| | Class 2 Device Recall Zimmer Silhouette Axial Torque Handle |  |
| Date Initiated by Firm | November 07, 2008 |
| Date Posted | March 04, 2009 |
| Recall Status1 |
Terminated 3 on December 22, 2009 |
| Recall Number | Z-1031-2009 |
| Recall Event ID |
50176 |
| Product Classification |
Wrench - Product Code HXC
|
| Product | Zimmer Silhouette Axial Torque Handle, Zimmer Spine, Inc., Minneapolis, MN; REF 7010-0151-01.
Reusable instrument that is used to apply the appropriate torque to the lock nut that is part of the Silhouette Spinal Fixation System. |
| Code Information |
Lots P050120 and P041466. |
Recalling Firm/ Manufacturer |
Zimmer Inc. 345 E Main St Warsaw IN 46580-2746
|
| For Additional Information Contact | 800-777-7505 |
Manufacturer Reason for Recall | Some of the torque wrenches apply less than the minimum torque to the screw locking cap when the torque indicators are aligned as described in the surgical technique. |
FDA Determined Cause 2 | Process control |
| Action | The firm's sales force was notified by an Urgent: Device Recall letter dated 10/7/08, which instructed them to stop using the instrument and quarantine it immediately; physically count all affected product in their territory and record data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer Spine; and return the affected devices with the completed Inventory Return Certification Form. User accounts/implanting physicians were informed by an Urgent: Device Recall letter dated 12/10/08 and instructed to monitor their patients. |
| Quantity in Commerce | 76 |
| Distribution | Worldwide Distribution --- USA including states of California, Minnesota, Ohio, Pennsylvania, and Tennessee, and countries of Denmark, Germany, Iran, Italy, Russia, Singapore, Spain, Switzerland, and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|