| Date Initiated by Firm | September 04, 2025 |
| Date Posted | October 16, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0179-2026 |
| Recall Event ID |
97659 |
| Product Classification |
Gynecological laparoscopic kit - Product Code OHD
|
| Product | Medical convenience kits
Item Number/Description
RCGY52D GYN LAPAROSCOPY SPH 0246965
UIGN82M GYN (PS 907645)
UIHY16AE ASC HYSTEROSCOPY PACK (PS 020634) |
| Code Information |
FHGY09S
UDI-DI 191072234360
lot 8532011
bag serial number
68874879;
LMGY39U
UDI-DI 191072209337
Lot 8490211
bag serial numbers
47860009
47861175;
OWGY11Q
UDI-DI 191072236555
lot 8389711
bag serial number
17622382;
RCGY52D
UDI-DI 191072232670
lot 8142411
bag serial number
17487909;
UIGN82M
UDI-DI 191072106674
lot 8084811
bag serial numbers
17391081
17391091
17391111
17391171
lot 8445711
bag serial numbers
17683242
lot 8530511
bag serial numbers
17712404;
UIHY16AE
UDI-DI 00191072234117
Lot 8171811
N/A
* If bag serial number is listed as "n/a", then entire lot is affected |
| FEI Number |
3007069947
|
Recalling Firm/ Manufacturer |
American Contract Systems Inc.
1601 Se Gateway Dr Ste 120
Grimes
IA
50111-5223
|
| For Additional Information Contact | Marlene Jones 470-280-4277 |
Manufacturer Reason for Recall | Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers.
Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users. |
| Quantity in Commerce | 32,433 total |
| Distribution | US Nationwide distribution in the states of SD, IA, MN, WA, IL.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|