 Class 2 Device Recall SIGNA Explorersee related information
Date Initiated by FirmJune 14, 2024
Date PostedAugust 02, 2024
Recall Status1 Open3, Classified
Recall NumberZ-2473-2024
Recall Event ID 94874
510(K)NumberK143251 
Product Classification System, nuclear magnetic resonance imaging - Product Code LNH
ProductSIGNA Explorer, with affected software versions: SV25.5, SV25.6, SV29.2, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
Code Information System ID: K2308, SHIMR02644, 415723SHMR6, 951682MR, 619291MR1, 2533, 303355MR, 305682MR1, 727869MR, 352344MR2, 407PIMRKCH, 630332MR, 316689MR, 508746JMR2, 406863WMR, 445MR2, 718567MR, 718250MR, 716839MR1, MSF094-1-6UQ9H2TJ, 412784MR1, 830643MMR2, 540776MR2, 360582450WPA, MSF094-1-6VI2CR9B, 630843OSCMR, AL4093MR01, DZ1638MR02, DZ2685MR01, DZ1930MR01, DZ4483MR01, DZ2684MR01, DZ2183MR01, DZ3053MR01, DZ1947MR01, DZ1305MR01, DZ1863MR01, DZ1035MR03, DZ1004MR03, DZ1004MR02, DZ4258MR01, DZ1601MR01, DZ1146MR01, DZ1504MR01, DZ1269MR01, DZ4165MR01, DZ3484MR01, DZ1339MR04, DZ1070MR01, DZ4047MR01, DZ2417MR01, DZ2418MR01, DZ3741MR01, DZ4609MR01, DZ2090MR01, DZ2785MR01, DZ2552MR01, DZ3365MR01, DZ4149MR01, DZ3445MR01, DZ1149MR02, DZ1337MR01, DZ2622MR01, DZ1268MR01, DZ3165MR01, DZ4313MR01, DZ1406MR03, DZ1771MR01, DZ1406MR02, DZ2092MR01, DZ1027MR01, DZ2014MR01, DZ1060MR05, DZ1297MR01, DZ2764MR01, DZ4227MR01, DZ2773MR01, DZ3805MR01, DZ2351MR01, DZ1475MR01, DZ1908MR01, DZ2361MR01, DZ1856MR02, DZ4929MR01, DZ3953MR01, DZ3078MR01, DZ2995MR01, DZ1255MR01, DZ4380MR01, DZ3087MR01, DZ1466MR01, MR3766AR, MR1946AR, MR9321AR, MR3338AR, 2722SIGNA15, MR3867AR, MR2812AR, MR6472AR, MR1196AR, MR8785AR, MR4330AR, MR8126AR, MR3498AR, MR8882AR, MR5700AR, MR6883AR, MR6459AR, MR7206AR, MR6816AR, MR8830AR, 4365SIGNA15, MR6899AR, MR9700AR, 264EXCHDE, MR3403AR, MR8229AR, MR2932AR, MR5239AR, MR3718AR, MR5150AR, MR7906AR, MR6855AR, MR8433AR, MR8920AR, MRR9946, MR9679AR, MR1936AR, MR2346AR, 4127SIGNA15, MR5043AR, MR8907AR, 3722SIG15T, MR7077AR, MR8570AR, MR2042AR, MR1437AR, MR1245AR, MR7856AR, MR2433AR, MR1720AR, MR5444AR, MR5790AR, 0910272080, 0910272089, 0910272087, 83151656091, 83193107602, 83147183761, 83207359389, 833MR000001, 83140192745A, 83215453018, 83151589682, 83200805122, 83164742085, 83212927900, 083027606532922, 083027609288917, 083027605498616, 083027608479920, 083027600426022, 083027606875420, 083027607691119, 083027604109319, 083027601632818, 083027600426222, 083027600874617, 083027606911319, 083027600486516, 083027606521417, 083027601703117, 083027607417021, MR52743961, 210110MR01, 083027608604518, MRRD0622, MR1726BO, MR7065BO, 2560124, BA4147MR01, MR51360451, MRR9040, MRR7716, 2849423, MRR10602, 1175978, MRRD0670, MRR9917, MRR8311, MRRD0764, MRR10597, MRR11632, MRR11499, MRRD0889, MRRD0859, MRR7919, MRRD1291, MRRD1157, MRRD0392, 2651970, MRR11526, MRRD1085, MRRD0909, MRR10914, 2705171, MRR11466, MRR11251, MRRD0632, MRRD0568, MRRD0436, MRR11576, MRR4548A, MRR11178, MRR10328, MRR10064, 1337081, 2839886, 74954, 2626380, 2761824, MRRD1232, 1347499, MRRD0634, MRR11488, MRR11621, MRR11456, MRRD0466, MRR10927, MRRD0412, MRR11001, 2599138, MRR11619, MRRD1004, MRRD0679, MRR11293, MRRD0437, 1192292, MRR11186, MRR11404, MRRD0673, MRRD0352, MRRD0314, MRRD1180, MRRD1159, MRR10160, MRRD0401, MRRD1400, MRRD0393, 1200823, 60357, MRRD1254, 34574, MRRD0538, 2707330, MRRD0927, MRR11566, MRR8015, MRRD0595, MRRD0470, MRR8517, 2682131, MRR11494, MRR11297, MRR11108, MRRD0924, MRR10744, MRR9306, MRRD0616, MRR11626, MRR11375, MRR1183, MRR11031, MRR11087, MRR10855, MRR1560SX2, MRRD1324, MR1543SX8, MRR11547, MRR10840, MRR11391, MRR10939, MRRD1002, MRRD1122, MRRD0813, MRR11657, MRRD0988, MRR11177, 33262, MRR9279, MRRD0579, MRR11646, MRR11618, MRR11636, MRR11579, MRR11577, MRR11562, MRR11536, MRR11088, MRR10899, MRR10997, MRR11000, 1292548, 2705318, 1202470, 2613126, 1167491, 1196362, 1200954, 