| Class 2 Device Recall Beckman Coulter Expo 32 ADC |  |
Date Initiated by Firm | November 01, 2006 |
Date Posted | January 09, 2007 |
Recall Status1 |
Terminated 3 on February 27, 2012 |
Recall Number | Z-0471-2007 |
Recall Event ID |
36958 |
Product Classification |
Flow Cytometers - Product Code JQP
|
Product | Expo32 ADC Software Version 1.1C, Part Number 6418337-New Users Software Kit, and Part Number 175454 Upgrade Software Kit used with the Coulter EPICS XL Flow Cytometers |
Code Information |
all codes with the Software version |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc 200 S Kraemer Blvd Brea CA 92822-6208
|
For Additional Information Contact | Nora Zerounian 714-961-3634 |
Manufacturer Reason for Recall | The firm has confirmed that the ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. This issue is limited to only protocols utilizing the Ratio parameter. |
FDA Determined Cause 2 | Other |
Action | A Product corrective Action (PCA) letter was mailed on 11/1/06 to Expo 32 software accounts. They are informed that the ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialogue box is accessed. The issue is limited to only protocols utilizing the Ratio parameter. An immediate user preventive action to remedy this issue each time acquisition of samples that require a protocol utilizing the ratio parameter is desired is given within the letter: 1) Perform acquisition of samples in batches which use protocols containing the same ratio parameter assignments. 2) Before running samples, open the protocol intended for use and open the Parameter Selection dialog box to ensure the correct ratio parameter assignments for numerator and denominator are selected. Make any necessary changes and save the protocol. 3) Run samples. (Letter was sent by US mail). A response form is included. |
Quantity in Commerce | 28 units |
Distribution | Nationwide including the states of MA, AL, CO, MI, IN, MD, MO NY, PA, TX, AR, CA, NJ, FL, and TN |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|