| Date Initiated by Firm | January 29, 2004 |
| Date Posted | July 20, 2004 |
| Recall Status1 |
Terminated 3 on November 08, 2005 |
| Recall Number | Z-0624-04 |
| Recall Event ID |
28210 |
| 510(K)Number | K033134 |
| Product Classification |
Stents, Drains And Dilators For The Biliary Ducts - Product Code FGE
|
| Product | ParaMount Mini GPS Stent System, Peripheral System, Balloon Expandable Stent and Delivery System 7mm x 21mm 80 cm .014" REF PMP4-7-21-80 |
| Code Information |
Lots 262849 through 445216 |
| FEI Number |
2183870
|
Recalling Firm/ Manufacturer |
ev3, Inc. 651 Campus Dr Saint Paul MN 55112-3495
|
Manufacturer Reason for Recall | After stent deployment, resistance can occur when pulling the deflated balloon back into the guiding catheter and this can result in an increased procedure time. |
FDA Determined Cause 2 | Other |
| Action | A recall letter was sent by Fed Ex to all affected customers for delivery on 01/30/04. Follow-up and product retrieval was done by the firm''s representatives. |
| Quantity in Commerce | 446 |
| Distribution | The product was distributed nationwide in the United States and internationally to China and Ireland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = FGE
|