• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medtronic

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medtronicsee related information
Date Initiated by FirmApril 05, 2006
Date PostedJune 01, 2006
Recall Status1 Terminated 3 on June 20, 2007
Recall NumberZ-0959-06
Recall Event ID 35236
510(K)NumberK021256 
Product Classification Catheter, Percutaneous - Product Code DQY
ProductCATH. GUIDE 6F SR3.5 W/SH LAUN CATHETER LA6SR35SH LA 6F 100CM SR35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peripheral vascular system. Product Number: LA6SR35SH
Code Information Lot Numbers: 0000021624 0000048665 0000053776 0000060280 0000068019 0000070770 0000077718 0000128585 204128 210659 214624 217180 221581 
Recalling Firm/
Manufacturer
Medtronic, Inc.
35 -37A Cherry Hill Dr
Danvers MA 01923-2565
For Additional Information ContactFred Boucher
978-739-3116
Manufacturer Reason
for Recall
Potential for non-sterility due to loss of package integrity
FDA Determined
Cause 2
Other
ActionMedtronic notified accounts and sales representatives on 4/05/06 advising the recalled product be segregated and held for return to the sales rep. and documented on the reconciliation form.
Quantity in Commerce195 units
DistributionNationwide Foreign: Austria, Australia, Bahrain, Bulgaria, Canada, Czech Republic, Denmark, Germany, Hungary, Italy, Jordan, France, Greece, Japan, Hong Kong, India, Italy, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DQY
-
-