| Class 2 Device Recall Stryker MedSurg bed | |
Date Initiated by Firm | August 10, 2009 |
Date Posted | September 18, 2009 |
Recall Status1 |
Terminated 3 on April 25, 2011 |
Recall Number | Z-2218-2009 |
Recall Event ID |
52230 |
Product Classification |
Bed, ac-powered adjustable hospital - Product Code FNL
|
Product | Stryker MedSurg bed, 115v, Stryker Medical, Portage, MI; Model 3002 S3.
Beds are A/C powered hospital beds that are intended for medical purposes within the hospital setting. |
Code Information |
Note these serial numbers are inclusive to models 3002, 3002 S3 and 3221. Serial Numbers: 080115984, 080115985, 080115986, 080115987, 080115988, 080115989, 080115990, 080115991, 080115992, 080115993, 080115994, 080115995, 080115996, 080115997, 080115998, 080115999, 080116000, 080116001, 080116002, 080116003, 080116004, 080116005, 080116006, 080116007, 080116008, 080116009, 080116010, 080116011, 080116012, 080116013, 080116014, 080116015, 080116016, 080116017, 080116018, 080116019, 080116020, 080116021, 080116022, 080116023, 080116024, 080116025, 080116026, 080116027, 080116028, 080116137, 080116138, 080116139, 080116140, 080116141, 080116142, 080116143, 080116144, 080116145, 080116146, 080116147, 080116148, 080116149, 080116150, 080116151, 080116152, 080116153, 080116154, 080116155, 080116156, 080116157, 080116158, 080116159, 080116160, 080116161, 080116162, 080116163, 080116164, 080116165, 080116166, 080116167, 080116168, 080116169, 080116170, 080116171, 080116172, 080116173, 080116174, 080116175, 080116176, 080116177, 080116178, 080116179, 080116180, 080116181, 080116319, 080116320, 080116321, 080116322, 080116323, 080116588, 080116614, 080116615, 080116637, 080116638, 080116639, 080116640, 080116641, 080116642, 080116643, 080116644, 080116645, 080116646, 080116647, 080116648, 080116649, 080116650, 080116651, 080116652, 080116653, 080116654, 080116655, 080116656, 080116657, 080116658, 080116659, 080116660, 080116661, 080116662, 080116663, 080116664, 080116665, 080116666, 080116667, 080116668, 080116669, 080116670, 080116671, 080116672, 080116673, 080116674, 080116675, 080116676, 080116677, 080116678, 080116679, 080116680, 080116681, 080116682, 080116683, 080116684, 080116685, 080116686, 080116687, 080116688, 080116689, 080116690, 080116691, 080116692, 080116693, 080116694, 080116695, 080116696, 080116697, 080116698, 080116699, 080116700, 080116701, 080116702, 080116703, 080116704, 080116705, 080116706, 080116707, 080116708, 080116709, 080116710, 080116711, 080116712, 080116713, 080116714, 080116715, 080116716, 080116717, 080116718, 080116719, 080116720, 080116721, 080116722, 080116723, 080116724, 080116725, 080116726, 080116727, 080116728, 080116729, 080116730, 080116731, 080116732, 080116733, 080116734, 080116735, 080116736, 080116737, 080116738, 080116739, 080116740, 080116741, 080116742, 080116743, 080116744, 080116745, 080116746, 080116747, 080116748, 080116749, 080116750, 080116751, 080116752, 080116753, 080116754, 080116755, 080116756, 080116757, 080116758, 080116759, 080116760, 080116761, 080116762, 080116763, 080116764, 080116765, 080116766, 080116767, 080116768, 080116769, 080116770, 080116771, 080116772, 080116773, 080116774, 080116775, 080116776, 080116777, 080116778, 080116779, 080116780, 080116781, 080116782, 080116783, 080116784, 080116785, 080116786, 080116787, 080116788, 080116789, 080116790, 080116791, 080116792, 080116793, 080116794, 080116795, 080116796, 080116797, 080116798, 080116799, 080116800, 080116801, 080116802, 080116803, 080116804, 080116805, 080116806, 080116807, 080116808, 080116809, 080116810, 080116811, 080116812, 080116813, 080116814, 080116815, 080116816, 080116817, 080116818, 080116819, 080116820, 080116821, 080116822, 080116823, 080116824, 080116825, 080116826, 080116827, 080116828, 080116829, 080116830, 080116831, 080116832, 080116833, 080116834, 080116835, 080116836, 080116837, 080116838, 080116839, 080116840, 080116841, 080116842, 080116843, 080116844, 080116845, 080116846, 080116847, 080116848, 080116849, 080116850, 080116851, 080116852, 080116853, 080116854, 080116855, 080116856, 080116857, 080116858, 080116859, 080116860, 080116861, 080116862, 080116863, 080116864, 080116865, 080116866, 080116867, 080116868, 080116869, 080116870, 080116871, 080116872, 080116873, 080116874, 080116875, 080116876, 080116877, 080116878, 080116879, 080116880, 080116881, 080116882, 080116883, 080116884, 080116885, 080116886, 080116887, 080116888, 080116889, 080116890, 080116891, 080116892, 080116893, 080116894, 080116895, 080116896, 080116897, 080116898, 080116899, 080116900, 080116901, 080116902, 080116903, 080116904, 080116905, 080116906, 080116907, 080116908, 080116909, 080116910, 080116911, 080116912, 080116913, 080116914, 080116915, 080116916, 080116917, 080116918, 080116919, 080116920, 080116921, 080116922, 