| Date Initiated by Firm | September 04, 2024 |
| Date Posted | October 23, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0162-2025 |
| Recall Event ID |
95415 |
| Product Classification |
Catheter care tray - Product Code OHR
|
| Product | Medline CENTRAL CATHETER INSERTION TRA, REF CVI680C; medical procedure convenience kit |
| Code Information |
UDI/DI 40653160992944 (each), 10653160992943 (case) , Lot Numbers: 23LBO846
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield
IL
60093-2753
|
| For Additional Information Contact | Karin Johnson 866-359-1704 |
Manufacturer Reason for Recall | It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 09/05/2024 via email and USPS. The notice explained the issue, hazard, and requested the destruction of the affected product. Distributors were directed to notify their consignees. |
| Quantity in Commerce | 1180 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Panama. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|