• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall VITALITY DR HE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall VITALITY DR HEsee related information
Date Initiated by FirmMarch 23, 2007
Date PostedJuly 11, 2007
Recall Status1 Terminated 3 on March 02, 2008
Recall NumberZ-1054-2007
Recall Event ID 37769
PMA NumberP010012 
Product Classification implantable defibrillator - Product Code NIK
Product VITALITY DR HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. This ICD is designed to detect and terminate ventricular tachycardia (VT) and ventricular fibrillation (VF) and provide bradycardia therapy (atrial and ventricular pacing). Therapies include both low- and high-eneergy shocks using either a biphasic or monophasic waveform. Bradycardia pacing, including adaptive-rate features, is available to detect and treat bradyarrhythmias and to support the cardiac rhythm after defibrillation therapy. The device offers dual-chamber bradycardia features (atrial and/or ventricular pacing and sensing).
Code Information Model T180, serial numbers: 200020, 200025, 200029, 200059, 200063, 200065, 200068, 200088, 200109, 200111, 200210, 200237, 200358, 200360, 200362, 200440, 200443, 200480, 200581, 200584, 200627, 200663, 200700, 200717, 200725, 200727, 200730, 200743, 200774, 200801, 200847, 200863, 200868, 200884, 200916, 200918, 200962, 200979, 201013, 201026, 201064, 201077, 201082, 201108, 201115, 201124, 201138, 201149, 201151, 201166, 201171, 201182, 201186, 201222, 201224, 201227, 201243, 201263, 201281, 201283, 201287, 201290, 201297, 201298, 201308, 201332, 201388, 201393, 201396, 201451, 201479, 201481, 201524, 201530, 201539, 201548, 201550, 201582, 201589, 201592, 201617, 201629, 201634, 201649, 201685, 201700, 201701, 201702, 201704, 201731, 201732, 201759, 201763, 201768, 201838, 201841, 201861, 201872, 201874, 201891, 201901, 201905, 201908, 201909, 201927, 201929, 201930, 201931, 201933, 201934, 201936, 201937, 201938, 201939, 201940, 201941, 201942, 201943, 201944, 201945, 201946, 201947, 201948, 201949, 201950, 201951, 201952, 201953, 201954, 201955, 201956, 201957, 201958, 201959, 201960, 201961, 201962, 201963, 201964, 201965, 201966, 201967, 201968, 201969, 201970, 201971, 201973, 201974, 201975, 201976, 201977, 201978, 201979, 201980, 201981, 201982, 201983, 201984, 201985, 201988, 201990, 201991, 201992, 201993, 201994, 201995, 201997, 201998, 201999, 202000, 202001, 202002, 202003, 202004, 202005, 202006, 202007, 202008, 202009, 202010, 202012, 202013, 202014, 202015, 202017, 202018, 202019, 202020, 202021, 202022, 202023, 202024, 202025, 202026, 202028, 202029, 202031, 202032, 202033, 202034, 202036, 202037, 202038, 202039, 202041, 202042, 202043, 202044, 202045, 202046, 202047, 202048, 202049, 202050, 202051, 202052, 202053, 202054, 202055, 202056, 202057, 202058, 202059, 202060, 202061, 202062, 202063, 202066, 202068, 202069, 202070, 202071, 202072, 202073, 202074, 202075, 202076, 202077, 202078, 202079, 202080, 202081, 202082, 202083, 202085, 202086, 202087, 202089, 202090, 202091, 202092, 202094, 202095, 202096, 202097, 202099, 202100, 202101, 202102, 202103, 202104, 202105, 202107, 202108, 202109, 202110, 202111, 202112, 202113, 202114, 202115, 202116, 202117, 