| | Class 2 Device Recall Prestige Coil System |  |
| Date Initiated by Firm | January 03, 2024 |
| Date Posted | February 08, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1042-2024 |
| Recall Event ID |
93751 |
| 510(K)Number | K220699 |
| Product Classification |
Device, Vascular, for Promoting Embolization - Product Code KRD
|
| Product | The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to embolize peripheral vascular abnormalities. The devices are to be placed to create blood stasis, reducing flow into the anomaly, and thrombosing the target site. The Prestige Coil System is comprised of the implant coil and a pusher assembly. There are a total of 81 Prestige coil models.
Vascular Embolization Device
Prestige Coil Model Numbers and Product description:
PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack
PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18
PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac
PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac
PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack
PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack
PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac
PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack
PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack
PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack
PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack
PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack
PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac
PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac
PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack
PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack
PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack
PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack
PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18
PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac
PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac
PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac
PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack
PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack
PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack
PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack
PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack
PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack
PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack
PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac
PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac
PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack
PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac
PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack
PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac
PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack
PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack
PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac
PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular
PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac
PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac
PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac
PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack
PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack
PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack
PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack
PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack
PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack
PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18
PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack
PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack
PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack
PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack
PRES0930H |
| Code Information |
Refer to Attachment 1 for list of affected products including the products' Lot Number, Product model number, UDI and expiration date.
Prestige Coils ( Lot Number/Model Number/Product Description/UDI-DI Code:
F201000210 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F201000231 PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18 00818053023217
F201000319 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac 00818053027307
F201000320 PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac 00818053027314
F201000321 PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack 00818053027444
F201000322 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F201000323 PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac 00818053027178
F201000324 PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack 00818053027185
F201000325 PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack 00818053027192
F201000326 PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack 00818053027208
F201000329 PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack 00818053027550
F201000330 PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack 00818053027581
F201000331 PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac 00818053027109