1184373, MR1355680, MRR11187, MRR8992, 2807598, MRR11614, 2831309, MRR10084, MRR8562, MRRD0958, MRR11647, MRRD0865, MRRD1383, MRRD0877, MRR11630, MRR3140, MRR6664, MRR8267, MRRD1081, 2713328, MRR11610, MRR10845, 2728519, MRR11653, MRRD0419, 2776720, 2668464, MR255906MR0, 1238483, MRRD0905, MRRD1181, MRRD0914, MRR11372, MRRD1020, MRR10964, MRR10813, MRYR2150, MRR10826, MRRD0952, MRRD1050, MRR8789, MRR11471, 2742381, MRR9929, 2713415, MRR8057A, MRR10274, 1344119, MRR11223, MRR11081, 2535122, MRR11038, MRR8310, 2708848, MRRD1037, 4737487MR11, MRR11640, MRR11555, MRR11382, MRR11358, MRR11188.MRR10995, MRR8405, MRR8460, 2831204, 26290-1, 2730904, 1202525, 1202522, 1202515, MRR8605, 2626827, MRRD0480, 47924-1, MRRD0828, MRRD0411, BR93131XXX, BG4600MR01, BG4612MR01, MR51201071, MR52096141, 250782MR, MRR6466, MRR1120, MRR8034, MRR1946, CUCMR1, MR4038CL, MRRD0718, MR2527CL, MRR8837, MRR2854, MRR10957, MR1914CL, MRRD0680, MRR4116, MRR1134, MR5061CL, MEDSMR1, MRR8531, MRRD0997, MR9406CL, MR8714CL, HBLUCOMR01, MRR5840, MR7894CL, MRR2425, 082427040285, 082427040212, 082427040225, 082427040216, 082427040203, 082427220123, 082427220082, 082427220064, 082427130173, 082427130093, 082427100312, 082427060090, 082427270068, 082427160302, 082427160292, 082427040223, 082427040237, 082427040231, 082427230156, 082427090448, 082427090400, 082427200216, 082427010155, 082427090373, 082427090380, 082427100288, 082427090337, 082427160241, 082427090323, 082427190147, 082427160375, 082427230154, 082427120225, 082427080107, 082427080078, 082427250080, 082427140145, 082427030173, 082427030163, 082427160227, 082427090362, 082427090358, 082427100296, 082427070233, 082427070182, 082427250082, 082427140186, 082427280019, 082427090384, 082427240095, 082427220077, 082427220071, 082427030180, 082427190150, 082427100407, 082427090403, 082427290030, 082427160235, 082427160319, 082427160310, 082427160318, 082427160256, 082427160362, 082427160238, 082427240074, 082427070215, 082427010134, 082427010129, 082427090401, 082427090334, 082427090305, 082427090328, 082427090434, 082427090419, 082427090436, 082427090321, 082427090393, 082427090325, 082427090303, 082427090295, 082427210191, 082427210198, 082427190139, 082427120214, 082427120290, 082427110140, 082427200151, 082427200238, 082427200201, 082427020097, 082427020111, 082427060119, 082427060095, 082427190143, 082427160242, 082427100347, 082427100305, 082427100396, 082427100307, 082427100294, 082427100338, 082427130117, 082427080071, 082427310174, 082427230148, 082427230262, 082427230253, 082427230245, 082427270071, 082427270076, 082427250083, 082427250081, 082427250065, 082427250122, 082427250096, 082427250094, 082427250079, 082427250072, 082427140183, 082427140180, 082427140157, 082427160225, 082427250084, 082427060092, 082427060088, 082427040272, 082427040254, 082427040238, 082427230121, 082427230145, 082427300036, 082427100397, 082427100418, 082427100405, 082427080129, 082427160252, 082427160343, 082427160197, 082427160307, 082427160250, 082427270061, 082427120369, 082427120308, 082427120331, 082427120333, 082427120334, 082427120345, 082427120330, 082427120191, 082427120326, 082427200147, 082427070189, 082427070277, 082427070202, 082427070318, 082427090433, 082427090296, 082427140178, 082427140179, 082427140188, 082427140187, 082427290041, 082427290064, 082427240080, 082427240105, 082427240121, 082427030138, 082427030212, 082427220084, 082427220126, 082427220081, 082427220105, 082427310129, MRRD0996, MRR97761, MRR9674, MRR9230, MRRD0698, MRRD0619, MR5368CO, MR3630CO, MR4479CO, MRR10843, MRR9801, 2839703GOLD, 2672732SIGNA, 2697548EXCITE, 2670275HORIZON, 2624537EXCITE, 2605060SIGEXT, 2586967SIGNA, 2567464SIGNAMR, MRR9706, 2812553GOLD, MRRD1139, MRR10955, MRR10892, MRR9513, 2887023XT, 2874631HDE1, MRRD1317, MR4574CO, MRR10359, 2806308HD, MRR10144, 2778960SIGNA, MRRD1323A, 2824773LX, MR5440CO, MRR10073, MRRD1434, MR2048CO, 