• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ELEOS Limb Salvage System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ELEOS Limb Salvage Systemsee related information
Date Initiated by FirmSeptember 22, 2024
Date PostedMay 28, 2025
Recall Status1 Open3, Classified
Recall NumberZ-1857-2025
Recall Event ID 96780
510(K)NumberK161520 
Product Classification Prosthesis, knee, femorotibial, constrained, cemented, metal/polymer - Product Code KRO
ProductELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System
Code Information Part Number: 25001040E. UDI-DI B27825001040E0. Lot Numbers: 1989873, 1989874
FEI Number 3013450937
Recalling Firm/
Manufacturer
Onkos Surgical, Inc.
77 E Halsey Rd
Parsippany NJ 07054-3714
For Additional Information ContactSean Curry
201-650-7349
Manufacturer Reason
for Recall
Potential challenge with intraoperative assembly of the external taper of a single midsection with the internal taper of a mating proximal femur.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOnkos Medical notified consignees via email on about 09/22/2024. Consignees were instructed to segregate and quarantine any product on hand immediately, complete and return the response form, and arrange for existing implants to be returned.
Quantity in Commerce24 units
DistributionUS Nationwide distribution in the states of AZ, CA, FL, FL, MD, MO, NC, TN.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = KRO
-
-