| | Class 2 Device Recall Medline |  |
| Date Initiated by Firm | April 08, 2024 |
| Date Posted | August 28, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2741-2024 |
| Recall Event ID |
94581 |
| Product Classification |
Spinal epidural anesthesia kit - Product Code OFT
|
| Product | Medline procedural kits labeled as:
PREOP NERVE BLOCK TRAY, Pack Number DYNJRA2040H |
| Code Information |
DYNJRA2040H, Lot Number 23CBH893;
DYNJRA2040H, Lot Number 23HBX547;
DYNJRA2040H, Lot Number 23JBN538;
DYNJRA2040H, Lot Number 24ABQ696;
DYNJRA2040H, Lot Number 24CBG746 |
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Karin Johnson 886-359-1704 |
Manufacturer Reason for Recall | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods. |
| Distribution | Worldwide |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|