| Class 2 Device Recall Carbomedics | |
Date Initiated by Firm | April 30, 2002 |
Date Posted | July 20, 2004 |
Recall Status1 |
Terminated 3 on March 24, 2004 |
Recall Number | Z-0738-04 |
Recall Event ID |
28335 |
510(K)Number | K960169 |
Product Classification |
Holder, Heart-Valve, Prosthesis - Product Code DTJ
|
Product | Carbomedics VT-200 Valve Tester |
Code Information |
All lots |
Recalling Firm/ Manufacturer |
Carbomedics, Inc 1300 E Anderson Ln Austin TX 78752-1708
|
Manufacturer Reason for Recall | Excessive cleaning and resterilization cause cracking and crazing of device causing the device to malfunction during use. |
FDA Determined Cause 2 | Other |
Action | Letters were sent to surgeons and hospital operating room supervisors in mid-May 2002 which included the following information:
It is imperative that hospitals regularly and thoroughly inspect all testing instruments for indications of wear and degredation. And, CarboMedics is in the process of re-issuing suggested maintenance and sterilization guidlines. |
Quantity in Commerce | 4606 units |
Distribution | US nationwide and to the following foreign countries: Denmark, Ireland, China, Cyprus, Greece, France, UK, Protugal, Ecuador, India, Japan, Hong Kong, Italy, Sweden, Belgium, Norway, Argentina, Spain, Malta, Austria, Switzerland, Netherlands, Germany, Australia, Finland |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = DTJ
|
|
|
|