|
Class 2 Device Recall |
 |
Date Initiated by Firm |
February 20, 2002 |
Date Posted |
September 12, 2003 |
Recall Status1 |
Terminated 3 on April 11, 2005 |
Recall Number |
Z-1204-03 |
Recall Event ID |
26581 |
510(K)Number |
K912470
|
Product Classification |
Table, Radiographic, Non-Tilting, Powered - Product Code IZZ
|
Product |
Multidiagnost 96 |
Code Information |
The firm utilizes a site number specific to each unit: 17197, 17368, 17325, 76449, 17247, 17339, 17304, 13444, 6616, 6558, 6559, 6426, 6539, 6541, 6583, 35366, 37859, 31000, 67642, 62363, 62362, 62278, 62300, 25769, 25901, 25884, 25768, 25900, 25888, 25834, 25967, 25710, 25907, 62279, 62310, 62403, 62391, 62309, 62404, 73921, 50135, 44742, 44670, 44703, 44760, 44699, 46921, 52354 |
Recalling Firm/ Manufacturer |
Philips Medical Systems 22100 Bothell Everett Highway Bothell WA 98041
|
For Additional Information Contact |
425-487-7000
|
Manufacturer Reason for Recall |
Spontaneous movement of table without input from an operator
|
FDA Determined Cause 2 |
Other |
Action |
On 2/20/02 the firm issued Field Change Order to their field personnel, instructing them to visit each site and upgrade each unit. |
Quantity in Commerce |
48 units |
Distribution |
The firm distributed units to hospitals and medical centers located throughout the US. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZZ and Original Applicant = PHILIPS MEDICAL SYSTEMS, INC.
|
|
|
|