080116923, 080116924, 080116925, 080116926, 080116927, 080116928, 080116929, 080116930, 080116931, 080116932, 080116933, 080116934, 080116935, 080116936, 080116937, 080116938, 080116939, 080116940, 080116941, 080116942, 080116943, 080116944, 080116945, 080116946, 080116947, 080116948, 080116949, 080116950, 080116951, 080116952, 080116953, 080116954, 080116955, 080116956, 080116957, 080116958, 080116959, 080116960, 080116961, 080116962, 080116963, 080116964, 080116965, 080116966, 080116967, 080116968, 080116969, 080116970, 080116971, 080116972, 080116973, 080116974, 080116975, 080116976, 080116977, 080116978, 080116979, 080116980, 080116981, 080116982, 080116983, 080116984, 080116985, 080116986, 080116987, 080116988, 080116989, 080116990, 080116991, 080116992, 080116993, 080116994, 080116995, 080116996, 080116997, 080116998, 080116999, 080117000, 080117001, 080117002, 080117003, 080117004, 080117005, 080117006, 080117007, 080117008, 080117009, 080117010, 080117011, 080117012, 080117013, 080117014, 080117015, 080117016, 080117017, 080117018, 080117019, 080117020, 080117021, 080117022, 080117023, 080117066, 080215001, 080215002, 080215003, 080215004, 080215005, 080215006, 080215007, 080215008, 080215009, 080215010, 080215011, 080215012, 080215013, 080215014, 080215015, 080215016, 080215017, 080215018, 080215019, 080215020, 080215021, 080215022, 080215023, 080215024, 080215025, 080215026, 080215027, 080215028, 080215029, 080215030, 080215031, 080215032, 080215033, 080215034, 080215035, 080215036, 080215037, 080215038, 080215039, 080215040, 080215041, 080215042, 080215043, 080215044, 080215045, 080215046, 080215047, 080215048, 080215049, 080215050, 080215051, 080215052, 080215053, 080215054, 080215055, 080215056, 080215057, 080215058, 080215059, 080215060, 080215061, 080215062, 080215063, 080215064, 080215065, 080215066, 080215067, 080215068, 080215069, 080215070, 080215071, 080215072, 080215073, 080215074, 080215075, 080215076, 080215077, 080215078, 080215079, 080215080, 080215081, 080215082, 080215083, 080215084, 080215085, 080215086, 080215087, 080215088, 080215089, 080215090, 080215091, 080215092, 080215093, 080215094, 080215095, 080215096, 080215097, 080215098, 080215099, 080215100, 080215101, 080215102, 080215103, 080215104, 080215105, 080215106, 080215107, 080215108, 080215109, 080215110, 080215111, 080215112, 080215113, 080215114, 080215115, 080215116, 080215117, 080215118, 080215119, 080215120, 080215121, 080215122, 080215123, 080215124, 080215125, 080215126, 080215127, 080215128, 080215129, 080215130, 080215131, 080215132, 080215133, 080215134, 080215135, 080215136, 080215137, 080215138, 080215139, 080215140, 080215141, 080215142, 080215143, 080215144, 080215145, 080215146, 080215147, 080215148, 080215149, 080215150, 080215151, 080215152, 080215153, 080215154, 080215155, 080215156, 080215157, 080215158, 080215159, 080215160, 080215161, 080215162, 080215163, 080215164, 080215165, 080215166, 080215167, 080215168, 080215169, 080215170, 080215171, 080215172, 080215173, 080215174, 080215175, 080215176, 080215177, 080215178, 080215179, 080215180, 080215181, 080215182, 080215183, 080215184, 080215185, 080215186, 080215187, 080215188, 080215189, 080215190, 080215191, 080215192, 080215193, 080215194, 080215195, 080215196, 080215197, 080215198, 080215199, 080215200, 080215201, 080215202, 080215203, 080215204, 080215205, 080215206, 080215207, 080215208, 080215209, 080215210, 080215211, 080215212, 080215213, 080215214, 080215215, 080215216, 080215217, 080215218, 080215219, 080215220, 080215221, 080215222, 080215223, 080215224, 080215225, 080215226, 080215227, 080215228, 080215229, 080215230, 080215231, 080215232, 080215233.080215234, 080215235, 080215236, 080215237, 080215238, 080215239, 080215240, 080215241, 080215242, 080215243, 080215244, 080215245, 080215246, 080215247, 080215248, 080215249, 080215250, 080215251, 080215252, 080215253, 080215254, 080215255, 080215256, 080215257, 080215258, 080215259, 080215260, 080215261, 080215262, 080215263, 080215264, 080215265, 080215266, 080215267, 080215268, 080215269, 080215270, 080215271, 080215272, 080215273, 080215274, 080215275, 080215276, 080215277, 080215278, 080215279, 080215280, 080215281, 080215282, 080215283, 080215284, 080215285, 080215286, 080215287, 080215288, 080215289, 080215290, 080215291, 080215292, 080215293, 080215294, 080215295, 080215296, 080215297, 080215298, 080215299, 080215300, 080215301, 080215302, 080215303, 080215304, 080215305, 080215306, 080215307, 080215308, 080215309, 080215310, 080215311, 080215312, 080215313, 080215314, 080215315, 080215316, 080215317, 080215318, 080215319, 080215320, 080215321, 080215322, 080215323, 080215324, 080215325, 080215326, 080215327, 080215328, 080215329, 080215330, 080215331, 080215332, 080215333, 080215334, 080215335, 080215336, 080215337, 080215338, 