F201000332 PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac 00818053027147
F201000333 PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack 00818053027215
F201000334 PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack 00818053027222
F201000335 PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack 00818053027239
F201000336 PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack 00818053027246
F201000450 PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 00818053023224
F201000451 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac 00818053027321
F201000452 PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac 00818053027338
F201000453 PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac 00818053027383
F201000454 PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack 00818053027536
F201000455 PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack 00818053027574
F201000516 PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack 00818053027604
F201000517 PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack 00818053027611
F201000518 PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack 00818053027642
F201000520 PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack 00818053027659
F201000521 PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack 00818053027666
F201000522 PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac 00818053027277
F201000523 PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac 00818053027284
F201000524 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F201000526 PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac 00818053027154
F201000527 PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack 00818053027598
F201000564 PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac 00818053027376
F201000571 PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack 00818053027505
F201000572 PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack 00818053027536
F201000575 PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac 00818053027284
F201000788 PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular 00818053027703
F201000789 PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac 00818053027291
F201000790 PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac 00818053027352
F201000791 PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac 00818053027413
F201000792 PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack 00818053027444
F201000793 PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack 00818053027598
F201000794 PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack 00818053027185
F201000795 PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack 00818053027192
F201001014 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack 00818053023989
F201001025 PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack 00818053023996
F201001126 PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 00818053023248
F201001139 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack 00818053023989
F201100055 PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack 00818053024153
F201100056 PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack 00818053024177
F201100057 PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack 00818053024207
F201100058 PRES0930HRG10 Prestige Coil 9mm x 30cm Helical-10 Regular 00818053024337
F201100059 PRES0102CXPPL Prestige Coil 1mm x 2cm Prestige Plus Complex Xtra Pac 00818053027260
F201100060 PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac 00818053027277
F201100061 PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac 00818053027314
F201100062 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac 00818053027321
F201100063 PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack 00818053027505
F201100064 PRES0615CPKPL Prestige Coil 6mm x 15cm Prestige Plus Complex Pack 00818053027529
F201100065 PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack 00818053027574
F201100066 PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac 00818053027109
F201100067 PRES0101CXPPL Prestige Coil 1mm x 1cm Prestige Plus Complex Xtra Pac 00818053027253
F201100132 PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack 00818053024153
F201100133 PRES0308HPK10 Prestige Coil 3mm x 8cm Helical-10 Pack 00818053024160
F201100134 PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack 00818053024177
F201100135 PRES0508HPK10 Prestige Coil 5mm x 8cm Helical-10 Pack 00818053024221
F201100136 PRES0930HRG10 Prestige Coil 9mm x 30cm Helical-10 Regular 00818053024337
F201100137 PRES0101CXPPL Prestige Coil 1mm x 1cm Prestige Plus Complex Xtra Pac 00818053027253
F201100138 PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac 00818053027147
F201100139 PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack 00818053027246
F201100229 PRES0615HRG10 Prestige Coil 6mm x 15cm Helical-10 Regular 00818053024276
F201100247 PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 00818053023248
F201100248 PRES1860COM18 Prestige Coil 18mm x 60cm Complex-18 00818053023279
