Date Initiated by Firm | July 19, 2005 |
Date Posted | September 01, 2005 |
Recall Status1 |
Terminated 3 on March 13, 2006 |
Recall Number | Z-1460-05 |
Recall Event ID |
32801 |
510(K)Number | K042479 |
Product Classification |
Enzyme Immunoassay, Gentamicin - Product Code LCD
|
Product | VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests. |
Code Information |
01-5846, Exp. 31 May 2006; 01-6036 Exp 31 October 2006; 01-6239 Exp 31 January 2007 |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Judy M. Strzepek 908-218-8524 |
Manufacturer Reason for Recall | Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours. |
FDA Determined Cause 2 | Other |
Action | Letters dated 7/19/2005 with instructions to process QC fluid according to the IFU when processing patient samples. When replacement product arrives, instructions are to destroy the remaining inventory of the recalled lots. |
Quantity in Commerce | 443 boxes |
Distribution | The product was shipped nationwide and to Ortho-Clinical/J&J affiliates in Australia, Canada, Singapore, England, France, Germany, Italy and Spain. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = LCD
|