080215339, 080215340, 080215341, 080215342, 080215343, 080215344, 080215345, 080215346, 080215347, 080215348, 080215349, 080215350, 080215351, 080215352, 080215353, 080215354, 080215355, 080215356, 080215357, 080215358, 080215359, 080215360, 080215361, 080215362, 080215363, 080215364, 080215365, 080215366, 080215367, 080215368, 080215369, 080215370, 080215371, 080215372, 080215373, 080215374, 080215375, 080215376, 080215377, 080215378, 080215379, 080215380, 080215381, 080215382, 080215383, 080215384, 080215385, 080215386, 080215387, 080215388, 080215389, 080215390, 080215391, 080215392, 080215393, 080215394, 080215395, 080215396, 080215397, 080215398, 080215399, 080215400, 080215401, 080215402, 080215403, 080215404, 080215405, 080215406, 080215407, 080215408, 080215409, 080215410, 080215411, 080215412, 080215413, 080215414, 080215415, 080215416, 080215417, 080215418, 080215419, 080215420, 080215421, 080215422, 080215423, 080215424, 080215425, 080215426, 080215427, 080215428, 080215429, 080215430, 080215431, 080215432, 080215433, 080215434, 080215435, 080215436, 080215437, 080215438, 080215439, 080215440, 080215441, 080215442, 080215443, 080215444, 080215445, 080215446, 080215447, 080215448, 080215449, 080215450, 080215451, 080215452, 080215453, 080215454, 080215455, 080215456, 080215457, 080215458, 080215459, 080215460, 080215461, 080215462, 080215463, 080215464, 080215465, 080215466, 080215467, 080215468, 080215469, 080215470, 080215471, 080215472, 080215473, 080215474, 080215475, 080215476, 080215477, 080215478, 080215479, 080215480, 080215481, 080215482, 080215483, 080215484, 080215485, 080215486, 080215487, 080215488, 080215489, 080215490, 080215491, 080215492, 080215493, 080215494, 080215495, 080215496, 080215497, 080215498, 080215499, 080215500, 080215501, 080215502, 080215503, 080215504, 080215505, 080215506, 080215507, 080215508, 080215509, 080215510, 080215511, 080215512, 080215513, 080215514, 080215515, 080215516, 080215517, 080215518, 080215519, 080215520, 080215521, 080215522, 080215523, 080215524, 080215525, 080215526, 080215527, 080215528, 080215529, 080215530, 080215531, 080215532, 080215533, 080215534, 080215535, 080215536, 080215537, 080215538, 080215539, 080215540, 080215541, 080215542, 080215543, 080215544, 080215545, 080215546, 080215547, 080215548, 080215549, 080215550, 080215551, 080215552, 080215553, 080215554, 080215555, 080215556, 080215557, 080215558, 080215559, 080215560, 080215561, 080215562, 080215563, 080215564, 080215565, 080215566, 080215567, 080215568, 080215569, 080215570, 080215571, 080215572, 080215573, 080215574, 080215575, 080215576, 080215577, 080215578, 080215579, 080215580, 080215581, 080215582, 080215583, 080215584, 080215585, 080215586, 080215587, 080215588, 080215589, 080215590, 080215591, 080215592, 080215593, 080215594, 080215595, 080215596, 080215597, 080215598, 080215599, 080215600, 080215601, 080215602, 080215603, 080215604, 080215605, 080215606, 080215607, 080215608, 080215609, 080215610, 080215611, 080215612, 080215613, 080215614, 080215615, 080215616, 080215617, 080215618, 080215619, 080215620, 080215621, 080215622, 080215623, 080215624, 080215625, 080215626, 080215627, 080215628, 080215629, 080215630, 080215631, 080215632, 080215633, 080215634, 080215635, 080215636, 080215637, 080215638, 080215639, 080215640, 080215641, 080215642, 080215643, 080215644, 080215645, 080215646, 080215647, 080215648, 080215649, 080215650, 080215651, 080215652, 080215653, 080215654, 080215655, 080215656, 080215657, 080215658, 080215659, 080215660, 080215661, 080215662, 080215663, 080215664, 080215665, 080215666, 080215667, 080215668, 080215669, 080215670, 080215671, 080215672, 080215673, 080215674, 080215675, 080215676, 080215677, 080215678, 080215679, 080215680, 080215681, 080215682, 080215683, 080215684, 080215685, 080215686, 080215687, 080215688, 080215689, 080215690, 080215691, 080215692, 080215693, 080215695, 080215696, 080215697, 080215698, 080215699, 080215700, 080215701, 080215702, 080215703, 080215704, 080215705, 080215706, 080215707, 080215708, 080215709, 080215710, 080215711, 080215712, 080215713, 080215714, 080215715, 080215716, 080215717, 080215718, 080215719, 080215720, 080215721, 080215722, 080215723, 080215724, 080215725, 080215726, 080215727, 080215728, 080215729, 080215730, 080215756, 080215771, 080215772, 080215773, 080215774, 080215775, 080215776, 080215777, 080215778, 080215779, 080215780, 080215781, 080215782, 080215783, 080215784, 080215785, 080215786, 080215787, 080215788, 080215789, 080215790, 080215791, 080215792, 080215793, 080215794, 080215795, 080215796, 080215797, 080215798, 080215799, 080215800, 080215801, 080215802, 080215803, 080215804, 080215805, 080215806, 080215807, 080215808, 080215809, 080215810, 080215811, 080215812, 080215813, 080215814, 080215815, 080215816, 080215817, 