202118, 202119, 202121, 202122, 202123, 202124, 202125, 202127, 202128, 202129, 202130, 202131, 202132, 202134, 202135, 202136, 202137, 202138, 202139, 202140, 202141, 202143, 202144, 202145, 202146, 202147, 202148, 202149, 202150, 202152, 202154, 202155, 202156, 202157, 202158, 202159, 202160, 202161, 202162, 202163, 202164, 202165, 202166, 202167, 202168, 202169, 202170, 202171, 202172, 202173, 202174, 202175, 202176, 202177, 202178, 202179, 202180, 202181, 202182, 202183, 202185, 202186, 202187, 202188, 202189, 202191, 202192, 202193, 202194, 202195, 202196, 202197, 202198, 202199, 202201, 202202, 202203, 202204, 202205, 202206, 202207, 202208, 202210, 202211, 202213, 202214, 202215, 202216, 202217, 202218, 202219, 202220, 202221, 202222, 202223, 202224, 202225, 202226, 202227, 202228, 202229, 202230, 202231, 202233, 202234, 202235, 202236, 202237, 202238, 202239, 202241, 202242, 202243, 202244, 202245, 202246, 202247, 202248, 202249, 202251, 202252, 202253, 202254, 202255, 202257, 202258, 202259, 202260, 202261, 202262, 202263, 202264, 202265, 202266, 202267, 202268, 202269, 202270, 202271, 202274, 202275, 202276, 202277, 202278, 202279, 202280, 202282, 202283, 202284, 202285, 202286, 202287, 202288, 202289, 202291, 202292, 202293, 202294, 202295, 202296, 202297, 202298, 202299, 202300, 202301, 202302, 202303, 202305, 202306, 202307, 202308, 202309, 202312, 202313, 202314, 202315, 202316, 202317, 202318, 202319, 202320, 202322, 202323, 202324, 202325, 202326, 202327, 202328, 202329, 202330, 202331, 202332, 202333, 202334, 202335, 202336, 202338, 202339, 202340, 202341, 202342, 202343, 202344, 202345, 202346, 202347, 202349, 202350, 202351, 202352, 202353, 202354, 202355, 202356, 202357, 202358, 202359, 202360, 202361, 202362, 202363, 202364, 202365, 202367, 202368, 202369, 202370, 202371, 202372, 202373, 202374, 202375, 202376, 202377, 202378, 202379, 202380, 202381, 202382, 202383, 202384, 202385, 202386, 202387, 202388, 202389, 202390, 202391, 202392, 202393, 202394, 202395, 202396, 202397, 202398, 202399, 202400, 202401, 202404, 202405, 202407, 202408, 202409, 202411, 202412, 202413, 202414, 202415, 202416, 202417, 202418, 202419, 202420, 202421, 202422, 202423, 202424, 202425, 202426, 202427, 202428, 202429, 202431, 202433, 202434, 202435, 202436, 202437, 202438, 202439, 202440, 202441, 202442, 202443, 202444, 202445, 202446, 202447, 202448, 202449, 202450, 202451, 202452, 202453, 202454, 202455, 202456, 202457, 202458, 202459, 202460, 202461, 202462, 202463, 202464, 202465, 202466, 202467, 202469, 202470, 202471, 202472, 202473, 202474, 202475, 202476, 202477, 202480, 202481, 202482, 202483, 202484, 202485, 202486, 202487, 202488, 202490, 202491, 202492, 202493, 202494, 202495, 202496, 202497, 202499, 202500, 202501, 202502, 202503, 202505, 202506, 202507, 202508, 202509, 202511, 202513, 202514, 202516, 202517, 202518, 202520, 202522, 202523, 202524, 202526, 202528, 202529, 202530, 202534, 202535, 202536, 202537, 202538, 202539, 202540, 202541, 202542, 202544, 202546, 202548, 202549, 202550, 202551, 202552, 202553, 202554, 202555, 202556, 202557, 202558, 202560, 202561, 202562, 202563, 202564, 202565, 202566, 202567, 202568, 202569, 202570, 202572, 202573, 202574, 202575, 202577, 202578, 