 Class 2 Device Recall Beckman Coulter Expo 32 ADCsee related information
Date Initiated by FirmNovember 01, 2006
Date PostedJanuary 09, 2007
Recall Status1 Terminated 3 on February 27, 2012
Recall NumberZ-0471-2007
Recall Event ID 36958
Product Classification Flow Cytometers - Product Code JQP
ProductExpo32 ADC Software Version 1.1C, Part Number 6418337-New Users Software Kit, and Part Number 175454 Upgrade Software Kit used with the Coulter EPICS XL Flow Cytometers
Code Information all codes with the Software version
Recalling Firm/
Manufacturer
Beckman Coulter Inc
200 S Kraemer Blvd
Brea CA 92822-6208
For Additional Information ContactNora Zerounian
714-961-3634
Manufacturer Reason
for Recall
The firm has confirmed that the ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. This issue is limited to only protocols utilizing the Ratio parameter.
FDA Determined
Cause 2
Other
ActionA Product corrective Action (PCA) letter was mailed on 11/1/06 to Expo 32 software accounts. They are informed that the ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialogue box is accessed. The issue is limited to only protocols utilizing the Ratio parameter. An immediate user preventive action to remedy this issue each time acquisition of samples that require a protocol utilizing the ratio parameter is desired is given within the letter: 1) Perform acquisition of samples in batches which use protocols containing the same ratio parameter assignments. 2) Before running samples, open the protocol intended for use and open the Parameter Selection dialog box to ensure the correct ratio parameter assignments for numerator and denominator are selected. Make any necessary changes and save the protocol. 3) Run samples. (Letter was sent by US mail). A response form is included.
Quantity in Commerce28 units
DistributionNationwide including the states of MA, AL, CO, MI, IN, MD, MO NY, PA, TX, AR, CA, NJ, FL, and TN
Total Product Life CycleTPLC Device Report