Date Initiated by Firm |
October 18, 2007 |
Date Posted |
December 11, 2007 |
Recall Status1 |
Terminated 3 on July 18, 2008 |
Recall Number |
Z-0375-2008 |
Recall Event ID |
44972 |
510(K)Number |
K894914 K912387 K001154
|
Product Classification |
Ion Specific Electrode (Sodium) - Product Code JGS
|
Product |
Abbott Point of Care, i-STAT G Cartridge; (Glucose/Glu); List No: 06F09-01; Part No: 120100; Abbott Point of Care, Inc., East Windsor, NJ 08520, USA; Product of Canada. |
Code Information |
Catalog Number: 06F09-01; 06F09-02 Lot Numbers: A07102A; A07124; A07138; D07117; D07119; D07120; and D07121. |
Recalling Firm/ Manufacturer |
Abbott Point of Care Inc. 104 Windsor Center Dr East Windsor NJ 08520
|
For Additional Information Contact |
609-443-9300
|
Manufacturer Reason for Recall |
Difficult to Close -- Certain lots of i-STAT cartridges are difficult to close, will not close or do not stay closed.
|
FDA Determined Cause 2 |
Packaging change control |
Action |
Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately. |
Distribution |
Worldwide Distribution- including USA and country of Canada. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JGS and Original Applicant = I-STAT CORP.
|