| | Class 2 Device Recall MiniMed Insulin Pump |  |
| Date Initiated by Firm | January 31, 2025 |
| Date Posted | March 03, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1258-2025 |
| Recall Event ID |
96200 |
| Product Classification |
Pump, infusion, insulin, to be used with invasive glucose sensor - Product Code OYC
|
| Product | MiniMed insulin pump, REF: 620G (MMT-1750) |
| Code Information |
All Lot Numbers; All UDI numbers |
| FEI Number |
3003166194
|
Recalling Firm/ Manufacturer |
Medtronic MiniMed, Inc.
18000 Devonshire St
Northridge
CA
91325-1219
|
| For Additional Information Contact | Medtronic 24-Hr Technical Support 1800-646-4633 Ext. 1 |
Manufacturer Reason for Recall | All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. |
FDA Determined Cause 2 | Device Design |
| Action | On January 31, 2025, Medtronic MiniMed issued a "Urgent Medical Device Correction" notification to all affected consignees. The notification was distributed via: e-mail, certified mail, SMS text message or phone call.
1. Monitor Your Glucose Levels:
2. Discuss how to prepare for situations like this with your healthcare professional.
3. Respond to Alerts and Symptoms.
4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter by either completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://info.medtronicdiabetes.com/pump-delivery-volume] to acknowledge this communication. |
| Quantity in Commerce | 228 units |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY and the countries of Algeria, Andorra, Argentina, Armenia, Aruba, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State Of Bosnia, Herzegovina , Botswana, Brazil, Brunei, Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Cayman, Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic, Republic Of Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Luxembourg, Malawi, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Republic Of Montenegro, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Northern Ireland, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion Romania, Russia, Saint Pierre And Miquelon, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Tanzania, United Republic Of Thailand, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic Of Vietnam, Yemen, Zimbabwe.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|