHP83
DMPXQ03VHP83
DMPXQ043HP83
DMPXQ048HP83
DMPXQ04CHP83
DMPXQ04LHP83
DMPXQ04ZHP83
DMPXQ055HP83
DMPXQ057HP83
DMPXQ05FHP83
DMPXQ05GHP83
DMPXQ05UHP83
DMPXQ05VHP83
DMPXQ061HP83
DMPXQ06FHP83
DMPXQ06PHP83
DMPXQ06WHP83
DMPXQ06ZHP83
DMPXQ076HP83
DMPXQ078HP83
DMPXQ07DHP83
DMPXQ07EHP83
DMPXQ07VHP83
DMPXQ07YHP83
DMPXQ083HP83
DMPXQ085HP83
DMPXQ08EHP83
DMPXQ08SHP83
DMPXQ09AHP83
DMPXQ09JHP83
DMPXQ09KHP83
DMPXQ09RHP83
DMPXQ09VHP83
DMPXQ0A0HP83
DMPXQ0A6HP83
DMPXQ0ADHP83
DMPXQ0AQHP83
DMPXQ0BHHP83
DMPXQ0BRHP83
DMPXQ0BWHP83
DMPXQ0BZHP83
DMPXQ0C1HP83
DMPXQ0C3HP83
DMPXQ0C9HP83
DMPXQ0CDHP83
DMPXQ0CGHP83
DMPXQ0CHHP83
DMPXQ0CPHP83
DMPXQ0CQHP83
DMPXQ0CRHP83
DMPXQ0CYHP83
DMPXQ0D0HP83
DMPXQ0D1HP83
DMPXQ0D2HP83
DMPXQ0D4HP83
DMPXQ0D7HP83
DMPXQ0DDHP83
DMPXQ0DEHP83
DMPXQ0DKHP83
DMPXQ0DQHP83
DMPXQ0DXHP83
DMPXQ0DZHP83
DMPXQ0E1HP83
DMPXQ0E3HP83
DMPXQ0E9HP83
DMPXQ0EAHP83
DMPXQ0EBHP83
DMPXQ0EGHP83
DMPXQ0EKHP83
DMPXQ0ELHP83
DMPXQ0ENHP83
DMPXQ0EQHP83
DMPXQ0ERHP83
DMPXQ0ETHP83
DMPXQ0EZHP83
DMPXQ0F1HP83
DMPXQ0F4HP83
DMPXQ0F5HP83
DMPXQ0FBHP83
DMPXQ0FDHP83
DMPXQ0FEHP83
DMPXQ0FKHP83
DMPXQ0FYHP83
DMPXQ0FZHP83
DMPXQ0G5HP83
DMPXQ0G8HP83
DMPXQ0G9HP83
DMPXQ0GGHP83
DMPXQ0GJHP83
DMPXQ0GKHP83
DMPXQ0GLHP83
DMPXQ0GQHP83
DMPXQ0GZHP83
DMPXQ0H2HP83
DMPXQ0H4HP83
DMPXQ0HJHP83
DMPXQ0HRHP83
DMPXQ0HSHP83
DMPXQ0HTHP83
DMPXQ0HUHP83
DMPXQ0HWHP83
DMPXQ0J6HP83
DMPXQ0J8HP83
DMPXQ0JEHP83
DMPXQ0JHHP83
DMPXQ0JPHP83
DMPXQ0JZHP83
DMPXQ0K0HP83
DMPXQ0K5HP83
DMPXQ0K6HP83
DMPXQ0K7HP83
DMPXQ0KAHP83
DMPXQ0KCHP83
DMPXQ0KFHP83
DMPXQ0KGHP83
DMPXQ0KHHP83
DMPXQ0KNHP83
DMPXQ0KPHP83
DMPXQ0KRHP83
DMPXQ0KSHP83
DMPXQ0L2HP83
DMPXQ0L3HP83
DMPXQ0L4HP83
DMPXQ0L6HP83
DMPXQ0LAHP83
DMPXQ0LDHP83
DMPXQ0LJHP83
DMPXQ0LKHP83
DMPXT0STHP50
DMPXT174HP50
DMPXV36MHP50
DMPXV3CPHP50
DMPXV4M3HP50
DMPXV4QFHP50
DMPXV4VVHP50
DMPXV4XXHP50
DMPXV4YPHP50
DMPXV50AHP50
DMPXV51FHP50
DMPXV51UHP50
DMPXV53NHP50
DMPXV545HP50
DMPXV56MHP50
DMPXV5C5HP50
DMPXV5THHP50
DMPXW090HP50
DMPXW09AHP50
DMPXW0AMHP50
DMPXW0AYHP50
DMPXW0B0HP50
DMPXW0B3HP50
DMPXW0BDHP50
DMPXW0BPHP50
DMPXW0BTHP50
DMPXW0BWHP50
DMPXW0CTHP50
DMPXW0CYHP50
DMPXW0D7HP50
DMPXW0DEHP50
DMPXW0DHHP50
DMPXW0DLHP50
DMPXW0E2HP50
DMPXW0EEHP50
DMPXW0FUHP50
DMPXW0GUHP50
DMPXW0J5HP50
DMPXW0LJHP50
DMPXW0WMJ28L
DMPXW1QMHP50
DMPXX05VHP50
DMPXX0F5HP50
DMPXX0PYHP50
DMPXX0Y3HP50
DMPXX12PHP50
DMPXX137HP50
DMPXX13GHP50
DMPXX13HHP50
DMPXX18CHP50
DMPXX1BWHP50
DMPXX1CMHP50
DMPXX1DNHP50
DMPXX1ERHP50
DMPXX1F1HP50
DMPXX1GAHP50
DMPXX26HHP50
DMPY1014HP83
DMPY101EHP83
DMPY1089HP50
DMPY108CHP51
DMPY10F5HP50
DMPY10FQHP50
DMPY10PJHP50
DMPY10RLHP50
DMPY30E0HP50
DMPY30EXHP50
DMPY30Z9J28L
DMPY30ZKJ28L
DMPY4008HP50
DMPY400QHP50
DMPY4013HP50
DMPY4018HP50
DMPY402KHP50
DMPY403DHP50
DMPY404DHP50
DMPY4057HP50
DMPY4061HP50
DMPY4066HP50
DMPY407ZHP50
DMPY4088HP50
DMPY408LHP50
DMPY409SHP50
DMPY409YHP50
DMPY40ACHP50
DMPY40ASHP50