202579, 202580, 202581, 202582, 202583, 202584, 202585, 202586, 202588, 202590, 202591, 202592, 202594, 202596, 202597, 202598, 202599, 202600, 202601, 202603, 202604, 202606, 202608, 202609, 202610, 202611, 202612, 202614, 202616, 202618, 202619, 202621, 202622, 202624, 202625, 202626, 202628, 202630, 202631, 202633, 202634, 202635, 202636, 202638, 202639, 202640, 202641, 202642, 202643, 202645, 202646, 202647, 202648, 202649, 202650, 202651, 202653, 202655, 202656, 202658, 202659, 202660, 202662, 202666, 202670, 202673, 202674, 202675, 202676, 202677, 202679, 202680, 202681, 202682, 202687, 202688, 202689, 202691, 202692, 202693, 202694, 202696, 202697, 202698, 202699, 202702, 202703, 202704, 202705, 202706, 202709, 202710, 202711, 202712, 202713, 202714, 202715, 202717, 202718, 202722, 202724, 202725, 202734, 202735, 202736, 202738, 202740, 202744, 202745, 202748, 202749, 202750, 202751, 202752, 202757, 202758, 202760, 202761, 202762, 202764, 202766, 202771, 202773, 202774, 202777, 202778, 202779, 202780, 202781, 202783, 202784, 202785, 202786, 202787, 202788, 202789, 202790, 202791, 202793, 202794, 202795, 202796, 202797, 202798, 202799, 202801, 202802, 202803, 202804, 202805, 202806, 202808, 202809, 202812, 202813, 202816, 202818, 202819, 202820, 202821, 202822, 202823, 202825, 202826, 202827, 202828, 202829, 202830, 202831, 202832, 202834, 202835, 202836, 202837, 202838, 202839, 202840, 202841, 202842, 202843, 202844, 202848, 202849, 202851, 202852, 202854, 202856, 202858, 202860, 202861, 202863, 202866, 202867, 202868, 202869, 202871, 202876, 202877, 202878, 202879, 202880, 202881, 202882, 202883, 202884, 202886, 202887, 202888, 202889, 202892, 202895, 202896, 202897, 202899, 202900, 202901, 202902, 202903, 202904, 202906, 202908, 202909, 202910, 202911, 202912, 202913, 202914, 202917, 202918, 202919, 202921, 202922, 202923, 202924, 202926, 202927, 202928, 202929, 202930, 202931, 202932, 202933, 202934, 202935, 202938, 202939, 202940, 202941, 202942, 202943, 202944, 202945, 202946, 202947, 202951, 202952, 202953, 202957, 202958, 202959, 202960, 202961, 202963, 202964, 202965, 202966, 202970, 202971, 202974, 202976, 202979, 202984, 202986, 202988, 202989, 202990, 202992, 202993, 202995, 202997, 202999, 203001, 203011, 203013, 203015, 203016, 203017, 203018, 203021, 203022, 203023, 203028, 203030, 203034, 203036, 203037, 203038, 203040, 203047, 203048, 203055, 203064, 203066, 203067, 203070, 203072, 203078, 203087, 203095, 203097, 203118, 203119, 203121, 203127, 203134, 203137, 203148, 203153, 203155, 203160, 203163, 203164, 203165, 203166, 203167, 203168, 203173, 203174, 203182, 203183, 203184, 203185, 203186, 203187, 203188, 203190, 203195, 203196, 203197, 203204, 203206, 203207, 203210, 203211, 203212, 203213, 203215, 203216, 203217, 203219, 203220, 203223, 203224, 203225, 203227, 203228, 203232, 203235, 203236, 203237, 203239, 203240, 203241, 203244, 203248, 203250, 203258, 203261, 203267, 203269, 203271, 203276, 203280, 203283, 203285, 203289, 203290, 203291, 203293, 203294, 203301, 203302, 203308, 203309, 203312, 203318, 203324, 203327, 203328, 203329, 203335, 203336, 203341, 203346, 203349, 203350, 203353, 203355, 203365, 203367, 203380, 203382, 203386, 203387, 203388, 203390, 203397, 