F201100249 PRES0102CXPPL Prestige Coil 1mm x 2cm Prestige Plus Complex Xtra Pac 00818053027260
F201100250 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F201100253 PRES0510CPKPL Prestige Coil 5mm x 10cm Prestige Plus Complex Pack 00818053027499
F201100255 PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack 00818053027505
F201100257 PRES0103HXPPL Prestige Coil 1mm x 3cm Prestige Plus Helical Xtra Pac 00818053027123
F201100258 PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack 00818053027208
F201100266 PRES0308HPK10 Prestige Coil 3mm x 8cm Helical-10 Pack 00818053024160
F201100267 PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack 00818053027215
F201100268 PRES0620HRG10 Prestige Coil 6mm x 20cm Helical-10 Regular 00818053024283
F201100291 PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack 00818053027550
F201100296 PRES0510HPK10 Prestige Coil 5mm x 10cm Helical-10 Pack 00818053024238
F201100297 PRES2065COM18 Prestige Coil 20mm x 65cm Complex-18 00818053023286
F201100320 PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 00818053023248
F201100321 PRES2465COM18 Prestige Coil 24mm x 65cm Complex-18 00818053023309
F201100403 PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack 00818053027185
F201100404 PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack 00818053027192
F201100405 PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack 00818053027215
F201100412 PRES0412CXP10 Prestige Coil 4mm x 12cm Complex-10 Xtra Pack 00818053027628
F201100499 PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac 00818053027376
F201200013 PRES0304HPK10 Prestige Coil 3mm x 4cm Helical-10 Pack 00818053024146
F201200015 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack 00818053023989
F201200016 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F201200017 PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack 00818053027550
F201200018 PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack 00818053027581
F201200019 PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack 00818053027208
F201200020 PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack 00818053023996
F201200377 PRES1655COM18 Prestige Coil 16mm x 55cm Complex-18 00818053023262
F201200378 PRES1860COM18 Prestige Coil 18mm x 60cm Complex-18 00818053023279
F201200765 PRES0408HPK10 Prestige Coil 4mm x 8cm Helical-10 Pack 00818053024191
F201200766 PRES0303CXPPL Prestige Coil 3mm x 3cm Prestige Plus Complex Xtra Pac 00818053027369
F201200767 PRES0407CPKPL Prestige Coil 4mm x 7cm Prestige Plus Complex Pack 00818053027437
F201200768 PRES4510CPKPL Prestige Coil 4.5mm x 10cm Prestige Plus Complex Pack 00818053027475
F201200771 PRES0204HXPPL Prestige Coil 2mm x 4cm Prestige Plus Helical Xtra Pac 00818053029943
F201200772 PRES0208HXPPL Prestige Coil 2mm x 8cm Prestige Plus Helical Xtra Pac 00818053029981
F201200773 PRES0515HPKPL Prestige Coil 5mm x 15cm Prestige Plus Helical Pack 00818053029684
F201200789 PRES1030HRG10 Prestige Coil 10mm x 30cm Helical-10 Regular 00818053024344
F201200920 PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 00818053023224
F201200921 PRES1655COM18 Prestige Coil 16mm x 55cm Complex-18 00818053023262
F201200922 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack 00818053023989
F201200923 PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack 00818053023996
F201200924 PRES0608HPK10 Prestige Coil 6mm x 8cm Helical-10 Pack 00818053024252
F201200925 PRES0730HRG10 Prestige Coil 7mm x 30cm Helical-10 Regular 00818053024306
F201200926 PRES0930HRG10 Prestige Coil 9mm x 30cm Helical-10 Regular 00818053024337
F201200927 PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac 00818053027413
F201201093 PRES0304HPK10 Prestige Coil 3mm x 4cm Helical-10 Pack 00818053024146
F201201095 PRES0255HXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Helical Xtra Pac 00818053027161
F201201101 PRES0720HRG10 Prestige Coil 7mm x 20cm Helical-10 Regular 00818053024290
F201201102 PRES1030HRG10 Prestige Coil 10mm x 30cm Helical-10 Regular 00818053024344
F201201103 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac 00818053027307
F201201104 PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack 00818053027505
F201201105 PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac 00818053027109
F201201106 PRES0206HXPPL Prestige Coil 2mm x 6cm Prestige Plus Helical Xtra Pac 00818053029967
F201201107 PRES0310HXPPL Prestige Coil 3mm x 10cm Prestige Plus Helical Xtra Pac 00810068560060
F201201108 PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac 00818053027178
F201201109 PRES0410HPKPL Prestige Coil 4mm x 10cm Prestige Plus Helical Pack 00818053029646
F201201110 PRES0608HPK10 Prestige Coil 6mm x 8cm Helical-10 Pack 00818053024252
F201201111 PRES0407CPKPL Prestige Coil 4mm x 7cm Prestige Plus Complex Pack 00818053027437
F201201112 PRES1030CPKPL Prestige Coil 10mm x 30cm Prestige Plus Complex Pack 00818053029745
F201201113 PRES0255HXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Helical Xtra Pac 00818053027161
F210100021 PRES0408HPK10 Prestige Coil 4mm x 8cm Helical-10 Pack 00818053024191
F210100022 PRES0608HPK10 Prestige Coil 6mm x 8cm Helical-10 Pack 00818053024252
