• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Livewire Steerable Electrophysiology Catheter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Livewire Steerable Electrophysiology Cathetersee related information
Date Initiated by FirmDecember 22, 2025
Date PostedFebruary 06, 2026
Recall Status1 Open3, Classified
Recall NumberZ-1321-2026
Recall Event ID 98277
510(K)NumberK221067 
Product Classification Catheter, recording, electrode, reprocessed - Product Code NLH
ProductReprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH;
Code Information 401582RH UDI-DI (case) 20888277407517¿ (ea) 10888277407510 Lots EP250618 EP250630; 401940RH UDI-DI 10888277407480 Lot EP250618; 401904RH UDI-DI (case) 20888277407609¿ (ea) 10888277407602 Lots EP250618 EP250630; 401914RH UDI-DI (case) 20888277407524¿ (ea) 10888277407527 Lot EP250618; 401932RH UDI-DI (case) 20888277407678¿ (ea) 10888277407671 Lots EP250618 EP250630; 401652RH UDI-DI 10888277407572 Lots EP250618 EP250630
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barlay
866-359-1704
Manufacturer Reason
for Recall
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
FDA Determined
Cause 2
Under Investigation by firm
ActionOn December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer.
Quantity in Commerce41
DistributionDistribution US nationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = NLH
-
-