203398, 203405, 203418, 203424, 203426, 203427, 203438, 203453, 203462, 203464, 203465, 203468, 203481, 203488, 203493, 203497, 203498, 203501, 203598, 203599, 203602, 203603, 203606, 203610, 203611, 203615, 203616, 203619, 203627, 203630, 203635, 203637, 203646, 203652, 203662, 203668, 203673, 203674, 203675, 203676, 203680, 203682, 203683, 203685, 203689, 203690, 203693, 203698, 203700, 203703, 203706, 203707, 203712, 203716, 203718, 203722, 203724, 203731, 203734, 203741, 203748, 203751, 203756, 203762, 203769, 203776, 203779, 203782, 203785, 203786, 203788, 203791, 203792, 203793, 203803, 203850, 203856, 203860, 203861, 203862, 203863, 203864, 203865, 203877, 203879, 203886, 203891, 203893, 203898, 203899, 203902, 203903, 203910, 203913, 203917, 203924, 203925, 203927, 203928, 203929, 203930, 203931, 203932, 203935, 203937, 203938, 203939, 203940, 203941, 203942, 203943, 203945, 203946, 203947, 203948, 203949, 203950, 203951, 203952, 203953, 203954, 203955, 203956, 203959, 203960, 203963, 203965, 203966, 203967, 203968, 203969, 203970, 203971, 203972, 203973, 203974, 203975, 203976, 203978, 203980, 203984, 203985, 203986, 203987, 203988, 203989, 203990, 203993, 203996, 204014, 204017, 204018, 204020, 204025, 204029, 204030, 204031, 204035, 204037, 204045, 204046, 204052, 204053, 204054, 204055, 204059, 204061, 204063, 204064, 204070, 204074, 204075, 204076, 204091, 204093, 204094, 204098, 204101, 204103, 204106, 204113, 204114, 204117, 204118, 204119, 204122, 204124, 204130, 204135, 204146, 204150, 204152, 204153, 204162, 204171, 204176, 204181, 204191, 204193, 204197, 204199, 204205, 204206, 204207, 204212, 204214, 204216, 204217, 204219, 204220, 204225, 204227, 204233, 204243, 204249, 204252, 204253, 204254, 204256, 204259, 204260, 204271, 204272, 204274, 204275, 204283, 204291, 204294, 204296, 204302, 204314, 204316, 204323, 204330, 204336, 204337, 204339, 204342, 204343, 204344, 204356, 204361, 204371, 204372, 204374, 204378, 204412, 204415, 204429, 204430, 204434, 204452, 204465, 204467, 204470, 204471, 204472, 204473, 204474, 204475, 204477, 204479, 204485, 204486, 204489, 204492, 204493, 204495, 204498, 204499, 204505, 204509, 204513, 204517, 204536, 204554, 204557, 204558, 204559, 204572, 204581, 204584, 204587, 204594, 204597, 204601, 204605, 204606, 204608, 204611, 204620, 204629, 204633, 204637, 204640, 204642, 204643, 204645, 204648, 204665, 204669, 204670, 204672, 204673, 204676, 204677, 204679, 204680, 204689, 204699, 204702, 204705, 204708, 204712, 204715, 204723, 204724, 204729, 204730, 204734, 204735, 204738, 204746, 204752, 204753, 204758, 204760, 204770, 204773, 204778, 204786, 204789, 204790, 204794, 204797, 204799, 204804, 204805, 204807, 204823, 204825, 204834, 204835, 204840, 204842, 204854, 204856, 204858, 204861, 204863, 204864, 204867, 204872, 204874, 204885, 204896, 204897, 204903, 204909, 204913, 204914, 204916, 204917, 204920, 204921, 204922, 204923, 204936, 204940, 204945, 204946, 204947, 204948, 204950, 204951, 204952, 204955, 204957, 204958, 204959, 204960, 204961, 204962, 204963, 204964, 204968, 204969, 204970, 204973, 204974, 204975, 204979, 204981, 204982, 204983, 204984, 204985, 204986, 204987, 204989, 204990, 204991, 204992, 204993, 204994, 204995, 204996, 