| Date Initiated by Firm | January 11, 2008 |
| Date Posted | September 18, 2008 |
| Recall Status1 |
Terminated 3 on November 12, 2009 |
| Recall Number | Z-1971-2008 |
| Recall Event ID |
47369 |
| 510(K)Number | K032641 K992815 |
| Product Classification |
Hip joint - Product Code MEH
|
| Product | Margron DTC Hip Replacement System; Margron hip - Femoral stem + suture holes Z+4, sterile, Portland Orthopaedics Inc., St. Clair, MI; REF 1-730-974. Product is used for Orthopedic surgery. |
| Code Information |
All lots. |
| FEI Number |
3003755751
|
Recalling Firm/ Manufacturer |
Portland Orthopaedics Pty, Ltd. 1 South Street, Suite 3, Level 5 Saint George Hospital Kogarah Australia
|
| For Additional Information Contact | 810-3293348 |
Manufacturer Reason for Recall | Australian data indicates the product is associated with a higher than average rate of hip replacement revision surgery. |
FDA Determined Cause 2 | Device Design |
| Action | Stocks were pulled from U.S. consignees on or before 1/11/08. Consignee and implanting surgeon letters dated 6/20/08 were sent advising of the issue and recommending patient monitoring. For assistance, contact Portland Orthopaedics at 1-810-329-3348. |
| Quantity in Commerce | 2403 of all products |
| Distribution | Nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|