2566917SIG15T, MRR10762, MR1205CO, MR8006CO, MRRD1175, MR8073CO, MR2257CO, MRRD0572, 2883683SIGNA, CI1040MR01, CI1089MR01, HR4029MR01, DK1083MR05, DK1083MR03, MRRD1444, 2726108MRHD, MRRD1356, MRRD1260, MR2582DO, MR5218EC, MRR9885, 2932192EX2, EG2782MR01, EG2761MR01, EG1790MR02, EG2763MR01, EG2629MR01, EG2661MR01, EG1238MR02, EG2604MR01, EG1709MR01, EG1526MR04, EG1237MR01, EG1118MR01, EG2717MR01, EG2558MR02, EG2572MR01, EG1037MR03, EG1284MR02, EG2754MR01, EG3013MR01, EG2940MR01, EG2948MR01, EG5723US04, EG1005MR03, EG1246MR01, EG1053MR02, EG2713MR01, EG2595MR01, EG1543MR01, EG1121MR01, EG1505MR02, EG1522MR02, EG1998MR01, EG1054MR03, EG2579MR01, EG1826MR02, EG2557MR02, EG2650MR02, EG2570MR01, EG1929MR02, EG1277MR01, EG1196MR01, EG2641MR01, EG1245MR01, EG2920MR01, EG1525MR04, EG1126MR01, EG2601MR01, EG1008MR03, EG1042MR01, EG3010MR01, EG1783MR02, EG1990MR01, EG1279MR01, EG2581MR01, EG1743MR01, EG2266MR01, EG2716MR01, EG2556MR02, EG2633MR01, EG2663MR01, EG2582MR01, EG1112MR02, EG2594MR01, EG1043MR02, EG1804MR01, EG2569MR01, EG1744MR01, EE4026MR01, MR51028551, MR51796811, MR51028131, ET1021MR02, ET1021MR01, ET1296MR01, MR50061091, FI1502MR01, FI1345MR01, FI1064MR01, FI1041MR01, FI1021MR02, X121122503, M4161670, M2045019, X171897213, M359673901, M2321310, X706835017, M196514411, X669977018, M4400430, M2485631, M2271518, GF1007MR01, 7361387, X141549007, M189104208, X2881410, GON5412312, M6063014, M347028307, M2748747, X151054614, 7461157, X178185401, L2202007, MR54068981, MQ1038MR04, MQ1038MR03, 6529149, X306788201, M1265008, M1265006, X472561001, M2618005, X756087033, M4183856, A128275409, X128275412, X185801209, M24313809, M222271601, 8264764, M6414733, M6414732, A1279217, M9886878, X116370513, 9826031, M2269855, 6968494, M2844564, M2533237, M1245737, M4503514, X288892404, M4152262, 08830630, 08486357, M200598212, X86070201, M414406701, 6098456, A1588543, X216471209, 8224770, X304369405, X304369401, X225779101, M190091214, RE1045MR02, 8526228, X970562025, 6242121, X2266305, X226587717, M477182001, GP1022MR02, 07143820, M4204524, 7141789, 07816651, X751968037, M2848451, M245413303, X533536502, B75162215, 7210650, M4216043, M6015639, M4497429, HAMB02MR01, 389077MR01, PA0148MR01, 396399MR02, 712330MR01, PC0075MR03, PX4178MR02, PX4178MR01, PC6132MR03, PC1351MR01, 393115MR01, 233637MR01, 160234MR01, PC1401MR01, 561545MR01, PC0702MR01, PNUE07MR01, 543871MR01, 560776MR02, 560776MR01, PC0573MR02, PC2007MR01, CO3030MR02, PC8740MR01, 519759MR01, 527556MR02, 505560MR01, TP0010MR01, PC1455MR01, PC3885MR01, HC5284MR03, PC0396MR10, 123702MR03, PC0396MR04, 505168MR01, 410725MR01, CC5340MR02, 265361MR03, PC5995MR01, PC0815MR03, PC4537MR02, PNUE06MR01, HC1711MR01, 420127MR02, PNUE04MR01, PC0125MR03, 126548MR01, PC4036MR01, 631485MR01, 487692MR01, HC9027MR02, PC0345MR02, 360937MR01, HC0628MR01, 220112MR01, PC3847MR01, TP0011MR01, 124454MR01, TP0008MR02, TP0009MR01, 145599MR01, 505562MR01, 468993MR01, PC3827MR03, HC6474MR02, H44436MR04, H41222MR02, H7024MR06, H1339MR03, H55722MR02, H17659MR1, H4562MR01, H42261MR01, H3044MR01, H17660MR3, H44432MR01, H6933MR01, H1787MR11, H1892MR12, H17666MR1, H44435MR03, H10634MR08, H4352MR05, H41223MR02, H4676MR03, H10651MR06, H4767MR02, H4079MR03, H3943MR01, H10635MR09, H17646MR04, H17655MR01, H7247MR01, H4262MR05, H2126MR16, H1926MR05, H1141MR24, H5531MR01, MRRD0868, MRR10967, 24527633818MRI, 24501195MR-1, 0828270079, 0828270077, 0828270069, 0828270048IB01, 0828270030IB01, 0828270050, 0828270049, 0828270045, 0828270020IB02, HU1749MR01, HU1243MR02, HU1099MR02, HU1842MR01, HU1583MR01, HU1897MR01, HU1672MR01, HU1744MR01, HU1641MR01, HU1835MR01, HU1831MR01, HR4269MR01, HU1886MR01, 083027846253819, 083027841277017, R83027844086323, 083027842020719, 083027846984620, 083027848070318, 083027846535119, R83027843424622, 083027840081018, 083027848396318, 083027849417418, 083027840855323, 083027848785516, 