204997, 204998, 204999, 205000, 205001, 205002, 205003, 205004, 205006, 205007, 205008, 205009, 205010, 205011, 205013, 205014, 205017, 205018, 205020, 205021, 205022, 205023, 205024, 205025, 205026, 205028, 205029, 205030, 205031, 205032, 205033, 205034, 205035, 205036, 205037, 205038, 205039, 205040, 205041, 205042, 205043, 205044, 205045, 205046, 205047, 205048, 205049, 205050, 205051, 205052, 205053, 205054, 205055, 205056, 205057, 205058, 205059, 205060, 205061, 205062, 205063, 205064, 205065, 205066, 205067, 205069, 205070, 205071, 205072, 205073, 205074, 205075, 205076, 205077, 205078, 205079, 205080, 205081, 205082, 205084, 205085, 205086, 205088, 205089, 205090, 205091, 205092, 205093, 205094, 205095, 205096, 205098, 205099, 205100, 205103, 205104, 205105, 205106, 205107, 205109, 205111, 205113, 205114, 205125, 205126, 205127, 205128, 205129, 205130, 205131, 205132, 205133, 205135, 205136, 205137, 205138, 205139, 205140, 205141, 205142, 205144, 205145, 205147, 205148, 205149, 205151, 205152, 205154, 205155, 205156, 205157, 205158, 205159, 205160, 205161, 205162, 205163, 205164, 205165, 205166, 205167, 205168, 205169, 205170, 205171, 205172, 205173, 205174, 205175, 205176, 205178, 205179, 205180, 205181, 205182, 205183, 205184, 205185, 205186, 205187, 205188, 205189, 205190, 205191, 205192, 205193, 205194, 205195, 205196, 205197, 205198, 205199, 205200, 205201, 205202, 205203, 205204, 205205, 205206, 205207, 205208, 205209, 205210, 205211, 205212, 205214, 205215, 205216, 205217, 205218, 205220, 205221, 205222, 205223, 205224, 205225, 205226, 205227, 205228, 205229, 205230, 205231, 205232, 205233, 205234, 205235, 205238, 205239, 205240, 205243, 205245, 205246, 205247, 205249, 205250, 205251, 205252, 205253, 205255, 205256, 205257, 205258, 205259, 205262, 205263, 205264, 205267, 205268, 205269, 205270, 205271, 205272, 205273, 205274, 205275, 205277, 205278, 205280, 205282, 205284, 205285, 205286, 205287, 205289, 205291, 205292, 205293, 205294, 205295, 205296, 205298, 205300, 205301, 205304, 205305, 205307, 205308, 205309, 205310, 205311, 205312, 205313, 205314, 205316, 205317, 205318, 205319, 205321, 205322, 205323, 205324, 205325, 205326, 205327, 205328, 205329, 205330, 205331, 205332, 205333, 205334, 205335, 205337, 205338, 205339, 205340, 205343, 205345, 205346, 205347, 205348, 205349, 205351, 205352, 205353, 205355, 205356, 205358, 205360, 205361, 205362, 205363, 205364, 205368, 205369, 205370, 205371, 205372, 205373, 205375, 205376, 205379, 205406, 205407, 205409, 205410, 205411, 205413, 205414, 205435, 205437, 205438, 205445, 205461, 205465, 205466, 205467, 205468, 205473, 205476, 205477, 205478, 205479, 205491, 205496, 205498, 205500, 205504, 205505, 205506, 205507, 205508, 205509, 205511, 205513, 205518, 205528, 205529, 205531, 205532, 205539, 205555, 205559, 205561, 205580, 205587, 205588, 205592, 205600, 205603, 205612, 205624, 205625, 205630, 205631, 205633, 205634, 205635, 205636, 205637, 205649, 205935, 205939, 205940, 205943, 200002, 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200011, 200012, 200013, 200014, 200015, 200016, 200017, 200018, 200021, 200022, 200023, 200024, 200026, 200027, 200028, 200030, 200031, 200032, 200033, 200035, 200036, 200037, 200038, 200039, 200040, 