• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall WB XL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall WB XLsee related information
Date Initiated by FirmApril 17, 2025
Date PostedApril 24, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1671-2025
Recall Event ID 96726
510(K)NumberK241575 
Product Classification Abutment, implant, dental, endosseous - Product Code NHA
ProductWB XL Anatomic Healing Abutment XC. Endosseous Dental Implant Abutments.
Code Information Article No. 064.4522S and 064.4523S; UDI (01)07630031775856(11)240613(17)290612(10)GNNA2, (01)07630031775856(11)241115(17)291114(10)JZYZ3, (01)07630031775856(11)241216(17)291215(10)KRJA6, (01)07630031775863(11)240604(17)290603(10)GJXT4, (01)07630031775863(11)241011(17)291010(10)JLLP9, (01)07630031775863(11)241120(17)291119(10)KCKL1, (01)07630031775863(11)241211(17)291210(10)KPRE4, (01)07630031775863(11)250219(17)300218(10)LWZM6; Lot No. (Expiration Date): GNNA2 (6/12/29), JZYZ3 (11/14/29), KRJA6 (12/15/29), GJXT4 (6/3/29), JLLP9 (10/10/29), KCKL1 (11/19/29), KPRE4 (12/10/29), and LWZM6 (2/18/30).
Recalling Firm/
Manufacturer
Straumann USA LLC
60 Minuteman Rd
Andover MA 01810-1008
For Additional Information ContactStraumann USA LLC
800-448-8168
Manufacturer Reason
for Recall
Incorrect blister labelling.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn "Urgent Field Safety Action" notification dated 4/17/2025 was mailed to consignees. The notification instructs consignees to immediately check inventory for affected devices and to quarantine any units identified as in stock. Any recalled devices identified are to be returned to Straumann USA LLC to the attention of Jennifer Jackson for credit or replacement. For affected devices that have been implanted, consignees are being instructed to evaluate the compatibility based on the provided instructions. Consignees are asked to return the provided Customer Confirmation Form using the UPS label included in the notification.
Quantity in Commerce1,786 units
DistributionWorldwide distribution - US Nationwide and the countries of Belgium, Canada, China, Netherlands, and Sweden.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = NHA
-
-