Date Initiated by Firm | June 17, 2005 |
Date Posted | August 02, 2006 |
Recall Status1 |
Terminated 3 on February 24, 2012 |
Recall Number | Z-1318-06 |
Recall Event ID |
34953 |
510(K)Number | K041874 |
Product Classification |
blood gas analyzer - Product Code CHL
|
Product | Blood Gas and Co-Ox Electrolyte & Metabolyte Analyzer, ABL-800 series. |
Code Information |
All analyzers manufactured from 1/2005-10/31/2005. The devices contain serial numbers 902-441RxxxNxxx. The Rxxx number denotes manufacturing run, and the Nxxx number denotes the individual instrument within the run. |
Recalling Firm/ Manufacturer |
Radiometer America Inc 810 Sharon Dr Westlake OH 44145-1598
|
For Additional Information Contact | Vince Sigmund 800-736-0600 |
Manufacturer Reason for Recall | Software defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase. |
FDA Determined Cause 2 | Other |
Action | The customers were apprised of this situation with a recall letter issued by the firm on 6/17/2005. In addition, new Operators Manuals were sent to the customers. In the letter, the customers were advised that a Service Engineer would be contacting them shortly in order to arrange for the installation of the software upgrade. The firm''s Service Engineers visited each customer location and installed software upgrades. Version 3.831 software was installed on the 700 Series Analyzers and Version 4.14 was installed on the 800 Series Analyzers. According to the firm, these software corrections were verified as having been completed by 3/10/2006. |
Quantity in Commerce | 108 devices |
Distribution | Nationwide. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = CHL
|