• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmJune 27, 2024
Date PostedSeptember 11, 2024
Recall Status1 Open3, Classified
Recall NumberZ-3109-2024
Recall Event ID 95128
Product Classification Ear, nose, and throat surgical tray - Product Code OGR
ProductMedline convenience kits containing semi-rigid suction liners labeled as follows: a) BAPTIST FLOYD BRONCHOSCOPY, Pack Number DYKE1456D; b) ENT MINOR, Pack Number DYNJ908099B; c) HEAD & NECK, Pack Number DYNJ908101B; d) NASAL FESS, Pack Number DYNJ908105B; e) NASAL-LF, Pack Number DYNJ908429; f) MASTOPEXY-CANNOVA, Pack Number DYNJ909948
Code Information a) DYKE1456D: UDI/DI case 40195327639939, UDI/DI each 10195327639938, Lot Numbers: 24CBK289, 24BBK989; b) DYNJ908099B: UDI/DI case 40195327507085, UDI/DI each 10195327507084, Lot Numbers: 24DBI371, 24BBA188; c) DYNJ908101B: UDI/DI case 40195327507092, UDI/DI each 10195327507091, Lot Numbers: 24CBE408, 24BBA260, 24ABB112; d) DYNJ908105B: UDI/DI case 40195327507122, UDI/DI each 10195327507121, Lot Numbers: 24ABV362; e) DYNJ908429: UDI/DI case 40195327078820, UDI/DI each 10195327078829, Lot Numbers: 24CBB310, 23LBV007; f) DYNJ909948: UDI/DI case 40195327512911, UDI/DI each 10195327512910, Lot Numbers: 24DBD609
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
Manufacturer Reason
for Recall
Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.
FDA Determined
Cause 2
Process control
ActionMedline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 06/27/2024 via US mail and email. The notice explained the problem with the device, potential risk to the patient, and requested the consignee immediately check their stock for the affected item number and the affected lot numbers and destroy any affected inventory. Distributors were directed to notify their customers of the recall and request the affected products be destroyed. For questions: contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce407 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Panama.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-