  Class 2 Device Recall Medtronic Marquis DR see related information
Date Initiated by Firm February 09, 2005
Date Posted March 11, 2005
Recall Status1 Terminated 3 on December 28, 2006
Recall Number Z-0599-05
Recall Event ID 31142
PMA Number P980016S023 
Product Classification Implantable Cardioverter Defibrillator (Non-Crt) - Product Code LWS
Product Medtronic Marquis DR, model 7274 Dual Chamber Implantable Cardioverter Defibrillator (ICD) devices are tiered therapy implantable arrhythmia control devices, which provide automatic detection of ventricular arrhythmias and delivery of user-selected therapies for the detected arrhythmia. Therapies are automatically delivered following detection or can be initiated via programming by the physician. These ICDs are multi-programmable and have the capability of transmitting via telemetry the programmed values, measured and collected data, and event markers.
Code Information Model 7274 Marquis DR (serial numbers listed include all afffected devices but may include some unaffected devices. Patients are recommended to check with their doctor for exact serial numbers affected.) H-Series: PKC100001H - PKC138887H and PKC142234H.  R-series: PKC200320R - PKC200572R.  A-Series: PKC120001S - PKC120092S and PKC600001S - PKC617090S.
Recalling Firm/
Manufacturer
Medtronic Inc. Cardiac Rhythm Managment
7000 Central Ave Ne
Fridley MN 55432
For Additional Information Contact Tim Samsel
763-514-3149
Manufacturer Reason
for Recall
Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.
FDA Determined
Cause 2
Other
Action Medtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.
Quantity in Commerce 42,388 total, (34,620 in US & 7,768 OUS)
Distribution worldwide
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA Database PMAs with Product Code = LWS and Original Applicant = MEDTRONIC CARDIAC RHYTHM DISEASE MANAGEMENT
PMAs with Product Code = LWS and Original Applicant = Medtronic, Inc.
-
-