080215818, 080215819, 080215820, 080215821, 080215822, 080215823, 080215824, 080215825, 080215826, 080215827, 080215828, 080215829, 080215830, 080215831, 080215832, 080215833, 080215834, 080215835, 080215836, 080215837, 080215838, 080215839, 080215840, 080215841, 080215842, 080215843, 080215844, 080215845, 080215846, 080215847, 080215848, 080215849, 080215850, 080215851, 080215852, 080215853, 080215854, 080215855, 080215856, 080215857, 080215858, 080215859, 080215860, 080215861, 080215862, 080215863, 080215864, 080215865, 080215866, 080215867, 080215868, 080215869, 080215870, 080215871, 080215872, 080215873, 080215874, 080215875, 080215876, 080215877, 080215878, 080215879, 080215880, 080215881, 080215882, 080215883, 080215884, 080215885, 080215886, 080215887, 080215888, 080215889, 080215890, 080215891, 080215892, 080215893, 080215894, 080215895, 080215896, 080215897, 080215898, 080215899, 080215900, 080215901, 080215902, 080215903, 080215904, 080215905, 080215906, 080215907, 080215908, 080215909, 080215910, 080215911, 080215912, 080215913, 080215914, 080215915, 080215916, 080215917, 080215918, 080215919, 080215920, 080215921, 080215922, 080215923, 080215924, 080215925, 080215926, 080215927, 080215928, 080215929, 080215930, 080215931, 080215932, 080215933, 080215934, 080215935, 080215936, 080215937, 080215938, 080215939, 080215940, 080215941, 080215942, 080215943, 080215944, 080215945, 080215946, 080215947, 080215948, 080215949, 080215950, 080215951, 080215952, 080215953, 080215954, 080215955, 080215956, 080215957, 080215958, 080215959, 080215960, 080215961, 080215962, 080215963, 080215964, 080215965, 080215966, 080215967, 080215968, 080215969, 080215970, 080215971, 080215972, 080215973, 080215974, 080215975, 080215976, 080215977, 080215978, 080215979, 080215980, 080215981, 080215982, 080215983, 080215984, 080215985, 080215986, 080215987, 080215988, 080215989, 080215990, 080215991, 080215992, 080215993, 080215994, 080215995, 080215996, 080215997, 080215998, 080216001, 080216002, 080216003, 080216004, 080216005, 080216006, 080216007, 080216008, 080216009, 080216010, 080216011, 080216012, 080216013, 080216014, 080216015, 080216016, 080216017, 080216018, 080216019, 080216020, 080216021, 080216022, 080216023, 080216024, 080216025, 080216026, 080216027, 080216028, 080216029, 080216030, 080216031, 080216032, 080216033, 080216074, 080216075, 080216076, 080216077, 080216078, 080216079, 080216080, 080216081, 080216082, 080216083, 080216084, 080216085, 080216086, 080216087, 080216088, 080216089, 080216090, 080216091, 080216092, 080216093, 080216094, 080216095, 080216096, 080216097, 080216098, 080216099, 080216100, 080216101, 080216102, 080216103, 080216104, 080216105, 080216106, 080216107, 080216108, 080216109, 080216110, 080216111, 080216112, 080216113, 080216114, 080216115, 080216116, 080216117, 080216118, 080216119, 080216120, 080216121, 080216122, 080216123, 080216124, 080216125, 080216126, 080216127, 080216128, 080216129, 080216130, 080216131, 080216132, 080216133, 080216134, 080216135, 080216136, 080216137, 080216138, 080216139, 080216140, 080216141, 080216142, 080216143, 080216144, 080216145, 080216146, 080216147, 080216148, 080216149, 080216150, 080216151, 080216152, 080216153, 080216154, 080216155, 080216156, 080216157, 080216158, 080216159, 080216160, 080216161, 080216162, 080216163, 080216164, 080216165, 080216166, 080216167, 080216168, 080216169, 080216170, 080216171, 080216172, 080216173, 080216174, 080216175, 080216176, 080216177, 080216178, 080216179, 080216180, 080216181, 080216182, 080216183, 080216184, 080216185, 080216186, 080216187, 080216188, 080216189, 080216210, 080216211, 080216212, 080216213, 080216214, 080216215, 080216216, 080216217, 080216218, 080216219, 080216220, 080216221, 080216222, 080216223, 080216224, 080216225, 080216226, 080216227, 080216228, 080216229, 080216230, 080216231, 080216232, 080216233, 080216234, 080216235, 080216236, 080216237, 080216238, 080216239, 080216240, 080216241, 080216242, 080216243, 080216244, 080216245, 080216246, 080216247, 080216248, 080216249, 080216250, 080216251, 080216252, 080216253, 080216254, 080216255, 080216256, 080216257, 080216272, 080216501, 080216502, 080216503, 080216504, 080216505, 080216506, 080216507, 080216508, 080216509, 080216510, 080216511, 080216512, 080216513, 080216514, 080216515, 080216516, 080216517, 080216518, 080216519, 080216520, 080216521, 080216522, 080216523, 080216524, 080216525, 080216526, 080216527, 080216528, 080216529, 080216530, 080216531, 080216532, 080216533, 080216534, 080216535, 080216536, 080216537, 080216538, 080216539, 080216540, 080216541, 080216542, 080216543, 080216544, 080216545, 080216546, 080216547, 080216548, 080216549, 080216550, 080216551, 080216552, 080216553, 080216554, 080216555, 080216556, 080216557, 080216558, 080216559, 080216560, 080216561, 