| | Class 2 Device Recall DatexOhmeda Aestiva/5 |  |
| Date Initiated by Firm | April 16, 2004 |
| Date Posted | July 20, 2004 |
| Recall Status1 |
Terminated 3 on November 14, 2005 |
| Recall Number | Z-0969-04 |
| Recall Event ID |
28936 |
| 510(K)Number | K993410 |
| Product Classification |
Apparatus, Gas-Scavenging - Product Code CBN
|
| Product | Datex-Ohmeda Aestiva/5 Active Gas Scavenging System, Part Number 1406-8216-000, used with a Model 7900 Ventillator, on a Aestiva/5 Anesthesia Machine, in a Magnetic Reasonance Suite. |
| Code Information |
Serial numbers AMTG00208, AMTG00212, AMTG00213, AMTG00216, AMTG00219, AMTG00220, AMTG00222, AMTH00113, AMTH00121, and AMTH00124 |
Recalling Firm/ Manufacturer |
Datex-Ohmeda Inc One Ohmeda Dr. Box 7550 Madison WI 53707
|
| For Additional Information Contact | Mr. Gary Muellenbach 800-345-2700 Ext. 3583 |
Manufacturer Reason for Recall | Increased circuit pressure as a result of a misassembled Aestiva/5 Active Gas Scavenging System. |
FDA Determined Cause 2 | Other |
| Action | Recall letters, dated April 26, 2004, stated that a firm representative would contact the consignee to schedule a visit to verify the proper function of each device unit. |
| Quantity in Commerce | 10 units |
| Distribution | The affected units were distributed nationwide in the United States and worldwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CBN
|
|
|
|