| Date Initiated by Firm | March 30, 2004 |
| Date Posted | July 20, 2004 |
| Recall Status1 |
Terminated 3 on July 24, 2006 |
| Recall Number | Z-0851-04 |
| Recall Event ID |
28753 |
| 510(K)Number | K760466 K801320 |
| Product Classification |
Screen, Intensifying, Radiographic - Product Code EAM
|
| Product | GE Single 15 inch TV Monitor Suspension for XT Bridge, Model 46-240485P2, Catalog Number B2056HA |
| Code Information |
All serial numbers |
Recalling Firm/ Manufacturer |
General Electric Med Systems 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | Larry A. Kroger 262-544-3894 |
Manufacturer Reason for Recall | The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall. |
FDA Determined Cause 2 | Other |
| Action | Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals. |
| Quantity in Commerce | 1236 devices |
| Distribution | The units were distributed nationwide in the United States and worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = EAM
|