200041, 200042, 200043, 200044, 200045, 200046, 200047, 200048, 200050, 200052, 200053, 200054, 200055, 200056, 200057, 200060, 200061, 200062, 200067, 200071, 200072, 200073, 200074, 200075, 200076, 200079, 200081, 200082, 200083, 200084, 200085, 200086, 200087, 200089, 200090, 200092, 200093, 200094, 200095, 200097, 200098, 200099, 200102, 200104, 200107, 200108, 200110, 200113, 200116, 200118, 200119, 200120, 200121, 200122, 200124, 200125, 200126, 200127, 200139, 200143, 200145, 200146, 200158, 200159, 200161, 200163, 200165, 200166, 200176, 200178, 200179, 200185, 200187, 200188, 200191, 200193, 200197, 200208, 200209, 200211, 200213, 200215, 200216, 200217, 200218, 200219, 200222, 200223, 200224, 200225, 200227, 200228, 200229, 200230, 200231, 200232, 200234, 200236, 200238, 200239, 200243, 200253, 200255, 200267, 200276, 200279, 200280, 200288, 200289, 200299, 200314, 200315, 200319, 200321, 200322, 200324, 200325, 200326, 200327, 200328, 200329, 200347, 200348, 200359, 200361, 200363, 200364, 200365, 200370, 200373, 200374, 200375, 200377, 200380, 200386, 200389, 200396, 200407, 200418, 200419, 200421, 200422, 200423, 200429, 200430, 200431, 200433, 200441, 200446, 200449, 200450, 200451, 200459, 200460, 200464, 200482, 200485, 200487, 200493, 200496, 200497, 200508, 200509, 200510, 200511, 200512, 200513, 200514, 200516, 200517, 200518, 200519, 200522, 200523, 200524, 200526, 200527, 200528, 200532, 200534, 200535, 200539, 200540, 200543, 200547, 200549, 200551, 200553, 200554, 200555, 200556, 200558, 200559, 200560, 200562, 200563, 200564, 200565, 200566, 200567, 200571, 200573, 200574, 200575, 200576, 200578, 200579, 200580, 200582, 200583, 200585, 200586, 200587, 200588, 200589, 200590, 200591, 200592, 200593, 200595, 200597, 200598, 200599, 200600, 200601, 200602, 200603, 200604, 200605, 200606, 200607, 200610, 200611, 200614, 200617, 200618, 200622, 200628, 200631, 200648, 200649, 200650, 200652, 200657, 200658, 200660, 200673, 200674, 200676, 200677, 200678, 200679, 200681, 200682, 200683, 200684, 200685, 200686, 200687, 200689, 200690, 200691, 200692, 200694, 200695, 200696, 200697, 200698, 200699, 200703, 200704, 200705, 200706, 200707, 200708, 200709, 200710, 200711, 200712, 200713, 200714, 200715, 200716, 200718, 200719, 200720, 200721, 200722, 200723, 200724, 200729, 200732, 200735, 200736, 200737, 200738, 200739, 200740, 200741, 200742, 200745, 200746, 200747, 200749, 200750, 200751, 200752, 200753, 200754, 200755, 200756, 200757, 200758, 200759, 200760, 200763, 200764, 200765, 200766, 200767, 200768, 200770, 200773, 200775, 200776, 200777, 200779, 200780, 200781, 200782, 200784, 200785, 200786, 200787, 200788, 200792, 200794, 200795, 200798, 200799, 200802, 200807, 200808, 200815, 200818, 200819, 200843, 200845, 200846, 200848, 200849, 200851, 200855, 200860, 200861, 200865, 200867, 200869, 200870, 200871, 200872, 200873, 200874, 200875, 200877, 200879, 200880, 200881, 200882, 200883, 200886, 200890, 200891, 200896, 200897, 200898, 200899, 200900, 200902, 200903, 200904, 200907, 200908, 200910, 200912, 200913, 200914, 200917, 200919, 200920, 200921, 200922, 200923, 200924, 200925, 200926, 200927, 200929, 200930, 200931, 200932, 200936, 200938, 200940, 200941, 200943, 200947, 