• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall MULTISTAR D XRay System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall MULTISTAR D XRay Systemsee related information
Date Initiated by FirmAugust 11, 2008
Date PostedNovember 04, 2008
Recall Status1 Terminated 3 on December 21, 2012
Recall NumberZ-0135-2009
Recall Event ID 49426
510(K)NumberK930635 
Product Classification angiographic x-ray system - Product Code IZI
ProductMULTISTAR D X-Ray System, Model number 3772501.
Code Information Model number 3772501. Serial numbers: 1001, 1007, 1008, 1009, 1016, 1017, 1020, 1033, 1040, 1046, 1049, 1050, 1054, 1058, 1060, 1061, 1065, 1070, 1071, 1073, 1076, 1084, 1088, 1089, 1092, 1095, 1096, 1098, 1099, 1100, 1102, 1104, 1106, 1107, 1108, 1111, 1121, 1125, 1130, 1133, 1135, 1141, 1143, 1146, 1147, 1148, 1149, 1150, 1153, 1157, 1158, 1159, 1165, 1166, 1167, 1172, 1175, 1176, 1177, 1178, 1179, 1196, 1197, 1198, 1203, 1204, 1207, 1212, 1213, 1214, 1216, 1236, 1237, 1241, 1249, 1254, 1258, 1262, 1269, 1273, 1279, 1293, 1295, 1299, 1300, 1301, 1508, 1510, 1512, 1514, 1517, 1522, 1525, 1530, 1531, 1533, 1536, 1537, 1538, 1542, 1544, 1547, 1548, 1551, 1554, 1561, 1564, 1569, 1570, 1580, 1585, 1587, 1591, 1594, 1598, 1600, 1602, 1605, 1607, 1608, 1609, 1612, 1613, 1614, 1615, 1618, 1619, 1620, 1621, 1622, 1626, 1628, 1629, 1630, 1633, 1636, 1641, 1643, 1645, 1648, 1653, 1656, 1659, 1660, 1661, 1662, 1665, 1667, 1671, 1672, 1673, 1674, 1676, 1677, 1682, 1683, 1685, 1686, 1693, 1694, 1695, 1697, 1698, 1699, 1701, 1705, 1706, 1709, 1710, 1712, 1715, 1716, 1721, 1722, 1723, 1728, 1731, 1732, 1734, 1735, 1736, 1737, 1739, 1740, 1741, 1742, 1744, 1745, 1746, 1747, 1748, 1752, 1753, 1754, 1755, 1756, 1758, 1757, 1759, 1760, 1766, 1767, 1768, 1770, 1771, 1772, 1774, 1776, 1777, 1778, 1779, 1780, 1781, 1783, 1784, 1786, 1787, 1788, 1790, 1791, 1796, 1797, 1798, 1801, 1804, 1805, 1807, 1808, 1810, 1812, 1813, 1814, 1816, 1817, 1819, 1820, 1821, 1823, 1824, 1825, 1826, 1830, 1831, 1832, 1833, 1841, 1844, 1846, 1848, 1849, 1853, 1854, 1856, 1857, 1858, 1860, 1862, 1865, and 9001.
Recalling Firm/
Manufacturer
Siemens Medical Solutions USA, Inc
51 Valley Stream Pkwy
Malvern PA 19355
Manufacturer Reason
for Recall
Unintended movement of the system table and/or C arm.
FDA Determined
Cause 2
Software design
ActionSiemens Medical Solutions issued a letter to consignees on 8/11/2008 to inform them of the problem and a workaround should the issue occur prior to software revision.
Quantity in Commerce259 units
DistributionNationwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IZI
-
-