DMPY40AXHP50
DMPY40BEHP50
DMPY40BSHP50
DMPY40DTHP50
DMPY40DXHP50
DMPY40HLHP50
DMPY40HNHP50
DMPY40J1HP50
DMPY40MNHP50
DMPY40MRHP50
DMPY40MWHP50
DMPY7001HP50
DMPY7008HP50
DMPY700UHP50
DMPY701GHP50
DMPY702KHP50
DMPY70C2HP50
DMPY70C3HP50
DMPY70CAHP50
DMPY70CEHP50
DMPY70CRHP50
DMPY70CYHP50
DMPY8002HP83
DMPY8007HP83
DMPY800FHP50
DMPY800KHP50
DMPY800SHP83
DMPY800THP50
DMPY800THP83
DMPY800YHP50
DMPY800ZHP50
DMPY8010HP50
DMPY8019HP50
DMPY801DHP50
DMPY801FHP83
DMPY80Q0J28L
DMPY902LHP50
DMPY907EJ28L
DMPY90ABHP50
DMPY90BWHP50
DMPY90CVHP50
DMPY90DUJ28L
DMPY90EPJ28L
DMPY90H6HP50
DMPY90KQHP50
DMPY90KXJ28L
DMPY90L6J28L
DMPY90NCJ28L
DMPY90NKJ28L
DMPY90NNJ28L
DMPY90QFJ28L
DMPYC002HP50
DMPYC006HP50
DMPYC007HP83
DMPYC008HP50
DMPYC00AHP50
DMPYC00CHP50
DMPYC00CHP83
DMPYC00FHP50
DMPYC00JHP50
DMPYC00PHP50
DMPYC00RHP83
DMPYC00SHP83
DMPYC00UHP50
DMPYC00VHP83
DMPYC00ZHP50
DMPYC010HP50
DMPYC013HP83
DMPYC014HP83
DMPYC015HP83
DMPYC016HP51
DMPYC019HP83
DMPYC01FHP83
DMPYC01HHP50
DMPYC01HHP83
DMPYC01JHP83
DMPYC01KHP83
DMPYC01LHP51
DMPYC01LHP83
DMPYC01UHP51
DMPYC01ZHP50
DMPYC020HP50
DMPYC022HP50
DMPYC023HP50
DMPYC026HP50
DMPYC02NHP50
DMPYC02QHP51
DMPYC02XHP50
DMPYC02YHP50
DMPYC038HP50
DMPYC03CHP50
DMPYC03PHP50
DMPYC03QHP50
DMPYC03RHP50
DMPYC03UHP50
DMPYC03WHP50
DMPYC041HP50
DMPYC045HP50
DMPYC04CHP50
DMPYC056HP51
DMPYD004HP50
DMPYD006HP50
DMPYD00AHP83
DMPYD00BHP50
DMPYD00BHP83
DMPYD00CHP50
DMPYD00DHP50
DMPYD00FHP50
DMPYD00HHP50
DMPYD00JHP50
DMPYD00KHP50
DMPYD00LHP50
DMPYD00RHP50
DMPYD00RHP83
DMPYD00UHP50
DMPYD00ZHP50
DMPYD012HP50
DMPYD016HP50
DMPYD016HP83
DMPYD018HP83
DMPYD019HP83
DMPYD01AHP50
DMPYD01DHP50
DMPYD01GHP50
F6QZ800UHP83
HeartMate Touch Communication System, Rental Model Number: HMT1100-R
UDI-DI Code: 05415067033758
Serial Numbers:
DMPWL0Y3HP83
DMPXQ006HP83
DMPXQ031HP83
DMPXQ0C7HP83
DMPXQ0JMHP83
HeartMate Touch Communication System Model Number: HMT1150
UDI-DI Code: 05415067032041
Serial Numbers:
DMPVJAYAHP51
DMPWL1A2HP83
DMPWR6PNHP50
DMPWV0DGHP51
DMPX12NSHP51
DMPX13DAHP51
DMPX22F4HP50
DMPX31NWHP51
DMPX50PBHP50
DMPX54VFHP50
DMPX55LLHP50
DMPX55MVHP50
DMPX7148HP51
DMPX71E7HP51
DMPX71RNHP51
DMPX75RFHP50
DMPX76JYHP50
DMPX77NMHP50
DMPX77RSHP50
DMPX77V5HP50
DMPX781ZHP50
DMPX789JHP50
DMPX78N2HP50
DMPX78Q5HP50
DMPX78W2HP50
DMPXC02HHP50
DMPXC04MHP50
DMPXC05PHP84
DMPXC0E1HP50
DMPXC0NKHP51
DMPXD1GAHP51
DMPXD1PRHP51
DMPXG1M1HP51
DMPXJ0L8HP83
DMPXK01LHP83
DMPXK05RHP83
DMPXK0BMHP83
DMPXK0NSHP83
DMPXK1QAHP83
DMPXK27PHP83
DMPXL0EKHP83
DMPXN01WHP51
DMPXN03PHP51
DMPXN06KHP51
DMPXN08NHP51
DMPXN0J7HP51
DMPXN1AJHP50
DMPXN1C3HP50
DMPXN1SJHP50
DMPXN2A7HP50
DMPXN2DUHP50
DMPXN2XFHP50
DMPXN3B3HP50
DMPXN3HWHP50
DMPXN3LGHP50
DMPXP0E2HP83
DMPXP0F4HP83
DMPXP0FZHP83
DMPXP0PBHP51
DMPXP0QAHP51
DMPXQ021HP51
DMPXQ041HP83
DMPXQ068HP83
DMPXQ097HP83
DMPXQ09PHP83
DMPXQ0AYHP83
DMPXQ0KEHP83