083027600909023, 083027603366516, 083027607457016, 083027104892617, R83027103776223, 083027100019412, 083027103536122, 083027102603721, 083027108786816, 083027102101447, 083027100049812, 083027109470819, 083027066508118, 083027820010312, 083027000949218, 083027006503017, 083027003155616, 083027009317018, 083027006678722, 083027005494719, 083027606048519, 083027001571315, 083027008063017, 083027242173520, 083027243088818, 083027246179917, 083027243411217, 083027245187515, 083027241541616, 083027247170216, R83027243887423, 083027242389615, 083027240009214, 083027061463720, 083027066503518, R83027064146723, 083027806275019, 083027006661822, R83027104625724, R83027824110923, 083027822898921, 083027826697920, 083027827198515, 083027820038511, 083027826405518, 083027820801321, 083027826430817, 083027824626519, 083027821740713, 083027826225520, 083027828631016, 083027821188412, 083027827461917, 083027823263019, 083027826405218, 083027863759923, 083027865719321, R83027864142823, 083027861954815, 083027864867814, 083027860007112, 083027865635520, 083027863759523, 083027865910620, 083027860008512, 083027860044912, 083027865032919, 083027866984117, 083027869960115, 083027843759323, 083027863735319, 083027280565115, 083027280009412, 083027289046323, 083027281989320, 083027282293320, 083027200753416, 083027204687420, 083027209120120, 083027203265719, 083027201689119, 083027203007317, 083027205547117, 083027200144616, 083027207170016, 083027207969111, 083027209389215, 083027821421617, 083027201984919, 083027212212422, 083027206642617, 083027209313010, 083027210326718, 083027001828115, 083027806086120, 083027600007311, 083027603635816, 083027602405020, R83027604013623, 083027606189211, 083027601581615, 083027601077320, 083027601078520, R83027603868323, R83027804089223, R83027064161124, 083027067560117, 083027060948417, 083027065592518, 083027042434320, 083027041266818, 083027041173416, 083027042722321, 083027043573918, 083027040391217, 083027045280015, 083027046625721, 083027601804818, 083027801967722, 083027804647721, 083027800993818, 083027803803219, 083027801497116, 083027803032115, 083027800439717, 083027804757116, 083027801749622, 083027800853915, 083027803478322, 083027802614121, R83027803164322, 083027842855011, R83037844277224, 083027844318723, R83027843920923, 083027843218922, 083027840149222, 083027844800321, 083027841233721, 083027840049116, 083027848271716, 083027842901111, 083027842583917, R83027844030823, 083027841239321, 083027842682716, 083027603806711, 083027026468322, 083027026592420, 083027022263118, 083027022598220, 8.3027E+13, 083027027146319, 083027029298718, 083027021879815, 083027022546321, 083027000004112, 083027021970520, 083027020068418, 083027024122521, 083027022581021, 083027004267914, 083027606787616, 083027600865518, 083027601425919, 083027601746512, 083027605960919, R83027603617222, 083027609619023, 083027602133220, 083027600908419, 083027609505617, 083027607006215, 083027600019013, 083027607456611, 083027601012510, 083027600030510, 083027602101609, 0835270035, 5512MR0024, 5512MR0008, 5512MR0013, 7982MR0032, 5512MR0007, 0835270018, 0835270006, 0835270004, 0835270016, 7982MR0012, 7982MR0019, 5512MR0015, 7982MR0055, 5512MR0019, 5512MR0017, 0835270005, 7982MR0001, 7982MR0002, 7982MR0021, 7982MR0011, 7982MR0041, 7982MR0048, IR2376MR01, IR2262MR01, MR51928461, MR51928451, IR2513MR01, IR2050MR01, IR2103MR01, IR1682MR01, IR1471MR02, MR51928441, MR48982456, IR2344MR01, IR2510MR01, IR2517MR01, IR2312MR01, IR2567MR01, IQ1540MR01, IQ1666MR01, IE1239MR01, IE1253MR02, IE1222MR01, IE1228MR02, A5156825, B197656404, A6184803, C5807417, A5388205, A344729904, C5294680, C5294618, B5152903, A6470604, A215175013, B215175025, A5331502, H5382005, A5413304, A129101307, A5290829, A5661503, A5183710, A5274202, A5697806, A5355209, A456623203, A5161411, A5437701, A5250211, A5236228, A5761301, A466145804, A5327026, A5644