080216562, 080216563, 080216564, 080216565, 080216566, 080216567, 080216568, 080216569, 080216570, 080216571, 080216572, 080216573, 080216574, 080216575, 080216576, 080216577, 080216578, 080216579, 080216580, 080216581, 080216582, 080216583, 080216584, 080216585, 080216586, 080216587, 080216588, 080216589, 080216590, 080216591, 080216592, 080216593, 080216594, 080216595, 080216596, 080216597, 080216598, 080216599, 080216600, 080216601, 080216602, 080216603, 080216604, 080216605, 080216606, 080216607, 080216608, 080216609, 080216610, 080216611, 080216612, 080216613, 080216614, 080216615, 080216616, 080216617, 080216618, 080216619, 080216620, 080216621, 080216622, 080216623, 080216624, 080216625, 080216626, 080216627, 080216628, 080216629, 080216630, 080216631, 080216632, 080216633, 080216634, 080216635, 080216636, 080216637, 080216638, 080216639, 080216640, 080216641, 080216642, 080216643, 080216644, 080216645, 080216646, 080216647, 080216648, 080216649, 080216650, 080216651, 080216652, 080216653, 080216654, 080216655, 080216656, 080216657, 080216658, 080216659, 080216660, 080216661, 080216662, 080216663, 080216664, 080216665, 080216666, 080216667, 080216668, 080216669, 080216670, 080216671, 080216672, 080216673, 080216676, 080216677, 080216678, 080216679, 080216680, 080216681, 080216682, 080216683, 080216684, 080216685, 080216686, 080216687, 080216688, 080216689, 080216690, 080216691, 080216692, 080216693, 080216694, 080216695, 080216696, 080216697, 080216698, 080216699, 080216700, 080216701, 080216702, 080216703, 080216704, 080216705, 080216706, 080216707, 080216708, 080216709, 080216710, 080216711, 080216712, 080216713, 080216714, 080216715, 080216716, 080216717, 080216718, 080216719, 080216720, 080216721, 080216722, 080216723, 080216724, 080216725, 080216726, 080216727, 080216728, 080216729, 080216730, 080216731, 080216732, 080216733, 080216734, 080216735, 080216736, 080216737, 080216738, 080216739, 080216740, 080216741, 080216742, 080216743, 080216744, 080216745, 080216746, 080216747, 080216748, 080216749, 080216750, 080216751, 080216752, 080216753, 080216754, 080216755, 080216756, 080216757, 080216758, 080216759, 080216760, 080216761, 080216762, 080216763, 080216764, 080216765, 080216766, 080216767, 080216768, 080216769, 080216770, 080216771, 080216772, 080216773, 080216774, 080216775, 080216776, 080216777, 080216778, 080216779, 080216780, 080216781, 080216782, 080216783, 080216784, 080216785, 080216786, 080216787, 080216788, 080216789, 080216790, 080216791, 080216792, 080216793, 080216794, 080216795, 080216796, 080216797, 080216798, 080216799, 080216800, 080216801, 080216802, 080216803, 080216804, 080216805, 080216806, 080216807, 080216808, 080216809, 080216810, 080216811, 080216812, 080216813, 080216814, 080216815, 080216816, 080216817, 080216818, 080216819, 080216820, 080216821, 080216822, 080216823, 080216824, 080216825, 080216826, 080216827, 080216828, 080216829, 080216830, 080216831, 080216832, 080216833, 080216834, 080216835, 080216836, 080216837, 080216838, 080216839, 080216840, 080216841, 080216842, 080216843, 080216844, 080216845, 080216846, 080216847, 080216848, 080216849, 080216850, 080216851, 080216852, 080216853, 080216854, 080216855, 080216856, 080216857, 080216858, 080216859, 080216860, 080216861, 080216862, 080216863, 080216864, 080216865, 080216866, 080216867, 080216868, 080216869, 080216870, 080216871, 080216872, 080216873, 080216874, 080216875, 080216876, 080216877, 080216878, 080216879, 080216880, 080216881, 080216882, 080216883, 080216884, 080216885, 080216886, 080216887, 080216888, 080216889, 080216890, 080216891, 080216892, 080216893, 080216894, 080216895, 080216896, 080216897, 080216898, 080216899, 080216900, 080216901, 080216902, 080216903, 080216904, 080216905, 080216906, 080216907, 080216908, 080216909, 080216910, 080216911, 080216912, 080216913, 080216914, 080216915, 080216916, 080216917, 080216918, 080216919, 080216920, 080216921, 080216922, 080216923, 080216924, 080216925, 080216926, 080216927, 080216928, 080216929, 080216930, 080216931, 080216932, 080216933, 080216934, 080216935, 080216936, 080216937, 080216938, 080216939, 080216940, 080216941, 080216942, 080216943, 080216944, 080216945, 080216946, 080216947, 080216948, 080216949, 080216950, 080216951, 080216952, 080216953, 080216954, 080216955, 080216956, 080216957, 080216958, 080216959, 080216960, 080216961, 080216962, 080216963, 080216964, 080216965, 080216966, 080216967, 080216968, 080216969, 080216970, 080216971, 080216972, 080216973, 080216974, 080216975, 080216976, 080216977, 080216978, 080216979, 080216980, 080216981, 080216982, 080216983, 080216984, 080216985, 080216986, 080216987, 080216988, 080216989, 080216990, 080216991, 080216992, 080216993, 080216994, 080216995, 080216996, 080216997, 080216998, 080216999, 080217000, 080315001, 080315002, 080315003, 