200948, 200949, 200951, 200952, 200953, 200954, 200961, 200963, 200964, 200965, 200966, 200967, 200972, 200973, 200975, 200976, 200977, 200980, 200981, 200982, 200989, 200990, 200991, 200992, 200993, 200994, 200996, 200997, 200999, 201000, 201004, 201008, 201009, 201010, 201011, 201012, 201014, 201015, 201016, 201018, 201019, 201020, 201021, 201023, 201024, 201025, 201027, 201029, 201030, 201032, 201033, 201034, 201035, 201037, 201040, 201041, 201042, 201063, 201065, 201066, 201067, 201068, 201069, 201070, 201071, 201072, 201074, 201075, 201076, 201079, 201080, 201081, 201083, 201084, 201085, 201086, 201087, 201088, 201089, 201090, 201092, 201093, 201094, 201096, 201097, 201098, 201099, 201100, 201101, 201102, 201103, 201104, 201105, 201107, 201109, 201110, 201111, 201112, 201114, 201116, 201117, 201118, 201119, 201120, 201122, 201123, 201125, 201126, 201127, 201128, 201129, 201130, 201131, 201132, 201133, 201135, 201137, 201139, 201140, 201141, 201142, 201143, 201144, 201145, 201146, 201147, 201150, 201152, 201153, 201154, 201155, 201156, 201158, 201159, 201160, 201161, 201162, 201163, 201164, 201165, 201168, 201169, 201170, 201173, 201174, 201175, 201176, 201177, 201178, 201180, 201181, 201183, 201185, 201187, 201188, 201190, 201191, 201192, 201193, 201194, 201196, 201197, 201198, 201199, 201202, 201203, 201206, 201207, 201208, 201209, 201210, 201211, 201212, 201214, 201215, 201216, 201217, 201220, 201221, 201223, 201226, 201228, 201229, 201231, 201233, 201234, 201235, 201236, 201237, 201238, 201239, 201240, 201241, 201242, 201244, 201246, 201247, 201248, 201249, 201250, 201251, 201252, 201253, 201254, 201255, 201257, 201258, 201260, 201261, 201264, 201265, 201266, 201267, 201269, 201270, 201273, 201274, 201275, 201276, 201277, 201278, 201279, 201282, 201284, 201285, 201286, 201289, 201291, 201292, 201293, 201294, 201295, 201296, 201299, 201300, 201301, 201302, 201303, 201304, 201305, 201306, 201307, 201309, 201310, 201313, 201314, 201317, 201318, 201319, 201320, 201321, 201322, 201324, 201325, 201326, 201327, 201328, 201329, 201330, 201331, 201335, 201337, 201338, 201340, 201341, 201342, 201343, 201345, 201346, 201347, 201350, 201351, 201352, 201353, 201354, 201355, 201356, 201357, 201358, 201359, 201360, 201361, 201362, 201363, 201364, 201365, 201366, 201367, 201368, 201369, 201370, 201371, 201372, 201373, 201374, 201375, 201377, 201378, 201379, 201380, 201381, 201382, 201383, 201384, 201385, 201387, 201389, 201390, 201391, 201394, 201395, 201397, 201398, 201399, 201400, 201401, 201402, 201403, 201404, 201405, 201406, 201407, 201409, 201410, 201411, 201412, 201413, 201414, 201416, 201417, 201418, 201419, 201420, 201422, 201423, 201424, 201425, 201426, 201427, 201428, 201429, 201431, 201433, 201434, 201436, 201437, 201439, 201440, 201441, 201442, 201443, 201444, 201445, 201446, 201447, 201448, 201449, 201450, 201452, 201453, 201454, 201455, 201456, 201457, 201459, 201460, 201461, 201462, 201463, 201464, 201465, 201466, 201467, 201468, 201469, 201470, 201472, 201473, 201474, 201475, 201476, 201477, 201478, 201480, 201482, 201484, 201485, 201486, 201487, 201488, 201489, 201490, 201491, 201492, 201493, 201494, 201495, 201497, 201498, 201499, 201500, 201501, 201502, 201504, 201505, 201506, 201507, 