|
Class 2 Device Recall GE Healthcare Signa Ovation 0.35T |
 |
Date Initiated by Firm |
November 03, 2008 |
Date Posted |
August 25, 2009 |
Recall Status1 |
Terminated 3 on May 24, 2012 |
Recall Number |
Z-1722-2009 |
Recall Event ID |
51314 |
510(K)Number |
K053009
|
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product |
GE Healthcare Signa Ovation 0.35T, Model Numbers: Ovation 1-3 (2276937), Ovation 4 (2377062-2, 2377062-5, 2377062-8, or 5118172), and Ovation 5 (5148725).
The 0.35T Signa Ovation with Excite Magnetic Resonance system is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The 0.35T Signa Ovation with Excite Magnetic Resonance system is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the body including, but not limited to, the musculoskeletal, vascular, cardiac, and neuron systems. |
Code Information |
Model Number: 2276937; Serial Numbers: 00000869965YM4, 00000801207YM2, 00000813952YM9, 00000952559YM3, 00000236062MR6, 00000869966YM2, 00000801206YM4, 00000813945YM3, 00000813950YM3, 00000869958YM9, 00000952561YM9, 00000813958YM6, 00000919518YM1, 00000813944YM6, 00000837353YM2, 00000813960YM2, 00000837350YM8, 00000264247MR1, 00000897708YM4, 00000897716YM7, 00000919512YM4, 00000919526YM4, 00000026540VE5, 00000919509YM0, 00000813959YM4, 00000801204YM9, 00000813947YM9, 00000952560YM1, 00000952565YM0, 00000897711YM8, 00000897717YM5, 00000919515YM7, 00000952552YM8, 00000813951YM1, 00000897701YM9, 00000869949YM8, 00000837359YM9, 00000919519YM9, 00000813946YM1, 00000869956YM3, 00000897706YM8, 00000952563YM5, 00000869947YM2, 00000801199YM1, 00000801193YM4, 00000837362YM3, 00000952548YM6, 00000952549YM4, 00000837360YM7, 00000952547YM8, 00000837346YM6, 00000869952YM2, 00000869954YM8, 00000919514YMO, 00000919524YM9, 00000837358YM1, 00000919521YM5, 00000869962YM1, 00000919513YM2, 00000897705YM0, 00000952555YM1, 00000813943YM8, 00000869961YM3, 00000813956YM0, 00000801203YM1, 00000801208YM0, 00000801209YM8, 00000813948YM7, 00000813953YM7, 00000813954YM5, 00000837347YM4, 00000837351YM6, 00000837352YM4, 00000837354YM0, 00000837357YM3, 00000837361YM5, 00000837363YM1, 00000837364YM9, 00000837365YM6, 00000869950YM6, 00000869951YM4, 00000869953YM0, 00000869955YM5, 00000869957YM1, 00000869960YM5, 00000869963YM9, 00000897698YM7, 00000897699YM5, 00000897700YM1, 00000897702YM7, 00000897703YM5, 00000897704YM3, 00000897707YM6, 00000897715YM9, 00000919507YM4, 00000919510YM8, 00000919516YM5, 00000919517YM3, 00000919520YM7, 00000919523YM1, 00000952550YM2, 00000952553YM6, 00000952557YM7, 00000952558YM5, 00000952562YM7, 00000952566YM8, 00000952554YM4, 00000919511YM6, 00000869948YM0, 00000897714YM2, 00000813957YM8, 00000813955YM2, 00000897712YM6, 00000813949YM5; Model Number: 5118172; Serial Numbers; 00000000564YR5, 00000000565YR2, 00000000563YR7, 00000000561YR1; Model Number: 2377062-2; Serial Number: 00000963007YM0; Model Number: 2377062-5; Serial Numbers: 00000000120YR6, 00000000123YR0, 00000000132YR1, 00000000124YR8, 00000000137YR0, 00000000130YR5, 00000000134YR7, 00000000138YR8, 00000962592YM2, 00000000146YR1, 00000000117YR2, 00000000116YR4, 00000000118YR0, 00000943610YM6, 00000000152YR9, 00000000140YR4, 00000000110YR7, 00000000135YR4, 00000000106YR5, 00000000129YR7, 00000000141YR2, 00000000125YR5, 00000000142YR0, 00000000113YR1, 00000000150YR3, 00000000105YR7, 00000000131YR3, 00000000139YR6, 00000000155YR2, 00000000108YR1, 00000000143YR8, 00000000101YR6, 00000000109YR9, 00000000153YR7, 00000951021YM5, 00000000119YR8, 00000000136YR2, 00000000128YR9, 00000000100YR8, 00000000112YR3, 00000000121YR4, 00000000154YR5; Model Number: 2377062-8; Serial Numbers: 00000000776YR5, 00000000792YR2, 00000000793YR0, 00000000801YR1, 00000000803YR7, 00000000829YR2, 00000000836YR7, 00000000758YR3, 00000000765YR8, 00000000755YR9, 00000000756YR7, 00000000774YR0, 00000000784YR9, 00000000785YR6, 00000000798YR9, 00000000827YR6, 00000000816YR9, 00000000779YR9, 00000000828YR4, 