080315004, 080315005, 080315006, 080315007, 080315008, 080315009, 080315010, 080315011, 080315012, 080315013, 080315014, 080315015, 080315016, 080315017, 080315018, 080315019, 080315020, 080315021, 080315022, 080315023, 080315024, 080315025, 080315026, 080315027, 080315028, 080315029, 080315030, 080315031, 080315032, 080315033, 080315034, 080315035, 080315036, 080315037, 080315038, 080315039, 080315040, 080315041, 080315042, 080315043, 080315044, 080315045, 080315046, 080315047, 080315048, 080315049, 080315050, 080315051, 080315052, 080315053, 080315054, 080315055, 080315056, 080315057, 080315058, 080315059, 080315060, 080315061, 080315062, 080315063, 080315064, 080315065, 080315066, 080315067, 080315068, 080315069, 080315070, 080315071, 080315072, 080315073, 080315074, 080315075, 080315076, 080315077, 080315078, 080315079, 080315080, 080315081, 080315082, 080315083, 080315084, 080315085, 080315086, 080315087, 080315088, 080315089, 080315090, 080315091, 080315092, 080315093, 080315094, 080315095, 080315096, 080315097, 080315098, 080315099, 080315100, 080315101, 080315102, 080315103, 080315104, 080315105, 080315106, 080315107, 080315108, 080315109, 080315110, 080315111, 080315112, 080315113, 080315114, 080315115, 080315116, 080315117, 080315118, 080315119, 080315120, 080315121, 080315122, 080315123, 080315124, 080315125, 080315126, 080315127, 080315128, 080315129, 080315130, 080315131, 080315132, 080315133, 080315134, 080315135, 080315136, 080315137, 080315138, 080315139, 080315140, 080315141, 080315142, 080315143, 080315144, 080315145, 080315146, 080315147, 080315148, 080315149, 080315150, 080315151, 080315152, 080315153, 080315154, 080315155, 080315156, 080315157, 080315158, 080315159, 080315160, 080315161, 080315162, 080315163, 080315164, 080315165, 080315166, 080315167, 080315168, 080315169, 080315170, 080315171, 080315172, 080315173, 080315174, 080315175, 080315176, 080315177, 080315178, 080315179, 080315180, 080315181, 080315182, 080315183, 080315184, 080315185, 080315186, 080315187, 080315188, 080315189, 080315190, 080315191, 080315192, 080315193, 080315194, 080315195, 080315196, 080315197, 080315198, 080315199, 080315200, 080315201, 080315202, 080315203, 080315204, 080315205, 080315206, 080315207, 080315208, 080315209, 080315210, 080315211, 080315212, 080315213, 080315214, 080315215, 080315216, 080315217, 080315218, 080315219, 080315220, 080315221, 080315222, 080315223, 080315224, 080315225, 080315226, 080315227, 080315228, 080315229, 080315230, 080315231, 080315232, 080315233, 080315234, 080315235, 080315236, 080315237, 080315238, 080315239, 080315240, 080315241, 080315242, 080315243, 080315244, 080315245, 080315246, 080315247, 080315248, 080315249, 080315250, 080315251, 080315252, 080315253, 080315254, 080315255, 080315256, 080315257, 080315258, 080315259, 080315260, 080315261, 080315262, 080315263, 080315264, 080315265, 080315266, 080315267, 080315268, 080315269, 080315270, 080315271, 080315272, 080315273, 080315274, 080315275, 080315276, 080315277, 080315278, 080315279, 080315280, 080315281, 080315282, 080315283, 080315284, 080315285, 080315286, 080315287, 080315288, 080315289, 080315290, 080315291, 080315292, 080315293, 080315294, 080315295, 080315296, 080315297, 080315298, 080315299, 080315300, 080315301, 080315302, 080315303, 080315304, 080315305, 080315306, 080315307, 080315308, 080315309, 080315310, 080315311, 080315312, 080315313, 080315314, 080315315, 080315316, 080315317, 080315318, 080315319, 080315320, 080315321, 080315322, 080315323, 080315324, 080315325, 080315326, 080315327, 080315328, 080315329, 080315330, 080315331, 080315332, 080315333, 080315334, 080315335, 080315336, 080315337, 080315338, 080315339, 080315340, 080315341, 080315342, 080315343, 080315344, 080315345, 080315346, 080315347, 080315348, 080315349, 080315350, 080315351, 080315352, 080315353, 080315354, 080315355, 080315356, 080315357, 080315358, 080315359, 080315360, 080315361, 080315362, 080315363, 080315364, 080315365, 080315366, 080315367, 080315368, 080315369, 080315370, 080315371, 080315372, 080315373, 080315374, 080315375, 080315376, 080315377, 080315378, 080315379, 080315380, 080315381, 080315382, 080315383, 080315384, 080315385, 080315386, 080315387, 080315388, 080315389, 080315390, 080315391, 080315392, 080315393, 080315394, 080315395, 080315396, 080315397, 080315398, 080315399, 080315400, 080315401, 080315402, 080315403, 080315404, 080315405, 080315406, 080315407, 080315408, 080315409, 080315410, 080315411, 080315412, 080315413, 080315414, 080315415, 080315416, 080315417, 080315418, 080315419, 080315420, 080315421, 080315422, 080315423, 080315424, 080315425, 080315426, 080315427, 080315428, 080315429, 080315430, 080315431, 080315432, 080315433, 080315434, 080315435, 080315436, 080315437, 080315438, 080315439, 080315440, 080315441, 080315442, 