• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall GE Advantx

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall GE Advantxsee related information
Date Initiated by FirmMarch 24, 2004
Date PostedJuly 20, 2004
Recall Status1 Terminated 3 on July 24, 2006
Recall NumberZ-0898-04
Recall Event ID 28747
510(K)NumberK890348 
Product Classification System, X-Ray, Angiographic - Product Code IZI
ProductAdvantx 1 LC2 Model 45562281 Angiographic X-ray System
Code Information All serial numbers
Recalling Firm/
Manufacturer
General Electric Med Systems
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information ContactLarry A. Kroger
262-544-3894
Manufacturer Reason
for Recall
Unintended movements of both the C-arc and the image intensifier have occured due to shearing of the gear motor shaft as a result of the misalignment of the gearbox.
FDA Determined
Cause 2
Other
ActionField modification instructions were issued on 3/24/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to the hospitals.
DistributionThe units were distributed nationwide in the United States and worldwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IZI
-
-