00000000789YR8, 00000000838YR3, 00000000773YR2, 00000000788YR0, 00000000796YR3, 00000000763YR3, 00000000766YR6, 00000000767YR4, 00000000787YR2, 00000000790YR6, 00000000800YR3, 00000000834YR2, 00000000840YR9, 00000000830YR0, 00000000832YR6, 00000000786YR4, 00000000764YR1, 00000000775YR7, 00000000818YR5, 00000000768YR2, 00000000757YR5, 00000000760YR9, 00000000762YR5, 00000000783YR1, 00000000809YR4, 00000000782YR3, 00000000812YR8, 00000000826YR8, 00000000835YR9, 00000000770YR8, 00000000771YR6, 00000000825YR0, 00000000759YR1, 00000000821YR9, 00000000841YR7, 00000000831YR8, 00000000808YR6, 00000000823YR5, 00000000824YR3, 00000000772YR4, 00000000810YR2, 00000000837YR5, 00000000769YR0, 00000000815YR1, 00000000833YR4, 00000000804YR5, 00000000778YR1, 00000000780YR7, 00000000795YR5, 00000000799YR7, 00000000805YR2, 00000000807YR8, 00000000819YR3, 00000000839YR1, 00000000811YR0, 00000000761YR7, 00000000777YR3, 00000000781YR5, 00000000791YR4, 00000000797YR1, 00000000794YR8; and Model Number: 5148725; Serial Numbers: 00000154413HM9, 00000176551HM0, 00000139846HM0, 00000192389HM5, 00000177143HM5, 00000193109HM6, 00000148905HM3, 00000147526HM8, 00000138166HM4, 00000000SPI275, 00000188489HM9, 00000167482HM9, 00000169104HM7, 00000171692HM7, 00000157419HM3, 00000266430MR8, 00000155516HM8, 00000155515HM0, 00000160809HM0, 00000175023HM1, 00000150718HM5, 00000162563HM1, 00000164072HM1, 00000165772HM5, 00000172421HM0, 00000172816HM1, 00000172869HM0, 00000173908HM5, 00000177792HM9, 00000178519HM5, 00000179361HM1, 00000180127HM3, 00000180746HM0, 00000182071HM1, 00000182249HM3, 00000182597HM5, 00000183130HM4, 00000187365HM2, 00000188810HM6, 00000189105HM0, 00000191017HM3, 00000191154HM4, 00000196647HM2, 00000200577HM5, 00000265078MR6, 00000157418HM5, 00000174348HM3, 00000189943HM4, 00000181736HM0, 00000146891HM7, 00000158641HM1, 00000147906HM2, 00000169833HM1, 00000170940HM1, 00000190359HM0, 00000191544HM6, 00000193220HM1, 00000193936HM2, 00000195772HM9, 00000144797HM8, 00000150152HM7, 00000170109HM3, 00000181735HM2, 00000197645HM5, 00000168323HM4 and 00000181517HM4. |
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-544-3894
|
Manufacturer Reason for Recall |
GE Healthcare has identified a potential pinch point hazard on Ovation MR Scanner when using CTL Array XL Body Flex and Opened Body coil. A patient's hand may be pinched between the magnet bore ceiling cover and the CTL Array, XL Body Flex, or Open Body coil if the patient puts his/her hand on the top of the coil.
2/20/09: Recall being expanded to include the OpenSpeed Systems.
|
FDA Determined Cause 2 |
Device Design |
Action |
GE Healthcare issued a "Product Safety Notification" letter dated October 23, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology/Cardiology & Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Product Correction and Contact Information for the Signa Ovation 0.35T Systems.
A second GE Healthcare "Urgent Medical Device Correction" letter dated March 5, 2009 was sent 3/9/09 to consignees which addressed the expanded recall to the Signa OpenSpeed0.7T MR Systems.
For further questions, contact your local GE Healthcare sales or service representative. |
Quantity in Commerce |
308 (143 US / 142 OUS / 23 TBD) |
Distribution |
Worldwide Distribution-- US (including states of AL, ALZ, AR, CA, CO,CT, DE, FL, FA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MT, MO, NE,NH, NV, NJ, NY, NC, OH, OK, OR, PA, PR, SC SD, TN, TX, VA, WA, WV and WI), CHINA, CYPRUS, ARGENTINA, BRAZIL EGYPT, DENMARK, FEDERAL REPUBLIC OF GERMANY, FEDERAL REPUBLIC OF FRANCE, GET 7 D CHINA, GREECE, HUNGARY, ITALY, IVORY COAST, JAPAN, KOREA, KUWAIT, MALAWI, MEXICO, OMAN, PANAMA, PHILIPPINES, RUSSIAN FEDERATION, SAUDI ARABIA, SPAIN , SWITZERLAND, TURKEY, UKRAINE, UNITED ARAB EMERITUS, UNITED KINGDOM and VENEZUELA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LNH and Original Applicant = GE MEDICAL SYSTEMS
|
|
|
|