DMPXR0B4J28L
DMPXV4ZXHP50
DMPXW0B6HP50
DMPXW0KUHP50
DMPXW0RQJ28L
DMPXW0YAJ28L
DMPXX07SHP50
DMPXX0VBHP51
DMPXX11QHP50
DMPXX1FJHP50
DMPY10P5HP50
DMPY30ZJJ28L
DMPY40ELJ28L
DMPY800XHP50
DMPY90JKJ28L
DMPY90KBJ28L
DMPY90N7J28L
DMPYC03HHP51
DMPYC03KHP50
DMPYC05RHP51
|
Recalling Firm/ Manufacturer |
Thoratec Corp. 6035 Stoneridge Dr Pleasanton CA 94588-3270
|
| For Additional Information Contact | Ms Beth Crain 651-756-5129 |
Manufacturer Reason for Recall | Due to software and controller systems that results in unexpected pump start or pump stop. |
FDA Determined Cause 2 | Software design |
| Action | On 01/03/2024, the firm sent an Urgent Medical Device Correction Letter via hand delivery to U.S. customers informing them that Abbott has received complaints for the HeartMate Touch Communication System application (App) version 1.0.32 due to unintentional start/stop of the HeartMate 3 Left Ventricle Assist Device (LVAD).
-Customer should be able to identify the software version of the HeartMate Touch App by tapping "Menu" and "About" on the left-hand panel on the HeartMate Touch screen.
-When the "STOP PUMP" sequence needs to be performed, follow the instructions in the Instructions For Use (IFU), Chapter 4 HeartMate Touch Communication System Pages 4-58 to 4-59.
-The user can prevent the unintentional LVAD start/stop from occurring by following the IFU and these additional guidance and recommendations. Note that communication may also be interrupted if the Wireless Adaptor is disconnected from the HeartMate Touch system prior to completion of the STOP PUMP sequence.
-Do not disconnect the HeartMate System Controller white cable or the Wireless Adapter from the Power Module until STOP PUMP sequence screen with the red progress bar is no longer visible and the STOP PUMP screen changes automatically.
-During pump priming, do not disconnect the pump until the timer reaches zero, the pump stops, and the Priming is complete message appears as instructed per the IFU.
Abbott is developing and will release a software update to prevent the sequence of events described above. This product correction will be included in an upcoming release of the HeartMate Touch software. An Abbott Representative will contact you for a software upgrade beginning Q2 2024.
For questions, contact Abbott Technical Support at 1-800-456-1477. |
| Quantity in Commerce | 1,560 systems as of 12/8/2023 |
| Distribution | Worldwide distribution. US nationwide, Australia, Austria, Canada, Colombia, Czech Republic, Germany, Hong Kong, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Martinique, New Zealand, Poland, Romania, Saudi Arabia, Singapore, Spain, Switzerland, Thailand and United Arab Emirates.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = DSQ
|
|
|
|