401, A5712303, A5359620, A6203402, A5218932, HXXXXX07, A5163336, A23086005, A5163316, A5719818, A5247924, A5276812, A5157407, A5738503, B5738501, A5128625, A5803203, A5300112, A5395312, A103421315, A5690009, A5194602, C5410301, A5405213, A5421025, C5165831, A5203406, A5155135, A389746501, A377686701, A5728839, A5728804, A5422026, A5185111, A5811749, C5126838, A5401312, A5102656, MR00097, MR00148, PC0816, MR00206, PC0807, YM3065, PC6184, PC6273, MR00212, PC6060, PC6011, PC6261, PC6142, PC0934, PC0856, PC6300, PC0921, PC0906, TEST20233, YM1476, PC6178, YM1476_old, PC0631, PC0821, PC0866, PC0830, PC6239, MR00113, PC6163, PC6216, PC0813, PC6289, MR00222, MR00142, PC6145, PC0847, PC6264, PC0210, PC0070, PC0876, MR00240, MR00122, PC6128, PC6195, YM0834, PC0840, PC6181, MR00036, PC0836, PC0006, PC0058, PC6155, PC0929, MR00153, PC6197, YM1745, MR00205, PC6202, PC0854, MR00253, MR00063, PC0040, PC5077, PC6282, MR00149, PC0846, PC6277, PC6149, YM1515, PC0072, MR00095, PC0858, PC0909, PC0201, PC0817, MR00235, PC6215, PC6150, PC6266, PC0859, PC0839, PC0805, PC6283, PC6191, PC0850, MR00014, MR00106, PC6122, PC6223, PC6231, MR00233, PC6212, PC0863, PC0809, MR00023, MR00246, PC6190, PC6159, PC6192, PC0864, PC0926, PC5009, MR00173, YM1768, PC6127, PC0956, MR00177, PC0837, PC0802, MR00209, PC0914, PC0868, PC0918, PC6187, MR00128, YM3057, MR00092, PC6229, PC6226, PC6165, PC0917, PC6227, PC6188, PC6164, PC0851, PC6189, PC6147, PC0904, PC6179, YM1892, PC6292, PC0870, MR00220, PC6180, PC6296, PC0810, PC5041, PC6055, PC0922, PC6221, MR00017, MR00044, MR00068, YM0651, PC0860, PC6123, PC6249, PC0827, PC0913, MR00031, PC0912, MR00215, PC6236, PC0835, PC0888, MR00107, PC6162, PC6153, MR00011, PC0612, PC0218, PC0826, PC6220, PC6204, PC0005, PC6120, MR00035, PC5075, PC6021, PC6177, PC5020, PC6301, MR00171, PC0928, PC0844, PC0100, YM0847, PC6242, PC6244, PC6232, PC0901, PC6272, PC0853, MR00178, PC0932, PC0903, PC6172, PC0845, MR00241, PC6183, PC0833, PC6211, PC0855, PC6200, PC0829, PC0806, MR00131, PC0872, PC6243, PC6295, MR00189, PC6135, PC6136, PC0838, PC0849, MR00003, PC0930, PC6250, PC6161, PC6203, MR00059, YM1456, PC0803, PC6207, PC0814, MR00030, PC6237, PC0822, MR00042, PC6235, PC6279, PC0033, PC0920, PC6255, PC0815, PC6157, PC0832, PC0056, PC6210, PC0905, PC5065, YM3078, YM0803, YM0780, PC6156, PC6152, PC0911, PC0931, PC0910, TEST20231, PC6214, PC0875, PC6201, PC0936, PC0862, PC0907, PC0823, PC6030, PC0828, PC0933, MR00234, PC0811, PC0924, MR00047, PC6206, PC0908, PC0848, PC0812, PC0915, PC6158, PC6217, PC0865, MR00108, MR00062, PC6144, PC6252, PC0867, PC6251, PC6175, PC6133, YM4111, PC0937, MR00048, PC0045, YM0534, MR00181, MR00239, PC0801, PC0037, PC0843, PC0916, PC5012, PC6119, PC6050, MR00126, PC0804, YM0810, YM3091, YM3092, YM0919, PC0919, MR00160, YM1450, PC0935, MR00127, MR00098, PC5058, YM3817, PC0055, MR00133, PC6194, PC6143, PC0871, PC6130, PC0927, PC6285, MR00144, MR00053, MR00052, PC6294, MR00054, PC6213, PC0857, MR00016, MR00214, PC0852, PC0842, PC0808, PC0925, PC0819, PC6225, YM1705, PC0869, PC0820, YM4160, PC6297, PC6240, PC0880, PC0825, PC5081, PC0013, YM0659, PC0099, PC6134, PC0861, YM0798, YM0662, PC6286, MR00213, MR00125, PC0841, MR00176, MR00034, PC0092, PC5070, PC5004, MR00018, PC0824, PC0081, JO1002MR04, JO1029MR02, JO4109MR02, KA4008MR01, *er*KA1029MR02, KA3109MR01, KA1029MR01, KA3300MR01, KA1040MR01, KA1035MR01, KA3443MR01, KA2801MR02, KA3260MR01, KA3466MR01, KG9999MR01, KA3377MR01, KA2737MR01, KA3416MR01, KA3378MR01, KA2974MR01, KA2824MR01, KA3355MR01, KA3406MR01, KE6239MR01, KE1000MR02, KE6184MR01, KE6323MR01, 0850270524, 0850270614, 0850270551, 0850270544, 0850270353, 0850270396, 0850270519, 0850270267, 0850270288, 0850270372, 0850270151, 0850270555, 0850270165, 0850270376, 0850270548, 0850270543, 0850270247, 0850270695, 0850270527, 0850270505, 0850270200, 0850270441, 0850270184, 0850270735, 0850270443, 