F210100023 PRES0830HRG10 Prestige Coil 8mm x 30cm Helical-10 Regular 00818053024320
F210100024 PRES0404CPKPL Prestige Coil 4mm x 4cm Prestige Plus Complex Pack 00818053027420
F210100025 PRES0255HXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Helical Xtra Pac 00818053027161
F210100026 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack 00818053023989
F210100027 PRES0304HPK10 Prestige Coil 3mm x 4cm Helical-10 Pack 00818053024146
F210100028 PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack 00818053024207
F210100029 PRES0510HPK10 Prestige Coil 5mm x 10cm Helical-10 Pack 00818053024238
F210100030 PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac 00818053027352
F210100032 PRES0103HXPPL Prestige Coil 1mm x 3cm Prestige Plus Helical Xtra Pac 00818053027123
F210100033 PRES0515HPKPL Prestige Coil 5mm x 15cm Prestige Plus Helical Pack 00818053029684
F210100475 PRES0308HPK10 Prestige Coil 3mm x 8cm Helical-10 Pack 00818053024160
F210100476 PRES0408HPK10 Prestige Coil 4mm x 8cm Helical-10 Pack 00818053024191
F210100477 PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack 00818053024207
F210100478 PRES0510HPK10 Prestige Coil 5mm x 10cm Helical-10 Pack 00818053024238
F210100479 PRES0610HPK10 Prestige Coil 6mm x 10cm Helical-10 Pack 00818053024269
F210100480 PRES0730HRG10 Prestige Coil 7mm x 30cm Helical-10 Regular 00818053024306
F210100481 PRES1030HRG10 Prestige Coil 10mm x 30cm Helical-10 Regular 00818053024344
F210100482 PRES4510CPKPL Prestige Coil 4.5mm x 10cm Prestige Plus Complex Pack 00818053027475
F210100483 PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack 00818053027505
F210100484 PRES0820CPKPL Prestige Coil 8mm x 20cm Prestige Plus Complex Pack 00818053027567
F210100485 PRES1030CPKPL Prestige Coil 10mm x 30cm Prestige Plus Complex Pack 00818053029745
F210100486 PRES0206HXPPL Prestige Coil 2mm x 6cm Prestige Plus Helical Xtra Pac 00818053029967
F210100487 PRES0208HXPPL Prestige Coil 2mm x 8cm Prestige Plus Helical Xtra Pac 00818053029981
F210100488 PRES0255HXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Helical Xtra Pac 00818053027161
F210100489 PRES0310HXPPL Prestige Coil 3mm x 10cm Prestige Plus Helical Xtra Pac 00810068560060
F210100490 PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac 00818053027178
F210100491 PRES0410HPKPL Prestige Coil 4mm x 10cm Prestige Plus Helical Pack 00818053029646
F210100492 PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack 00818053027185
F210100493 PRES0515HPKPL Prestige Coil 5mm x 15cm Prestige Plus Helical Pack 00818053029684
F210100494 PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack 00818053027192
F210100495 PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack 00818053027208
F210100496 PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack 00818053027215
F210100497 PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack 00818053027222
F210100498 PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack 00818053027239
F210100499 PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack 00818053027246
F210100500 PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac 00818053027291
F210100901 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac 00818053027307
F210100902 PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac 00818053027352
F210100903 PRES0303CXPPL Prestige Coil 3mm x 3cm Prestige Plus Complex Xtra Pac 00818053027369
F210100904 PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac 00818053027383
F210100905 PRES0355CXPPL Prestige Coil 3.5mm x 5cm Prestige Plus Complex Xtra Pac 00818053027406
F210100935 PRES0615HRG10 Prestige Coil 6mm x 15cm Helical-10 Regular 00818053024276
F210100936 PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac 00818053027314
F210100937 PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack 00818053027444
F210100938 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F210100939 PRES0510CPKPL Prestige Coil 5mm x 10cm Prestige Plus Complex Pack 00818053027499
F210100940 PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack 00818053027536
F210100941 PRES0204HXPPL Prestige Coil 2mm x 4cm Prestige Plus Helical Xtra Pac 00818053029943
F210100942 PRES0410HPKPL Prestige Coil 4mm x 10cm Prestige Plus Helical Pack 00818053029646
F210100943 PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack 00818053027215
F210200379 PRES0720HRG10 Prestige Coil 7mm x 20cm Helical-10 Regular 00818053024290
F210200380 PRES0830HRG10 Prestige Coil 8mm x 30cm Helical-10 Regular 00818053024320
F210200382 PRES0101CXPPL Prestige Coil 1mm x 1cm Prestige Plus Complex Xtra Pac 00818053027253
F210200383 PRES0102CXPPL Prestige Coil 1mm x 2cm Prestige Plus Complex Xtra Pac 00818053027260
F210200384 PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac 00818053027277
F210200385 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac 00818053027307
F210200386 PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac 00818053027338
F210200387 PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac 00818053027376
F210200388 PRES0355CXPPL Prestige Coil 3.5mm x 5cm Prestige Plus Complex Xtra Pac 00818053027406