201508, 201510, 201511, 201513, 201514, 201515, 201516, 201517, 201518, 201519, 201520, 201521, 201522, 201523, 201527, 201528, 201529, 201531, 201534, 201536, 201537, 201538, 201540, 201541, 201542, 201543, 201544, 201545, 201546, 201547, 201549, 201551, 201553, 201554, 201555, 201556, 201557, 201558, 201559, 201560, 201561, 201562, 201563, 201565, 201566, 201568, 201569, 201570, 201571, 201572, 201573, 201574, 201575, 201576, 201578, 201579, 201581, 201583, 201584, 201585, 201586, 201587, 201588, 201590, 201593, 201594, 201595, 201596, 201599, 201600, 201601, 201602, 201603, 201605, 201606, 201607, 201608, 201610, 201611, 201612, 201615, 201616, 201618, 201619, 201620, 201622, 201623, 201624, 201625, 201626, 201627, 201628, 201630, 201631, 201632, 201635, 201636, 201637, 201638, 201639, 201640, 201641, 201642, 201643, 201644, 201645, 201646, 201647, 201648, 201650, 201651, 201653, 201654, 201655, 201656, 201657, 201658, 201659, 201661, 201662, 201664, 201666, 201667, 201668, 201669, 201670, 201672, 201673, 201674, 201675, 201676, 201677, 201678, 201679, 201680, 201681, 201682, 201683, 201684, 201687, 201691, 201693, 201694, 201695, 201696, 201697, 201698, 201699, 201703, 201706, 201707, 201709, 201710, 201711, 201712, 201714, 201716, 201717, 201719, 201720, 201721, 201722, 201724, 201725, 201726, 201727, 201728, 201729, 201730, 201733, 201734, 201735, 201736, 201738, 201739, 201740, 201742, 201743, 201745, 201747, 201748, 201749, 201750, 201751, 201752, 201753, 201754, 201757, 201758, 201761, 201762, 201764, 201765, 201766, 201767, 201769, 201770, 201772, 201773, 201774, 201775, 201776, 201778, 201779, 201780, 201782, 201785, 201786, 201788, 201789, 201790, 201791, 201792, 201793, 201795, 201796, 201797, 201798, 201799, 201800, 201801, 201802, 201803, 201804, 201805, 201806, 201807, 201808, 201809, 201810, 201811, 201812, 201813, 201814, 201816, 201817, 201818, 201819, 201820, 201821, 201822, 201824, 201826, 201827, 201828, 201829, 201830, 201831, 201832, 201833, 201835, 201836, 201837, 201839, 201840, 201842, 201843, 201845, 201846, 201847, 201848, 201849, 201850, 201851, 201852, 201853, 201854, 201856, 201859, 201860, 201862, 201863, 201864, 201865, 201866, 201867, 201868, 201869, 201870, 201871, 201875, 201876, 201877, 201878, 201879, 201880, 201881, 201882, 201883, 201884, 201886, 201887, 201888, 201889, 201890, 201892, 201894, 201895, 201897, 201898, 201899, 201900, 201903, 201904, 201906, 201907, 201910, 201912, 201913, 201914, 201915, 201916, 201917, 201918, 201920, 201921, 201922, 201923, 201924, 201925, 201926, 202403, 202545, 204685.
Recalling Firm/
Manufacturer
Boston Scientific CRM Corp
4100 Hamline Ave N
Saint Paul MN 55112-5700
Manufacturer Reason
for Recall
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.
FDA Determined
Cause 2
Other
ActionAn Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.
Quantity in Commerce3,041
DistributionWorldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Polynesia, Portugal, Puerto Rico, Qatar, Saint Lucia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. Virgin Islands, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, and West Indies.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA DatabasePMAs with Product Code = NIK
-
-