| Class 2 Device Recall |  |
Date Initiated by Firm | April 29, 2005 |
Date Posted | June 16, 2005 |
Recall Status1 |
Terminated 3 on May 09, 2008 |
Recall Number | Z-0896-05 |
Recall Event ID |
31997 |
510(K)Number | K032096 |
Product Classification |
System, Image Processing, Radiological - Product Code LLZ
|
Product | View Forum 2003 software, SW 3.4L1 |
Code Information |
The firm utilizes site numbers: 534544, 520177, 520633, 532261, 504927, 532134, 534573, 531862, 521928, 522027, 521424, 533804, 532053 |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact | Sarah Baxter 425-487-7665 |
Manufacturer Reason for Recall | Automated analysis of vascular obstruction tool in software gives unreliable calibration and analysis results when using unprocessed images acquired on the Philips Allura FD10. |
FDA Determined Cause 2 | Other |
Action | On 4/29/05 the firm mailed a letter, via certified return receipt. to their customers. The letter notified users of the problem and offered a work around. Consignees were advised they would be contacted by a representative to upgrade the unit with new software. |
Quantity in Commerce | 18 units |
Distribution | The units were distributed to 18 medical centers thorughout the US. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LLZ
|
|
|
|