Date Initiated by Firm |
June 19, 2003 |
Date Posted |
August 06, 2003 |
Recall Status1 |
Terminated 3 on December 01, 2004 |
Recall Number |
Z-1077-03 |
Recall Event ID |
26722 |
Product |
Synchron LX Systems Primary Tube Sample Template
(a piece of ''labeling'' for use with the Synchron LX 20 and LX 20 Pro systems).
Part Number: 967178 |
Code Information |
No range of serial numbers apply. |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc 200 South Kraemer Blvd W 337 Brea CA 92822
|
For Additional Information Contact |
Kathleen M. Jaker 714-961-3666
|
Manufacturer Reason for Recall |
Dead volume on sight gauge for the Synchron LX 20 and LX 20 Pro Systems not stated, which may result in instrument error.
|
FDA Determined Cause 2 |
Other |
Action |
A revised template with a recall letter was sent to customers with a response request on 19 June 2003. |
Quantity in Commerce |
502 |
Distribution |
USA and Canada. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|