Class 2 Device Recall Zyoptix XP Microkeratome tray |
 |
Date Initiated by Firm |
April 23, 2008 |
Date Posted |
October 22, 2008 |
Recall Status1 |
Terminated 3 on April 02, 2012 |
Recall Number |
Z-0104-2009 |
Recall Event ID |
47839 |
510(K)Number |
K040204
|
Product Classification |
Ac-Powered Keratome - Product Code HNO
|
Product |
Zyoptix XP Microkeratome tray, Ref. #ZXP1000, which contains a number of components, one of which is the maxon motor. The label shows the product is manufactured by Bausch & Lomb Incorporated, Rochester, NY. |
Code Information |
Tray/motor serial numbers: 11061/77318906, 10980/77318909, 11019/77318920, 11029/77318925, 11084/77318926, 10969/77504001, 11046/77504002, 11000/77504005, 11088/77504006, 11045/77504007, 11068/77504008, 11108/77504009, 10972/77504011, 11106/77504014, 11021/77504015, 11177/77732302, 11202/77732307, 11091/77732316, 11031/77732318, 11066/77898701, 11026/77898702, 11077/77898704, 11120/77898705, 11103/77898706, 11040/77898707, 11192/77898708, 11190/77898712, 11174/77898713, 11153/77898714, 11030/77898715, 11111/78080317, 11047/78080319, 11150/78080321, 10904/78080323, 11149/78080324, 11096/78080325, 11041/78080333, 11081/78080334, 11064/78080335, 11221/78080339, 11122/78080340, 11034/78080341, 11140/78080342, 11089/78207127 |
Recalling Firm/ Manufacturer |
Bausch & Lomb Inc 1400 North Goodman Street Rochester NY 14609-3547
|
For Additional Information Contact |
Mike Judy 585-338-6000
|
Manufacturer Reason for Recall |
The plastic tip on the drive shaft may come off the shaft and cause the blade to stop oscillating.
|
FDA Determined Cause 2 |
Device Design |
Action |
Phone calls were initiated on 4/28/08 and an Urgent Recall letter dated 4/23/08 was issued 4/28/08 via overnight mail to U.S. customers explaining the reason for recall. The letter also requested that customers stop using the motors identified and return of the motors for evaluation. A return form was enclosed. |
Quantity in Commerce |
44 trays |
Distribution |
Worldwide Distribution --- including USA and countries of Brazil, The Netherlands, Canada, India, Malaysia, Mexico, and Singapore. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HNO and Original Applicant = BAUSCH & LOMB, INC.
|
|
|
|