| Class 2 Device Recall AMATECH | |
Date Initiated by Firm | July 12, 2004 |
Date Posted | August 07, 2004 |
Recall Status1 |
Terminated 3 on May 05, 2005 |
Recall Number | Z-1327-04 |
Recall Event ID |
29589 |
Product Classification |
Table, Operating-Room, Ac-Powered - Product Code FQO
|
Product | Neuro Adapter 2 used with the Berchtold B-810 Surgical Table or Getinge USA Fortis Surgical Table
Catalog No: MDTNA2 |
Code Information |
131613, 131614, 131615, 131616,152144, 152145, 152147, 152148 219039, 219040, 219042, 219043, 232800, 232801, 232802, 232803, 232804, 205395, 205396, 235327, 235329, 235330, 237473, 237465, 237467, 237468, 237469, 237470, 237471 |
Recalling Firm/ Manufacturer |
Allen Medical Systems One Post Office Square Acton MA 01720-3934
|
For Additional Information Contact | Nancy Royalty 877-458-4884 |
Manufacturer Reason for Recall | Neuro adapter may fail to engage and adequately clamp to the surgical table, resulting in movement |
FDA Determined Cause 2 | Other |
Action | Allen Medical notified the two distributors by telephone on 7/12/04 and 7/16/04. A customer notifcation letter issued requesting return of the device for correction. |
Quantity in Commerce | 29 units |
Distribution | SC |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|