0850270463, 0850270108, 0850270187, 0850270431, 0850270132, 0850270645, 0850270613, 0850270452, 0850270542, 0850270562, 0850270558, 0850270552, 0850270554, 0850270525, 0850270522, 0850270526, 0850270141, 0850270208, 0850270772, 0850270219, 0850270635, 0850270196, 0850270485, 0850270434, 0850270344, 0850270268, 0850270503, 0850270269, 0850270747, 0850270533, 0850270254, 0850270500, 0850270484, 0850270473, 0850270326, 0850270159, 0850270550, 0850270091, 0850270214, MR51359221, MR51333891, KW1013MR02, KW1001MR03, MR25692391, KW1006MR02, KW1005MR02, KW1019MR02, KW1018MR01, LV4026MR02, LV4004MR02, LB1009MR02, LB1064MR02, LB1430MR01, MR52673481, MR50004041, LY1160MR01, MR46599621, MR52666471, LY1011MR02, LY1013MR01, LY1175MR01, LY1177MR01, LY1011MR01, LT4114MR01, LT4022MR02, LT4028MR01, LT4069MR01, 0834270014, 834270018, 1553MR0009, 1553MR0004, 1553MR0016, 1553MR0020, 1553MR0012, ML1013MR01, MT1014MR01, BMALTA46, MRR1701A, MRRD0969, 96MRHD2563412, MRRD1363, MRRD1311, MRRD1283, MRHM0503, 2904666HDI, 5338MR15T, 1084MREINP, 570844TWINSPEED, MRRD1416.MRRD1389, 2673175LXMR, 57MRHD1178, MRRD0950, 2951520HDE, MRG2LS16226, MRRD1242, MRRD1453, MRRD1437, MRRD1449, MRRD1447, MRRD1213, 48MRE2744667, MRR8808, MRRD1233, MRG2YS13039, MRRD1370, MR292391MR0, MRR4342, MRR10970A, MRRD1287, MRRD1436, MRRD1411, MRRD1357, MRR2714, MRRD1076B, 31MRE2540752, MRRD1321, 2832672HDE, MRRD1410, MRRD1407, MRRD1380, MRRD1347, MRR8394, MRRD0536, MRR294, 2912069MRHDE, 2867718MRHDEAW, 2867718MRHDE, MRRD1335, MRRD1332, MRRD1279, MRR11558, MRR9546, 2856565HDE, MRR8240, MR287078MR0, 99201MR15, MRR0455, MR52353941, MR49554941, MR52293331, MR48582681, MA1488MR01, MR47289841, MA1151MR01, MR52300401, MR52299021, MR50497271, MR52293461, MR52298391, MA1955MR01, MA1955MR02, MA1588MR01, MA1087MR01, MA1737MR01, MA1252MR02, MA1278MR01, MA1233MR01, MA1650MR01, MA1152MR02, MA1465MR01, MA1502MR01, MA1047MR01, MA1727MR01, MA1083MR01, MR52293211, MA1479MR01, MA1460MR02, MA1881MR01, MA1339MR01, MR47780703, MA1296MR01, MA1864MR01, MR52298831, MA1938MR01, MA1263MR01, MA1219MR01, MA1769MR01, MA1020MR02, MA1711MR01, MR50554841, MA1818MR01, MA1197MR01, MA1003MR01, MA1941MR01, MZ1001MR01, 6073MR0017, 6073MR0040, 1573MR0004, 6073MR0016, 6073MR0024, 1573MR0005, 083027002514618, 083027003210019, ECT0220, RTD0123, NG1982MR01, NG1739MR01, NG1915MR01, NG1432MR01, NG1410MR01, NG1743MR02, NG1265MR01, 9447742, NG1359MR01, NG1356MR01, NG1043MR02, NG1256MR01, NO1021MR02, NO1284MR01, MR51714511, OM1139MR01, OM1121MR01, OM1020MR01, PK1373MR01, PK1314MR01, PK1248MR01, PK1623MR01, PK1300MR03, PK1300MR02, PK1242MR01, PK1628MR01, PK1453MR01, PK1391MR01, PK1629MR01, PK1627MR01, PK1390MR01, 2850446, MRR9466, MRRD0691, MRR10989, MRR9894, MRR8090, 38210315MRI, 2700602SIGNA, MR4122PE, MR5610PE, 1557MR0050, 1557MR0059, 1557MR0022, 1557MR0033, 1557MR0052, 1557MR0045, 1557MR0051, 1557MR0020, 1557MR0030, 1557MR0012, 1557MR0041, 1557MR0061, 0843270007, 1557MR0013, PL1797MR03, P00652MR01, P00666MR01, PL0677MR01, PL0620MR02, P00665MR01, PL2256MR03, PL2256MR01, PL1493MR01, P00020MR01, PL1754MR01, PL1745MR01, P00612MR01, PL1724MR01, PL3927MR01, PL1760MR01, PL0659MR01, P00360MR01, PL1772MR01, PL0439MR02, PL1016MR01, PL1709MR02, PL1720MR01, P00304MR01, P00224MR02, P00224MR01, PL0675MR04, P00027MR01, PL1813MR01, P00644MR01, P00403MR01, PL3547MR01, PL1501MR04, PL3596MR01, 600118MR02, PL2342MR01, PL3594MR01, PL1776MR01, PL1107MR02, PL0631MR01, PL1790MR01, PL1787MR01, P00477MR01, PL3889MR01, P00329MR01, PL1316MR01, PL2773MR02, PL3362MR01, PL1755MR01, PL0869MR01, P00529MR01, PL1060MR01, PL1730MR03, PL1355MR01, PL2309MR01, PL0001MR01, PPS75601, PPS54202, PPS26404, PPR02110, PER18426, PPR10411, PPR29115, PPR104800, PPS80401, PER01511, PPR59000, PPR62007, PPR94108, PPS49012, PPR92403, PPR23104, PER08728, PPS18202, PPS21302, PPR96413, PPR58104, PPS62102, 809269EXCITE, 787840EXCITE, 