080315443, 080315444, 080315445, 080315446, 080315447, 080315448, 080315449, 080315450, 080315451, 080315452, 080315453, 080315454, 080315455, 080315456, 080315457, 080315458, 080315459, 080315460, 080315461, 080315462, 080315463, 080315464, 080315465, 080315466, 080315467, 080315468, 080315469, 080315470, 080315471, 080315472, 080315473, 080315474, 080315475, 080315476, 080315477, 080315478, 080315479, 080315480, 080315481, 080315482, 080315483, 080315484, 080315485, 080315486, 080315487, 080315488, 080315489, 080315490, 080315491, 080315492, 080315493, 080315494, 080315495, 080315496, 080315497, 080315498, 080315499, 080315500, 080315501, 080315502, 080315503, 080315504, 080315505, 080315506, 080315507, 080315508, 080315509, 080315510, 080315511, 080315512, 080315513, 080315514, 080315515, 080315516, 080315517, 080315518, 080315519, 080315520, 080315521, 080315522, 080315523, 080315524, 080315525, 080315526, 080315527, 080315528, 080315529, 080315530, 080315531, 080315532, 080315533, 080315534, 080315535, 080315536, 080315537, 080315538, 080315539, 080315540, 080315541, 080315542, 080315543, 080315544, 080315545, 080315546, 080315547, 080315548, 080315549, 080315550, 080315551, 080315552, 080315553, 080315554, 080315555, 080315556, 080315557, 080315558, 080315559, 080315560, 080315561, 080315562, 080315563, 080315564, 080315565, 080315566, 080315567, 080315568, 080315569, 080315570, 080315571, 080315572, 080315573, 080315574, 080315575, 080315576, 080315577, 080315578, 080315579, 080315580, 080315581, 080315582, 080315583, 080315584, 080315585, 080315586, 080315587, 080315588, 080315589, 080315590, 080315591, 080315592, 080315593, 080315594, 080315595, 080315596, 080315597, 080315598, 080315599, 080315600, 080315601, 080315602, 080315603, 080315604, 080315605, 080315606, 080315607, 080315608, 080315609, 080315610, 080315611, 080315612, 080315613, 080315614, 080315615, 080315616, 080315617, 080315618, 080315619, 080315620, 080315621, 080315622, 080315623, 080315624, 080315625, 080315626, 080315627, 080315628, 080315629, 080315630, 080315631, 080315632, 080315633, 080315634, 080315635, 080315636, 080315637, 080315638, 080315639, 080315640, 080315641, 080315642, 080315643, 080315644, 080315645, 080315646, 080315647, 080315648, 080315649, 080315650, 080315651, 080315652, 080315653, 080315654, 080315655, 080315656, 080315657, 080315658, 080315659, 080315660, 080315661, 080315662, 080315663, 080315664, 080315665, 080315666, 080315667, 080315668, 080315669, 080315670, 080315671, 080315672, 080315673, 080315674, 080315675, 080315676, 080315677, 080315678, 080315679, 080315680, 080315681, 080315682, 080315683, 080315684, 080315685, 080315686, 080315687, 080315688, 080315689, 080315690, 080315691, 080315692, 080315693, 080315694, 080315695, 080315696, 080315697, 080315698, 080315699, 080315700, 080315701, 080315702, 080315703, 080315704, 080315705, 080315706, 080315707, 080315708, 080315709, 080315710, 080315711, 080315712, 080315713, 080315714, 080315715, 080315716, 080315717, 080315718, 080315719, 080315720, 080315721, 080315722, 080315723, 080315724, 080315725, 080315726, 080315727, 080315728, 080315729, 080315730, 080315731, 080315732, 080315733, 080315734, 080315735, 080315736, 080315737, 080315738, 080315739, 080315740, 080315741, 080315742, 080315743, 080315744, 080315745, 080315746, 080315747, 080315748, 080315749, 080315750, 080315751, 080315752, 080315753, 080315754, 080315755, 080315756, 080315757, 080315758, 080315759, 080315760, 080315761, 080315762, 080315763, 080315764, 080315765, 080315766, 080315767, 080315768, 080315769, 080315770, 080315771, 080315772, 080315773, 080315774, 080315775, 080315776, 080315777, 080315778, 080315779, 080315780, 080315781, 080315782, 080315783, 080315784, 080315785, 080315786, 080315787, 080315788, 080315789, 080315790, 080315791, 080315792, 080315793, 080315794, 080315795, 080315796, 080315797, 080315798, 080315799, 080315800, 080315801, 080315802, 080315803, 080315804, 080315805, 080315806, 080315807, 080315808, 080315809, 080315810, 080315811, 080315812, 080315813, 080315814, 080315815, 080315816, 080315817, 080315818, 080315819, 080315820, 080315821, 080315822, 080315823, 080315824, 080315825, 080315826, 080315827, 080315828, 080315829, 080315830, 080315831, 080315832, 080315833, 080315834, 080315835, 080315836, 080315837, 080315838, 080315839, 080315840, 080315841, 080315842, 080315843, 080315844, 080315845, 080315846, 080315847, 080315848, 080315849, 080315850, 080315851, 080315852, 080315853, 080315854, 080315855, 080315856, 080315857, 080315858, 080315859, 080315860, 080315861, 080315862, 080315863, 080315864, 080315865, 080315866, 080315867, 080315868, 080315869, 080315870, 080315871, 080315872, 080315873, 080315874, 080315875, 080315876, 080315877, 080315878, 080315879, 080315925, 080315926, 080315927, 