| Class 2 Device Recall VITROS 4600 Immunodiagnostic System | |
Date Initiated by Firm | January 26, 2024 |
Date Posted | March 01, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-1230-2024 |
Recall Event ID |
93927 |
510(K)Number | K031924 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished) |
Code Information |
VITROS System Software Versions 3.8.0 or 3.8.1
Regular UDI-DI 10758750012343
Refurbished UD-DI 10758750033201
Serial Numbers:
US:
J46000208
J46000209
J46000240
J46000497
J46000864
J46000865
J46000898
J46001030
J46001064
J46001125
J46001132
J46001293
J46001462
J46001703
J46001704
OUS:
J46000150 J46000933
J46000162 J46001029
J46000190 J46001041
J46000247 J46001042
J46000296 J46001087
J46000393 J46001119
J46000410 J46001186
J46000478 J46001187
J46000499 J46001210
J46000531 J46001217
J46000564 J46001289
J46000617 J46001541
J46000772 J46001565
J46000788 J46001571
J46000800 J46001609
J46000801 J46001629
J46000812 J46001639
J46000821 J46001645
J46000855 J46001680
J46000859 J46000948
J46000861 J46000983
J46000910
|
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Mr. Joe Falvo 585-453-3000 |
Manufacturer Reason for Recall | A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported. |
FDA Determined Cause 2 | Software Design Change |
Action | On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules.
The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024.
If you have further questions, please contact Global Services Organization at 1-800-421-3311. |
Quantity in Commerce | 15 US; 43 OUS |
Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil,
Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico,
Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
|
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|