  Class 2 Device Recall see related information
Date Initiated by Firm September 10, 2004
Date Posted November 05, 2004
Recall Status1 Terminated 3 on February 09, 2006
Recall Number Z-0178-05
Recall Event ID 30160
510(K)Number K002185  
Product Classification System, Imaging, Pulsed Doppler, Ultrasonic - Product Code IYN
Product HDI 4000 Ultrasound System & the Philips 4000 Ultrasound System (same system).
Code Information Part # 9896 053 15031 (8500-0066-01), 9896 053 18811 (8500-0080-01), 9896 053 22011 (8500-0087-01)  Site numbers: A42000896, A3B000796, A34000603, A2C000441, A32000483, A33000540, A28000273, A35000637, A35000649, A26000159, A45001061, A28000241, A37000732, A25000143, A35000663, A42000932, A32000505, A42000898, A33000546, A24000084, A26000165, A28000255, A28000266, A28000276, A45001050, A36000692, A45001066, A34000559, A24000052, A32000501, A2C000460, A3C000853, A34000582, A2A000350, A2C000458, A2B000408, A28000277, A27000197, A25000136, A33000550, A34000596, A24000104, A34000621, A33000549, A2B000374, A34000607, A43001003, A28000238, A34000568, A28000275, A36000674, A29000323, A2A000354, A29000318, A29000325, A27000188, A36000678, A28000253, A34000628, A27000204, A3B000804, A36000687, A3C000825, A25000140, A3B000799, A28000267, A2A000335, A2A000353, A31000489, A29000289, A2A000338, A39000780, A26000153, A35000636, A2A000348, A38000760, A42000916, A2B000395, A42000912, A24000101, A36000706, A35000665, A34000629, A42000907, A43001005, A25000131, A27000189, A25000123, A42000889, A2C000438, A2B000420, A28000231, A26000160, A37000735, A32000502, A34000595, A24000067, A42000885, A28000270, A24000054, A2B000358, A39000776, A2C000448, A3C000846, A24000087, A35000659, A2C000455, A39000769, A43001006, A26000161, A24000057, A42000945, A34000589, A37000723, A31000477, A31000488, A2B000411, A2C000440, A2C000452, A2C000456, A31000466, A31000467, A31000479, A31000480, A31000481, A32000509, A26000154, A27000181, A42000925, A33000523, A34000577, A3C000812, A29000290, A31000474, A34000602, A26000157, A2A000349, A42000901, A42000947, A23000047, A36000702, A25000146, A2A000345, A44001022, A28000280, A2C000449, A3C000850, A41000861, A24000060, A2A000330, A42000895, A27000212, A26000170, A27000192, A42000931, A37000738, A26000173, A26000179, A26000178, A29000300, A38000753, A39000783, A24000061, A2A000327, A27000186, A2C000463, A31000478, A2C000446, A31000492, A28000235, A28000229, A25000144, A2B000399, A27000198, A27000202, A2C000445, A34000588, A36000677, A3C000824, A43000982, A2B000365, A36000707, EMO00005, A36000705, 34000623, A26000176, A26000180, A25000106, A25000108, A25000115, A34000565, A34000578, A31000476, A32000504, A36000670, A2B000385, A3B000794, A2A000344, A36000694, A3B000806, A3C000826, A39000767, A42000899, A2C000437, A21000016, A25000118, A2C000444, A38000756, A37000720, A45001058, A37000721, A34000612, A32000490, A28000226, A28000260, A24000077, A25000111, A31000475, A33000534, A24000076, A25000116, A35000652, A31000470, A36000716, A35000630, A44001032, A41000872, A2C000451, A2C000457, A32000498, A29000319, A27000210, A24000051, A31000468, A34000557, A34000567, A3B000802, A42000934, A42000921, A35000650, A27000203, A23000040, A34000563, A21000018, A32000495, A36000717, A39000778, A24000096, A25000126, A26000164, A27000194, A42000893, A42000903, A32000515, A34000615, A34000616, A2C000439, A33000536, A34000587, A34000606, A32000507, A25000137, A24000056, A34000614, A25000133, A44001045, A25000147, A23000042, A35000664, A25000130, A25000148, A32000516, A34000611, A43000995, A37000736, A42000894, A31000482, A2A000331, A32000499, A43000980, A35000654, A44001019, A25000110, A34000556, A35000641, A35000642, A44001027, A44001028, A42000913, A3C000813, A42000928, A45001051, A2B000381, A2B000372, A2C000450, A32000513, A24000073, A3B000790, A34000597, A25000145, A31000493, A28000220, A34000627, A35000669, A28000265, A37000743, A35000645, A27000200, A29000310, A28000247, A26000171, A42000922, A27000184, A34000560, A28000243, A24000055, A35000644, A29000292, A21000015, A33000528, A32000517, A27000211, A28000257, A28000230, A2A000340, A34000599, A28000262, A2C000443, A2C000465, A35000643, A2A000337, A28000250, A31000491  
Recalling Firm/
Manufacturer
Philips Ultrasound, Inc.
22100 Bothell-Everett Hwy.
P.O. Box 3003
Bothell WA 98041-3003
For Additional Information Contact Sarah Baxter
425-487-7665
Manufacturer Reason
for Recall
Potential for device to cause burning of the skin.
FDA Determined
Cause 2
Other
Action On 9/6/04 the firm sent a certified return receipt letter to customers advising of the issue and informing that a service representative will visit to install new software to reduce thermal output.
Quantity in Commerce 333 domestic units, 29 demos, 704 international units
Distribution Units were distributed to medial facilities, hospitals, and veterinarians worldwide. Countries include Africa, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Cyprus, Domican Republic, Ecuador, France, Germany, Greece, Guam, Guatemala, Hungary, India, Israel, Italy, Jordan, Kuwait, Martinique, Mexico, Netherlands, New Zealand, Norway, Phlippines, Portugal, Puerto Rico, Peru, Santa Domingo, Saudia Arabia, Selangor, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Ukraine, United Kingdom, Zenica
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = IYN and Original Applicant = MEDISON AMERICA, INC.
-
-