| Date Initiated by Firm | September 04, 2025 |
| Date Posted | October 16, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0173-2026 |
| Recall Event ID |
97659 |
| Product Classification |
Labor and Delivery Kit - Product Code MLS
|
| Product | Medical convenience kits
Item number/ Product description
AHLD15X LABOR DELIVERY PACK - 213387
AKDV96B VAGINAL DELIVERY
AKDV96B VAGINAL DELIVERY
EVBH28D BIRTHING ROOM PACK
UPVD01K VAGINAL DELIVERY PACK |
| Code Information |
AHLD15X
UDI-DI 191072224767
Lot 8188711
Bag serial no.
17555039
17555044
17555049
17609166
17609181;
AKDV96B
UDI-DI 191072224019
Lot 7311311
Bag serial no.
47140875
47141919;
EVBH28D
UDI-DI 191072239051
Lot 8424611
Bag serial no.
17754395;
UPVD01K
UDI-DI 191072238993
Lot 8439111
Bag serial no.
17682598
Log 8499211
Bag serial no.
17742543
|
| FEI Number |
3007069947
|
Recalling Firm/ Manufacturer |
American Contract Systems Inc.
1601 Se Gateway Dr Ste 120
Grimes
IA
50111-5223
|
| For Additional Information Contact | Marlene Jones 470-280-4277 |
Manufacturer Reason for Recall | Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers.
Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users. |
| Quantity in Commerce | 32,433 total |
| Distribution | US Nationwide distribution in the states of SD, IA, MN, WA, IL.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|