• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmMarch 22, 2024
Date PostedAugust 22, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2714-2024
Recall Event ID 94356
Product Classification Orthopedic tray - Product Code OJH
ProductMedline medical procedure kits labeled as follows: a) ARTHROSCOPY, REF DYNJ909699; b) BACK CDS, REF CDS984000C; c) EXTREMITY, REF DYNJ900847A; d) HAND-LF, REF DYNJ902619G; e) HIP PINNING CDS, REF CDS984009C; f) HIP PINNING/TFN/ANTERIOR HIP, REF DYNJ902380K; g) KNEE, REF DYNJ904771B; h) KNEE ARTHROSCOPY, REF DYNJ905519D; i) KNEE ARTHROSCOPY, REF DYNJ906273F; j) KNEE ARTHROSCOPY, REF DYNJ909747; k) KNEE ARTHROSCOPY V, REF DYNJS2029D; l) LAMINECTOMY, REF DYNJ907159A; m) LOWER EXTREMITY, REF DYNJ904767A; n) LOWER EXTREMITY, REF DYNJ905518F; o) LOWER EXTREMITY, REF DYNJ909938; p) MINOR EXTREMITY, REF DYNJ909939; q) PODIATRY, REF DYNJ906275D; r) SHOULDER, REF DYNJ904769C; s) SHOULDER, REF DYNJ905511D; t) SHOULDER, REF DYNJ909749; u) SHOULDER ARTHROSCOPY, REF DYNJ906274F; v) SHOULDER ARTHROSCOPY CDS, REF CDS984008C; w) SOUTH EXTREMITY, REF DYNJ909212A; x) SPINE, REF DYNJ905512F; y) TOTAL, REF DYNJ909704; z) TOTAL HIP, REF DYNJ907157A; aa) TOTAL HIP BASIC, REF DYNJ902368K; bb) TOTAL JOINT, REF DYNJ905513G; cc) TOTAL JOINT, REF DYNJ908375C; dd) TOTAL JOINT II, REF DYNJS2027F; ee) TOTAL KNEE, REF DYNJ907158A; ff) TOTAL KNEE CDS, REF CDS983991C; gg) TOTAL SHOULDER, REF DYNJ902367J; hh) TURK-HAND, REF DYNJ905852C; ii) UPPER EXTREMITY, REF DYNJ905514D
Code Information a) REF DYNJ909699, UDI/DI 40195327446759 (case), 10195327446758 (unit), Lot Numbers: DYNJ909699, DYNJ909699, DYNJ909699, DYNJ909699, DYNJ909699, DYNJ909699; b) REF CDS984000C, UDI/DI 40889942011552 (case), 10889942011551 (unit), Lot Numbers: 22GBT604, 22IBI040, 23ABA955, 23DBT170, 23EBU747, 23HBB409; c) REF DYNJ900847A, UDI/DI 40888277771209 (case), 10888277771208 (unit), Lot Numbers: d) REF DYNJ902619G, UDI/DI 40195327214488 (case), 10195327214487 (unit), Lot Numbers: 22FBG650, 22FBK152, 22FBW287, 22HBA416, 22HBG372, 22HBS529, 22JBA184, 23EBC466, 23HBL953; e) REF CDS984009C, UDI/DI 40888277768971 (case), 10888277768970 (unit), Lot Numbers: 22FBL426, 22GBU282, 23CBD236, 23HBY917; f) REF DYNJ902380K, UDI/DI 40195327507009 (case), 10195327507008 (unit), Lot Numbers: 23KBX257; g) REF DYNJ904771B, UDI/DI 40195327137190 (case), 10195327137199 (unit), Lot Numbers: 22EMB001, 22IMF542, 22KMC063, 23AME421, 23CMC716, 23DME033, 23FMI125, 23HMC273, 23HMI602, 23LME735; h) REF DYNJ905519D, UDI/DI 40193489967396 (case), 10193489967395 (unit), Lot Numbers: 22GMG221, 22HMB217, 22LMJ233, 23AME962, 23AMI318, 23AMJ640, 23DMI673, 23IMG465, 23LMD163; i) REF DYNJ906273F, UDI/DI 40195327511013 (case), 10195327511012 (unit), Lot Numbers: 23JBH985; j) REF DYNJ909747, UDI/DI 40195327468201 (case), 10195327468200 (unit), Lot Numbers: 23HBZ535, 23KBF559; k) REF DYNJS2029D, UDI/DI 40193489214414 (case), 10193489214413 (unit), Lot Numbers: 22IDA884, 22LDA526, 23CDA885, 23FDA459, 23JDB741; l) REF DYNJ907159A, UDI/DI 40193489970419 (case), 10193489970418 (unit), Lot Numbers: 22EME489, 22FMH850, 22IMA282, 22JME894, 22LME340, 23AMB899, 23AMB937, 23AMF082, 23AMG074, 23AMJ063, 23BMF089, 23CMA038, 23FMF222, 23GMB990, 23GMF747, 23HMH495, 23JMC384, 23LMD528; m) REF DYNJ904767A, UDI/DI 40889942850953 (case), 10889942850952 (unit), Lot Numbers: 22EMI503, 22IMC879, 22KMD093, 22LME746, 23AMC120, 23DMA417, 23DMD168, 23FME590, 23GMH321, 23HMG168; n) REF DYNJ905518F, UDI/DI 40195327040407 (case), 10195327040406 (unit), Lot Numbers: 22FMI412, 22HMD694, 22IMA909, 22JMC316, 22KMC860, 22LMD325, 23AMF318, 23BMG871, 23EMF948, 23FMA357, 23GMD311, 23HMG163, 23JMA417, 23KMB755, 23LME256; o) REF DYNJ909938, UDI/DI 40195327511228 (case), 10195327511227 (unit), Lot