| Class 2 Device Recall 10cc Nonsterile Manifold Kit | |
Date Initiated by Firm | June 27, 2005 |
Date Posted | August 03, 2005 |
Recall Status1 |
Terminated 3 on April 19, 2007 |
Recall Number | Z-1079-05 |
Recall Event ID |
32510 |
Product Classification |
System, X-Ray, Angiographic - Product Code IZI
|
Product | Labeling is as follows:
Product Labeling:
''***Argon Medical Devices***MANIFOLD KIT***SINGLE USE ONLY***REF/REORDER NO. 094337***EtO STERILIZATION RECOMMENDED***Caution: For manufacturing, processing or repacking.***1445 Flat Creek Road Athens, TX 75751 800-927-4669***''
Carton Labeling:
''***Argon Medical Devices***MANIFOLD KIT***SINGLE USE ONLY***REF/REORDER NO. 094337***Quantity: 20 ea***NON-STERILE*FURTHER PROCESSING REQUIRED***EtO (ethylene oxide) Sterilization Recommended***1445 Flat Creek Road Athens, TX 75751 800-927-4669***'' |
Code Information |
Reorder #: 00-094337 - Lot #: 99442306 Reorder #: 00-094367 - Lot #: 99443999 Reorder #: 00-094367 - Lot #: 99444179 |
Recalling Firm/ Manufacturer |
Argon Medical Devices, Inc 1445 Flat Creek Rd Athens TX 75751-5002
|
Manufacturer Reason for Recall | The firm is recalling the product due to the potential for air aspiration or fluid leakage from a faulty syringe tip. |
FDA Determined Cause 2 | Other |
Action | The firm plans to distribute recall letters to all customers indicated on the customer list. TELEPHONE 6/27/2005. All products will be returned to Argon and disposed. |
Quantity in Commerce | 700 syringes |
Distribution | TN, IL, WI, MA, VA, IN, LA, TX, CA, JAPAN, GERMANY, THE NETHERLANDS |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|