F210200389 PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac 00818053027413
F210200390 PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack 00818053027598
F210200391 PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac 00818053027147
F210200440 PRES0820HRG10 Prestige Coil 8mm x 20cm Helical-10 Regular 00818053024313
F210200457 PRES0620HRG10 Prestige Coil 6mm x 20cm Helical-10 Regular 00818053024283
F210200460 PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack 00818053027451
F210200539 PRES0407CPKPL Prestige Coil 4mm x 7cm Prestige Plus Complex Pack 00818053027437
F210300005 PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack 00818053027574
F210300007 PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack 00818053027444
F210300017 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac 00818053027307
F210300024 PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack 00818053027574
F210300104 PRES0930HRG10 Prestige Coil 9mm x 30cm Helical-10 Regular 00818053024337
F210300177 PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac 00818053027383
F210300183 PRES1030CPKPL Prestige Coil 10mm x 30cm Prestige Plus Complex Pack 00818053029745
F210300185 PRES0410HPKPL Prestige Coil 4mm x 10cm Prestige Plus Helical Pack 00818053029646
F210300381 PRES0208HXPPL Prestige Coil 2mm x 8cm Prestige Plus Helical Xtra Pac 00818053029981
F210601079 PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack 00818053027192
F230601801 PRES1655COM18 Prestige Coil 16mm x 55cm Complex-18 00840303708094
F230601818 PRES2065COM18 Prestige Coil 20mm x 65cm Complex-18 00840303708117
F230601877 PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 00840303708070
F230602699 PRES2465COM18 Prestige Coil 24mm x 65cm Complex-18 00840303708131
F230602711 PRES1860COM18 Prestige Coil 18mm x 60cm Complex-18 00840303708100
F230604171 PRES2265COM18 Prestige Coil 22mm x 65cm Complex-18 00840303708124
F230604173 PRES1860COM18 Prestige Coil 18mm x 60cm Complex-18 00840303708100
F230604175 PRES2065COM18 Prestige Coil 20mm x 65cm Complex-18 00840303708117
F230604374 PRES2465COM18 Prestige Coil 24mm x 65cm Complex-18 00840303708131
F230605423 PRES2265COM18 Prestige Coil 22mm x 65cm Complex-18 00840303708124
F230605479 PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 00840303708070
F230605542 PRES1655COM18 Prestige Coil 16mm x 55cm Complex-18 00840303708094
F230605560 PRES2265COM18 Prestige Coil 22mm x 65cm Complex-18 00840303708124
F230700403 PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 00840303708056
F230700467 PRES0827COM18 Prestige Coil 8mm x 27cm Complex-18 00840303708018
F230700759 PRES1034COM18 Prestige Coil 10mm x 34cm Complex-18 00840303708032
F230700883 PRES1034COM18 Prestige Coil 10mm x 34cm Complex-18 00840303708032
F231100604 PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac 00818053027284 |
Recalling Firm/ Manufacturer |
BALT USA, LLC 29 Parker Ste 100 Irvine CA 92618-1667
|
| For Additional Information Contact | Mr. David Vu 949-788-1443 |
Manufacturer Reason for Recall | Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli. |
FDA Determined Cause 2 | Material/Component Contamination |
| Action | On 01/03/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via priority mail to customers informing them that Balt USA has became aware of a situation where discoloration along the delivery system pusher component of the Optima Coil System was identified during in-house inspection of released finished good products. The discoloration was observed on the outside of the PET jacket along the hypotube component.
Customers are instructed to:
1) Stop or cease use of the affected recall product.
2) Inform their hospital staff, such as safety officers, pharmacists, purchasing agent, head of neuroradiology and the neuroradiology staff, physicians including but not limited to endovascular neurosurgeons and interventional neurologists, as well as any other person if deemed necessary of the recall.
3) Identify and locate the Optima Coil System products concerned by this recall procedure.
4) Collect and put in quarantine the Optima Coil System products concerned by this recall procedure, then return them to Balt USA through the usual RGA (Return Goods Authorization) procedure by contacting Balt USA s Customer Service department.
5) Keep informed by Balt USA about the status of every unit of Optima Coil System products concerned by this recall procedure.
For questions or additional assistance, contact Balt USA Quality Department at:
Contact:
Quality Department
email: QA@baltgroup.com
Balt USA, LLC
29 Parker | Irvine CA, 92618 | USA
Telephone: +1 949.788.1443 | Fax: +1 949.788.1444
|
| Quantity in Commerce | 2802 coils |
| Distribution | U.S.: AZ, CA, CO, FL, GA, HI, IL, KS, MA, ME, MO, MS, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI and Puerto Rico
O.U.S.: Argentina, Australia, Brazil, Canada, Chile, Columbia, Dominican Republic, Ecuador, France, Indonesia, India, Japan, Korea, Malaysia, Mexico, Panama, Paraguay, Peru, Russia, Uruguay, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = KRD
|
|
|
|