New Search Back to Search Results
 Class 1 Device Recall Dura Starsee related information
Date Initiated by FirmJanuary 14, 2008
Date PostedFebruary 06, 2008
Recall Status1 Terminated 3 on October 20, 2009
Recall NumberZ-0748-2008
Recall Event ID 46378
PMA NumberP880003 
Product Classification Dilatation Catheter - Product Code LOX
ProductCORDIS "Dura Star" 2.50 x 10 Dilatation Catheter, Catalog # 70110250, Distributed by Cordis Corporation, Miami Lakes, FL 33014
Code Information Lot #s: 13175336 13180557 13189422 13190774 13208064 13209913 13209921 13211074 13217606 13225515 13245564 13250081 13251126 13252416 13266368 13269603 13272936 13278669 13285514 13289737 13297291 13298192 13300591 13303777 13305476 13309104 and 13314185
Recalling Firm/
Manufacturer
Cordis Corporation
14201 NW 60th Avenue
Miami Lakes FL 33014-2802
For Additional Information Contact
786-313-2000
Manufacturer Reason
for Recall
Slow Deflation or No Deflation
FDA Determined
Cause 2
Process control
ActionCordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.
Quantity in Commerce133,300 units total for recalls Z-0747-0824-2008
DistributionWorldwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.