| Class 2 Device Recall DxC 500 AU Clinical Chemistry Analyzer |  |
Date Initiated by Firm | July 10, 2024 |
Date Posted | August 21, 2024 |
Recall Status1 |
Open3, Classified |
Recall Number | Z-2691-2024 |
Recall Event ID |
95018 |
510(K)Number | K220977 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | DxC 500 AU Clinical Chemistry Analyzer, REF C63520
is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. |
Code Information |
Catalog Number: C63520
UDI-DI code: 14987666545065
Serial Numbers:
2023070010
2023060005
2023090015
2023070009
2023080011
2023080012
2023100019
2023100018
2023100020
2024010037
2024010036
2023110022
2024010038
2023090016
2024020040
2024020041
2024020039
2024020043
2024020042
2023110023
2024020045
2024020046
2024010024
2024010025 |
Recalling Firm/ Manufacturer |
Beckman Coulter Mishima K.K. 454-32 Higashino; Nagaizumi-Cho Sunto-Gun Japan
|
Manufacturer Reason for Recall | Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results. |
FDA Determined Cause 2 | Software design |
Action | during the week of 07/25/2024, the firm sent an "URGENT MEDICAL DEVICE RECALL" letter via email and postal mail to customers informing them that after the analyzer processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and an error code "9000" will be reported.
Customer are instructed:
. Monitor the rerun/reflex process if a rerun/reflex rack is held in the buffer area, manually retrieve the rack from the buffer area, manually order the rerun/reflex test and reload the sample onto the sample handler. Refer to Rerunning a Test section of the Instructions for Use on page 10-34.
Beckman Coulter has implemented the correction with software V1.4 which was released on 06/21/2024. A Beckman Coulter service representative will contact customers to schedule the software upgrade.
For any questions or assistance - contact the Beckman Coulter Customer Support via email https://www.beckmancoulter.com/en/support/contact-us |
Quantity in Commerce | 24 analyzers |
Distribution | Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|