 Class 1 Device Recall VOCSN MultiFunction Ventilators: VOCSNPro and VProsee related information
Date Initiated by FirmFebruary 03, 2025
Date PostedFebruary 26, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1196-2025
Recall Event ID 96136
510(K)NumberK162877 
Product Classification Ventilator, continuous, facility use - Product Code CBK
ProductVOCSN Multi-Function Ventilators: VOCSN+Pro (V+O+C+S+N+Pro, English), REF: PRT-00490-001; V+Pro (V+Pro, English), REF: PRT-01185-000, PRT-01185-002
Code Information REF/UDI-DI/Serial Numbers: PRT-00490-001/00855573007792/123627, 121892, 119639; PRT-01185-000/00855573007877/5038421, 116700, 117933, 5038112, 5038619, 5038717, 5038953, 5039086, 5039091, 5039361, 5039499, 5039503, 5039541, 5039636, 5039938, 5040019, 5040640, 5040762, 5041060, 5041078, 5041534, 5042558, 117344, 118464, 118509, 5039387; PRT-01185-002/00850018761154/5037655
FEI Number 3013095415
Recalling Firm/
Manufacturer
Ventec Life Systems, Inc.
22002 26th Ave Se
Bothell WA 98021-4903
For Additional Information Contact
425-686-1722
Manufacturer Reason
for Recall
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
FDA Determined
Cause 2
Component change control
ActionOn 02/03/2025, recall notices were sent to Durable Medical Equipment Distributors asking them to do the following: Clinical Action: Follow internal procedures to transfer the patient to an alternative source of ventilation, if available. If no other ventilation method is available, you may continue to use the affected product adhering to the guidelines below, while you await device correction by firm Service personnel: 1. Use only when the patient is actively monitored to ensure timely response to a potential malfunction. 2. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 3. Use external monitors (e.g. SpO2) and set the alarm thresholds appropriately for the patient. 4. Connect the ventilator to an external monitoring system, wherever possible. Also, Durable Medical Equipment Distributors are asked to acknowledge they forwarded the recall notice, and that they contacted the users to which the devices were provided, advised them of the recall and action instructions within accompanying Recall Notification Letter including how to have the devices repaired. Complete and return the acknowledgement form via email to Ventec-Recall@reacthealth.com For any questions contact the firm at at 844-MYVOCSN (844-698-6276), Monday through Friday between 8:00 a.m. - 5:00 p.m. PDT.
Quantity in Commerce30
DistributionUS distribution in CA, NY, and MO.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.