| | Class 2 Device Recall BacT/ALERT |  |
| Date Initiated by Firm | August 14, 2006 |
| Date Posted | May 03, 2007 |
| Recall Status1 |
Terminated 3 on December 08, 2009 |
| Recall Number | Z-0774-2007 |
| Recall Event ID |
36826 |
| 510(K)Number | K934593 |
| Product Classification |
Blood Culturing System - Product Code MDB
|
| Product | BacT/ALERT 3D System Software, Blood Culturing System |
| Code Information |
All Models |
Recalling Firm/ Manufacturer |
bioMerieux, Inc. 100 Rodolphe St Durham NC 27712-9402
|
| For Additional Information Contact | Brian Brundidge 919-620-2968 |
Manufacturer Reason for Recall | Plastic bottle samples assigned less sensitive default glass bottle algorithm instead of the plastic bottle algorithm on BacT/ALERT Microbial Detection Systems. |
FDA Determined Cause 2 | Other |
| Action | Consignees were notified via A Field Corrective Action Notification on/about 10/18/2006. A second notification was sent on October 31, 2006. |
| Quantity in Commerce | 10790 units, total for all models |
| Distribution | Worldwide, including USA, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dutch Antilles, Egypt, Ecuador, Estonia, Faeroe Islands, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Kenya, Korea, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania , Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = MDB
|
|
|
|