Numbers: 23JBI008; p) REF DYNJ909939, UDI/DI 40195327511235 (case), 10195327511234 (unit), Lot Numbers: 23JBH778, 23KBU569; q) REF DYNJ906275D, UDI/DI 40195327510986 (case), 10195327510985 (unit), Lot Numbers: 23JBI291, 23LBO829; r) REF DYNJ904769C, UDI/DI 40195327137039 (case), 10195327137038 (unit), Lot Numbers: 22FMB469, 22IMA132, 22KMD488, 22LME616, 23GMF950, 23HME593, 23JMG901, 23KMG320; s) REF DYNJ905511D, UDI/DI 40193489967549 (case), 10193489967548 (unit), Lot Numbers: DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D; t) REF DYNJ909749, UDI/DI 40195327468225 (case), 10195327468224 (unit), Lot Numbers: 23IBM342; u) REF DYNJ906274F, UDI/DI 40195327511020 (case), 10195327511029 (unit), Lot Numbers: 23JBQ500, 23LBI456; v) REF CDS984008C, UDI/DI 40888277768988 (case), 10888277768987 (unit), Lot Numbers: 22GBL020, 22JBE903, 23ABI396, 23ABM540, 23CBO804; w) REF DYNJ909212A, UDI/DI 40195327493241 (case), 10195327493240 (unit), Lot Numbers: 23JBA160, 23KBA970; x) REF DYNJ905512F, UDI/DI 40195327225378 (case), 10195327225377 (unit), Lot Numbers: 22JME529, 22KMD611, 22LMI893, 23BMI268, 23CMF328, 23FMG242, 23HMH500, 23IMC073, 23IMG464, 23JMI252, 23LMC546; y) REF DYNJ909704, UDI/DI 40195327446919 (case), 10195327446918 (unit), Lot Numbers: 23GBN059, 23KBA374; z) REF DYNJ907157A, UDI/DI 40193489970358 (case), 10193489970357 (unit), Lot Numbers: 22FMD804, 22JMG489, 22LMC517, 23BMA251, 23CMC137, 23CME146, 23DMA129, 23EMF952, 23FMG076, 23GMI555, 23IME182, 23JMI348, 23KMA447; aa) REF DYNJ902368K, UDI/DI 40195327505890 (case), 10195327505899 (unit), Lot Numbers: 23JBI661, 23KBT583; bb) REF DYNJ905513G, UDI/DI 40193489968027 (case), 10193489968026 (unit), Lot Numbers: 22FMG260, 22JMB517, 22JMF062, 23BME809, 23BMF391, 23BMG366, 23BMG367, 23BMI412, 23EMB428, 23FMA677, 23GMD648, 23JMC096, 23KMB846, 23LMC515; cc) REF DYNJ908375C, UDI/DI 40195327484263 (case), 10195327484262 (unit), Lot Numbers: 23IBL834, 23KBD095, 23LBB392; dd) REF DYNJS2027F, UDI/DI 40195327360314 (case), 10195327360313 (unit), Lot Numbers: 23EDA949, 23GDA203, 23HDB800, 23JDB140, 23KDA685, 23KDC088; ee) REF DYNJ907158A, UDI/DI 40183489970402 (case), 10195327137038 (unit), Lot Numbers: 22EME052, 22FMG234, 22IMC193, 22IMH124, 22LME781, 23CMD417, 23CME132, 23CMF812, 23DMF106, 23GMB888, 23HMH494, 23IME266, 23JMA733, 23KMH135; ff) REF CDS983991C, UDI/DI 40889942131410 (case), 10889942131419 (unit), Lot Numbers: 22FBB136, 22FBU282, 22GBN805, 22HBG300, 22HBS717, 22IBI043, 23CBT616, 23EBA215, 23EBU789, 23HBI490; gg) REF DYNJ902367J, UDI/DI 40195327505838 (case), 10195327505837 (unit), Lot Numbers: 23JBG975, 23KBN775, 23KBP709; hh) REF DYNJ905852C, UDI/DI 40195327214204 (case), 10195327214203 (unit), Lot Numbers: 22IBK464, 22IBM480, 23ABO113, 23ABO114, 23ABQ791, 23DBP733, 23EBR661, 23GBO389, 23HBX998, 23KBG164, 23LBF311; ii) REF DYNJ905514D, UDI/DI 40193489968034 (case), 10193489968033 (unit), Lot Numbers: 22EMA529, 22EMH888, 22FMB880, 22GME946, 22GMH574, 22GMI092, 22HMH887, 22LMD324, 22LMI453, 23AMG866, 23IMG780
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
886-359-1704
Manufacturer Reason
for Recall
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.
FDA Determined
Cause 2
Under Investigation by firm
ActionMedline Industries issued a Safety Alert to its consignees on 03/22/2024 via email and certified mail. The notice explained the problem with the device and requested that consignees follow the instructions for use. Distributors or those who further transferred the affected product were directed to notify their consignees.
Quantity in Commerce14977 units
DistributionWorldwide - US Nationwide and the countries of Panama, Mexico, Canada, Qatar, El Salvador, Oman, Colombia, Singapore.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-