Class 2 Device Recall PremierEdge MICROSURGICAL KNIFE 3.0mm Clear Cornea Knife Parallel Sides |
 |
Date Initiated by Firm |
July 18, 2007 |
Date Posted |
January 12, 2008 |
Recall Status1 |
Terminated 3 on September 25, 2009 |
Recall Number |
Z-0499-2008 |
Recall Event ID |
45870 |
Product Classification |
Ophthalmic Knife - Product Code HNN
|
Product |
PremierEdge MICROSURGICAL KNIFE, 3.0mm Clear Cornea Knife - Parallel Sides, Item Code/REF: PE 3230 , OASIS Glendora, CA 91741 |
Code Information |
Lot No: DK0207AA |
Recalling Firm/ Manufacturer |
Oasis Medical Inc 514 S Vermont Ave Bldg 510-528 Glendora CA 91741-6205
|
For Additional Information Contact |
Robyn Scopis 800-528-9786
|
Manufacturer Reason for Recall |
Sterility (package integrity) compromised: the device is puncturing the packing during handling. Once the packaging is damaged, the product is no longer sterile as labeled. THIS IS AN EXPANSION OF A PREVIOUS RECALL
|
FDA Determined Cause 2 |
Device Design |
Action |
A recall notification, Urgent Medical Device Recall Expansion Letter, dated 7.18.2007, was sent to customers advising the letter is to inform them of a voluntary product recall expansion involving the OASIS Medical Premier Edge Knives. A product label for ease in identifying the product is included. The letter further advises this voluntary recall has been expanded due to the product puncturing the packing during rough handling. Once the packaging is damaged, the product is no longer sterile as labeled. The letter instructs customers to immediately examine their inventory and quarantine product subject to recall. Once the product has been identified, product is to be returned by FedEx to OASIS
Medical. The letter further instructs that if the customer has further distributed this product, to identify their customers and notify them at once of this product recall. This recall should be carried out to the user level. |
Quantity in Commerce |
28,319 individual knives (Z-0485-0516-2008) |
Distribution |
Worldwide: USA, Kuwait, Lebanon, Lithuania, Mexico, Netherlands, Norway, South Africa, Saudi Arabia, Slovakia, Spain, Sweden, Switzwerland, Turkey, United Kingdom, Ecuador, Finland, France, Germany, Greece, Italy, Denmark, Columbia, Chile, Brazil, Belgium, and Australia |