Date Initiated by Firm | December 03, 2004 |
Date Posted | February 16, 2005 |
Recall Status1 |
Terminated 3 on August 24, 2007 |
Recall Number | Z-0550-05 |
Recall Event ID |
31007 |
510(K)Number | K011160 K031571 |
Product Classification |
System, X-Ray, Tomography, Computed - Product Code JAK
|
Product | Brand Name of Device: EmpowerCT/EmpowerCTA Injector System Ceiling Column Accessory; Common/Usual Name: CT Injector System Ceiling Column Accessory, Catalog Number 9886L. Ceiling Column Accessory: The CT and CTA Injector Systems employs an Articulating Arm that mounts to the Ceiling Column Accessory, which are designated as 9886S (11.25''), M (21.50''), and L (34.50''), i.e. 9886L. |
Code Information |
No lot #s or serial #s assigned to this accessory |
Recalling Firm/ Manufacturer |
E-Z-EM, Inc 750 Summa Ave Westbury NY 11590-5040
|
For Additional Information Contact | Ms. Tracey A. Campbell 516-333-8230 Ext. 2307 |
Manufacturer Reason for Recall | Out of specification welding could cause the ceiling mount accessory to detach from its mounting. The firm received 1 complaint from a hospital that the injector fell off the ceiling mounting bracket and landed on the floor. |
FDA Determined Cause 2 | Other |
Action | On 12/3/04, "Urgent: Medical Device Recall" Letters (dated 12/3/04), were mailed via return receipt. |
Distribution | Medical facilities, 1 distributor & 1 physician nationwide. There were also foreign accounts in Canada, United Kingdom, France, The Netherlands and Korea. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|