787263HDX, QA1095MR01, RO1138MR01, RO1430MR01, RO1232MR01, RO1494MR01, RO1016MR01, RO1463MR01, RO1462MR01, RO1345MR01, RO1321MR01, RO1149MR01, RO1080MR01, RO1150MR01, RO1200MR02, RO1200MR01, RO1227MR01, RO1198MR01, MR50839571, RO1329MR01, RO1572MR01, RO1172MR01, RO1678MR01, RO1098MR04, RO1428MR01, RO1038MR01, RO1646MR01, RO1210MR01, RO1015MR02, RO1102MR03, RO1595MR01, RO1584MR01, RO1355MR01, RO1090MR02, RO1664MR01, RO1255MR01, RO1640MR01, RO1356MR01, RO1438MR01, RO1358MR01, RO1473MR01, RO1122MR01, RO1602MR01, RO1146MR01, RO1429MR01, RO1325MR01, RO1160MR01, RO1622MR01, RO1362MR01, RO1363MR01, RO1453MR01, RO1231MR02, RO1272MR01, RO1364MR01, RO4172MR02, RO1303MR01, RO1471MR01, RO1466MR01, RO1365MR01, RO1205MR01, RO1092MR01, RO1128MR02, RO1598MR01, RO1673MR01, RO1051MR01, RO1087MR02, RO1465MR01, RO1257MR01, RO1252MR01, RO1644MR01, RO1674MR01, RO1487MR01, RO1308MR02, RO1593MR01, RO1229MR01, RO1159MR01, RO1059MR02, RO1431MR01, RO1060MR02, RO1562MR02, RO1460MR01, RO1336MR01, RO1508MR01, RO1371MR01, RO1042MR01, RO1373MR01, RO1125MR01, RO1152MR01, RO1194MR01, RO1675MR01, RO1550MR01, RO1374MR01, RO1577MR01, RO1544MR01, RO1412MR01, RO1416MR01, RO1144MR01, RO1432MR01, RO1184MR01, RO1455MR01, RO1133MR01, RO1153MR01, RO1447MR01, RO1139MR01, RF0129MR01, RU0076MR02, RF2748MR01, RF2631MR01, RF1815MR01, RU7536MR01, RF0367MR01, RF3042MR01, RU1078MR01, RU6340MR04, RF0567MR01, RU3923MR01, RU0473MR01, RU3889MR01, RF2072MR02, RU7618MR01, RU3926MR01, RF3270MR01, RU2580MR03, RU3611MR01, RU2652MR03, RF0001MR01, RU7524MR01, RU3408MR01, RU7550MR01, RF3564MR01, RU3447MR01, RF1647MR01, RU2157MR01, RU1115MR01, RF0252MR01, RF2584MR01, RF1352MR01, RF1332MR01, RU1198MR01, RU3623MR01, RU1086MR01, RF1695MR01, RU4440MR01, RF1258MR01, RF3584MR01, RF3081MR01, RF3082MR01, RF3317MR01, RU3867MR01, RU3261MR01, RU3291MR01, RU7872MR01, RU7180MR01, RF0852MR01, RU3462MR01, RU7545MR02, RU2447MR01, RU5552MR01, RU8891MR01, RU7564MR01, RF0484MR02, RU1017MR02, RU1017MR03, RU8916MR01, RF2406MR01, 600100MR04, RF1911MR01, RU4454MR01, RU1204MR04, RU1653MR02, RU3929MR01, RU3919MR01, RF1749MR01, RU0402MR01, RU2144MR01, RF3314MR01, RU7884MR01, RF3434MR01, RF2871MR01, RF1283MR01, RU7468MR01, RU6177MR01, RU3911MR01, SA2330MR01, SA2653MR01, SA2214MR01, SA2161MR01, SA2485MR01, SA2191MR01, SA1012MR05, SA1012MR04, SA1056MR01, MR53659451, MR53659481, MR53659491, MR53659501, RS4001MR01, RS4124MR01, RS4019MR01, RS4012MR01, RS4156MR02, RS4052MR01, RS4032MR02, RS4043MR01, 6073MR0012, 6073MR0009, 0847270018, 0847270029, SGMRIDUMMY123, 6073MR0026, 6073MR0025, 6073MR0022, 0847270001, SK1043MR01, ZA2103MR03, ZA1918MR03, ZA2224MR02, ZA1017MR01, ZA1038MR02, ZA1000MR04, 6225296, ZA3049MR01, ZA2015MR03, ZA1560MR01, ZA2372MR02, ZA2222MR02, 8282959, ZA2250MR01, ZA2119MR02, 09865187, ZA2101MR02, ZA1442MR02, ZA3192MR02, ZA3293MR01, ZA1058MR02, 09632508, ZA2100MR01, ZA1071MR02, ZA1006MR03, YP170A26, DPD59704, MPX43106, MPX35202, MP750816, MP825711, MP118827, ME208331, CE137808, NPV58123, FPG97802, DPN34709, FPD27500, NE209912, NPV63900, MPX98004, DEL26400, CP749713, MPX43803, FEC800352, DE641215, FE387840, FPD40200, DPD23001, DEL71943, MP385511, YE600109, DPN62302, DPA93505, DE492A30, MEX25550, FEG45915, NPV57900, NPV07308, YE167215, MP601318, FPG23603, DPN47008, ME051539, MEX71105, GPX25903, UE535A57, GPX22600, MP324987, MPX68102, DE669614, MEX71200, CPA91609, CEE15655, DPD37900, YP567303, NE741452, NE688850, CE671328, EE672844, FPG10102, FPG08204, MPD20500, DPN88433, MP3249117, NP083636, NP083635, 083027802635819, 083027800190520, 083027808117324, 083027800189520, SD1084MR01, MR52587891, MR52588111, MR51941681, MR50445311, SD1088MR01, SD1080MR01, SD1081MR01, SD1057MR01, MR51941821, SD1094MR01, SZ0002MR01, BD017MR01, O011MR02, O011MR01, N006MR01, K001MR01, K002MR01, Z001MR02, A013MR02, C018MR01, 824158MR01, 680542MR01, 501030MR02, 115030MR01, 167429MR01, 167682MR01, SY2074MR01, SY1012MR02, SY2077MR01, SY2078MR01, SY2013MR02, 