Date Initiated by Firm |
October 24, 2003 |
Date Posted |
December 18, 2003 |
Recall Status1 |
Terminated 3 on February 09, 2005 |
Recall Number |
Z-0236-04 |
Recall Event ID |
27465 |
Product Classification |
Indicator, Biological Sterilization Process - Product Code FRC
|
Product |
3M Attest 1264 Biological Indicators, in boxes of 100 per box |
Code Information |
Lot numbers 2005-07 DR, 2005-07 DS, 2005-07 DT, 2005-07 DU, 2005-08 DA, 2005-08 DB, 2005-08 DC, 2005-08 DN, and 2005-08 DO |
Recalling Firm/ Manufacturer |
3m Health Care 3m Center Saint Paul MN 55144-1001
|
For Additional Information Contact |
Cynthia Lamarucciola 651-733-7605
|
Manufacturer Reason for Recall |
The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on a positive control may revert to negative after 24 hours of incubation.
|
FDA Determined Cause 2 |
Other |
Action |
The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product. |
Quantity in Commerce |
9592 boxes of 100 biological indicators per box |
Distribution |
The product was distributed nationwide in the United States and worldwide. |