|
Class 2 Device Recall Philips Medical |
 |
Date Initiated by Firm |
July 24, 2006 |
Date Posted |
September 12, 2006 |
Recall Status1 |
Terminated 3 on August 11, 2020 |
Recall Number |
Z-1487-06 |
Recall Event ID |
35785 |
510(K)Number |
K033715
|
Product Classification |
physiological patient monitoring system - Product Code DQA
|
Product |
IntelliVue MultiMeasurement Server (MMS)/Multimeasurement Server, physiological patient monitoring system. Model: M3001A |
Code Information |
M3001A options A01, A01C06, A01C12, and A01C18 with serial number prefix DE512 and MMS software revision C.0 or B.1 M3001A options A02, A02C06, and A02C18 with serial number prefix DE441 and DE512 and MMS software revision C.0 or B.1 |
Recalling Firm/ Manufacturer |
Philips Medical Systems 3000 Minuteman Road Andover MA 01810-1032
|
For Additional Information Contact |
LeeAn Rogus 978-659-4252
|
Manufacturer Reason for Recall |
Patient monitor may display inaccurate reading when the Disposable Sp02 Sensor is not attached
|
FDA Determined Cause 2 |
Other |
Action |
Philips is using the contract services of Sterling, Indianapolis, IN to contact all consignees regarding the correction/upgrade to the device. The letter notification dated 7/21/06 was mailed UPS Delivery Tracking the week of 07/24/06. Additionally, Philips included with the Urgent Device Correction Notice a procedure to mitigate risk until the upgrade is installed. |
Quantity in Commerce |
48,699 units |
Distribution |
Worldwide, including USA, Canada, Argentina, Australia, Austria, Belgium, Brazil, Czech, Croatia, China, Denmark, Finland, France, Germany, Israel, Italy, Jordan, Kuwait, Luxemburg, Netherlands, Oman, Norway, Poland, Portugual, Poland, Russia, South Africa, Spain, Sweden, Swizerland, Turkey, UAE, Yemen, UK, Zimbabwe, Zambia, India, Indonesia, Japan, Korea, Philipines, Malaysia, Singapore, Sri Lanka, Taiwan, Vietnam, Venezuela. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = DQA and Original Applicant = PHILIPS MEDICAL SYSTEMS, INC.
|
|
|
|