0853270161, 0853270181, 0853270216, 0853270224, MR50718431, 1573MR0023, 1573MR0054, 1573MR0022, 1573MR0026, 0856270043, 1573MR0017, 1573MR0064, 1573MR0007, 0856270045, 0856270042, 0856270054, 1573MR0011, 0856270055, TN1588MR01, TN2004MR01, TN2004AW02, TN2002MR01, TN1278MR01, TN1632MR01, TN2005MR01, TN1699MR01, TN1018MR02, FSSADUMMMR01, TNDUMMMR01, TN1366MR01, 21028MRS01, 34717MRS01, 34717MRS02, 01071MRS01, 01062MRS01, 01005MRS02, 02008MRS01, 04011MRS01, 06236MRS01, 06211MRS01, 06184MRS07, 06184MRS02, 06184MRS03, 06185MRS01, 34787MRS01, 06172MRS01, 06167MRS01, 06072MRS01, 07023MRS02, 07023MRS01, 07040MRS01, 07039MRS01, 09019MRS01, 09018MRS01, 10022MRS01, 10021MRS01, 14008MRS01, 9435702, 16052MRS01, 16055MRS01, 16053MRS01, 16018MRS03, 16003MRS01, 17014MRS01, 20001MRS02, 25018MRS01, 25018MRS02, 25008MRS01, 26024MRS01, 26022MRS01, 31041MRS01, 31031MRS01, 31030MRS01, 31039MRS01, 34241MRS01, 54003MRS01, 34722MRS01, 34716MRS02, 34716MRS01, 34062MRS01, 34472MRS01, 34478MRS01, 34389MRS02, 34386MRS02, 34453MRS01, 34450MRS01, 34459MRS01, 34756MRS02, 34433MRS01, 34419MRS01, 34416MRS01, 35118MRS01, 35092MRS01, 35072MRS01, 35045MRS01, 46014MRS02, 46032MRS01, 37006MRS01, 38022MRS01, 79004MRS02, 71010MRS01, 71009MRS01, 71007MRS01, 40009MRS01, 41044MRS01, 34239MRS02, 41031MRS01, 43023MRS01, 43022MRS01, 00023MRS03, 44017MRS01, 45019MRS02, 45023MRS01, 45021MRS01, 45022MRS01, 47011MRS02, 33044MRS02, 33045MRS01, 49005MRS02, 52013MRS01, 53012MRS01, 55018MRS01, 55025MRS01, 55029MRS01, 55032MRS01, 63028MRS01, 60012MRS01, 31027MRS02, 61022MRS01, 61018MRS01, 65017MRS03, 65017MRS02, 66008MRS01, 67019MRS01, 68012MRS01, 06051MRS01, 10020MRS01, 10016MRS01, 19013MRS01, 20017MRS01, 21020MRS01, 28005MRS02, 8495586, 34308MRS01, 06182MRS01, 34455MRS01, 55038MRS01, 60001MRS01, MR49966381, UK1006MR01, MR51420931, UA4810MR01, MR53527441, UA4410MR01, MR53497281, AE1715MR01, AE1160MR01, 10718MRS01, 70001MRM01, 00055MRS03, 10740MRM01, IE1261MR01, 00298MRS01, IE1590MR01, 70383MRS03, 70383MRS04, 00136MRS02, 10893MRM01, 11670MRS01, 00577MRS02, 10498MRS01, 10724MRS01, 00824MRS01, 00313MRS03, 10695MRM04, 10079MRM83, 10719MRS01, 2647, MRRD1172, 2880725MR, MR48271631, UZ1359MR01, UZ1316MR01, UZ1357MR01, UZ1358MR01, MR52113731, UZ1352MR01, UZ1344MR01, UZ1360MR01, UZ1306MR01, 3882936037MRAND, 3882857623MRCTR, 3882647967MROCC, 38803561MRMET, 5973MR0010, 5973MR0066, 5973MR0049, 5973MR0043, 5973MR0035, 5973MR0073, 5973MR0045, 5973MR0039, 5973MR0041, 5973MR0029, 5973MR0020, 5973MR0036, 5973MR0019, 5973MR0098, 5973MR0062, 5973MR0034, ZW6019MR01
FEI Number 2183553
Recalling Firm/
Manufacturer
GE Medical Systems, LLC
3200 N Grandview Blvd
Waukesha WI 53188-1693
For Additional Information Contact
262-521-6560
Manufacturer Reason
for Recall
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
FDA Determined
Cause 2
Device Design
ActionGE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION to its consignees via letter on 06/14/2024. The letter explained the problem with the device, safety issue, and requested the following: "Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: To ensure that the displayed B1+RMS value on the Scan Prescription User Interface does not exceed the user prescribed limit specified in the Low SAR Mode User Interface: " If your software version includes the Optimized T2 FLAIR Sequence option, enable this option by setting the value to 1.00. This option is present on the Advanced Tab of the Scan Prescription User Interface. " If your software version does not include the Optimized T2 FLAIR Sequence option, adjust scan parameters such as TR , # Slices or Auto Refocus Flip Angle until the displayed B1+RMS value on the Scan Prescription User Interface does not exceed the user prescribed limit. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Quantity in Commerce2528 units
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report