080315928, 080315929, 080315930, 080315931, 080315932, 080315933, 080315934, 080315935, 080315936, 080315937, 080315938, 080315939, 080315940, 080315941, 080315942, 080315943, 080315944, 080315945, 080315946, 080315947, 080315948, 080315949, 080315950, 080315951, 080315952, 080315953, 080315954, 080315955, 080315956, 080315957, 080315958, 080315959, 080315960, 080315961, 080315962, 080315963, 080315964, 080315965, 080315966, 080315967, 080315968, 080315969, 080315970, 080315971, 080315972, 080315973, 080315974, 080315975, 080315976, 080315977, 080315978, 080315979, 080315980, 080315981, 080315982, 080315983, 080315984, 080315985, 080315986, 080315987, 080315988, 080315989, 080315990, 080315991, 080315992, 080315993, 080315994, 080315995, 080315996, 080315997, 080315998, 080315999, 080316000, 080316001, 080316002, 080316003, 080316004, 080316005, 080316006, 080316007, 080316008, 080316009, 080316010, 080316011, 080316012, 080316013, 080316014, 080316015, 080316016, 080316017, 080316018, 080316019, 080316020, 080316021, 080316022, 080316023, 080316024, 080316025, 080316026, 080316027, 080316028, 080316029, 080316030, 080316031, 080316032, 080316033, 080316034, 080316035, 080316036, 080316037, 080316038, 080316039, 080316040, 080316041, 080316042, 080316043, 080316044, 080316045, 080316046, 080316047, 080316048, 080316049, 080316050, 080316051, 080316052, 080316053, 080316054, 080316055, 080316056, 080316057, 080316058, 080316059, 080316060, 080316061, 080316062, 080316063, 080316064, 080316065, 080316066, 080316067, 080316068, 080316069, 080316070, 080316071, 080316072, 080316073, 080316074, 080316075, 080316076, 080316077, 080316078, 080316079, 080316080, 080316081, 080316082, 080316083, 080316084, 080316085, 080316086, 080316087, 080316088, 080316089, 080316090, 080316091, 080316092, 080316093, 080316094, 080316095, 080316096, 080316097, 080316098, 080316099, 080316100, 080316101, 080316102, 080316103, 080316104, 080316105, 080316106, 080316107, 080316108, 080316109, 080316110, 080316111, 080316112, 080316113, 080316114, 080316115, 080316116, 080316117, 080316118, 080316119, 080316120, 080316121, 080316122, 080316123, 080316124, 080316125, 080316126, 080316127, 080316128, 080316129, 080316130, 080316131, 080316132, 080316133, 080316134, 080316135, 080316136, 080316137, 080316138, 080316139, 080316140, 080316141, 080316142, 080316143, 080316144, 080316145, 080316146, 080316147, 080316148, 080316149, 080316150, 080316151, 080316152, 080316153, 080316154, 080316155, 080316156, 080316157, 080316158, 080316159, 080316160, 080316161, 080316162, 080316163, 080316164, 080316165, 080316166, 080316167, 080316168, 080316169, 080316170, 080316171, 080316172, 080316173, 080316174, 080316175, 080316176, 080316177, 080316178, 080316179, 080316180, 080316226, 080316227, 080316228, 080316229, 080316230, 080316231, 080316232, 080316233, 080316234, 080316235, 080316236, 080316237, 080316238, 080316239, 080316240, 080316241, 080316242, 080316243, 080316244, 080316245, 080316246, 080316247, 080316248, 080316249, 080316250, 080316251, 080316252, 080316253, 080316254, 080316255, 080316256, 080316257, 080316258, 080316259, 080316260, 080316261, 080316262, 080316263, 080316264, 080316265, 080316266, 080316267, 080316268, 080316269, 080316270, 080316271, 080316272, 080316273, 080316274, 080316275, 080316276, 080316277, 080316278, 080316279, 080316280, 080316281, 080316282, 080316283, 080316284, 080316285, 080316286, 080316287, 080316288, 080316289, 080316290, 080316291, 080316292, 080316293, 080316294, 080316295, 080316296, 080316297, 080316298, 080316299, 080316300, 080316301, 080316302, 080316303, 080316304, 080316305, 080316306, 080316307, 080316308, 080316309, 080316310, 080316311, 080316312, 080316313, 080316314, 080316315, 080316316, 080316317, 080316318, 080316319, 080316320, 080316321, 080316322, 080316323, 080316324, 080316325, 080316326, 080316327, 080316328, 080316329, 080316330, 080316331, 080316332, 080316333, 080316334, 080316335, 080316336, 080316337, 080316338, 080316339, 080316340, 080316341, 080316342, 080316343, 080316344, 080316345, 080316346, 080316347, 080316348, 080316349, 080316350, 080316351, 080316352, 080316353, 080316354, 080316355, 080316356, 080316357, 080316358, 080316359, 080316360, 080316361, 080316362, 080316363, 080316364, 080316365, 080316366, 080316367, 080316368, 080316369, 080316370, 080316371, 080316372, 080316373, 080316374, 080316375, 080316376, 080316377, 080316378, 080316379, 080316380, 080316381, 080316382, 080316383, 080316384, 080316385, 080316386, 080316387, 080316388, 080316389, 080316390, 080316391, 080316392, 080316393, 080316394, 080316395, 080316396, 080316397, 080316398, 080316399, 080316400, 080316401, 080316402, 080316403, 080316404, 080316405, 080316406, 080316407, 080316408, 080316409, 080316410, 080316411, 080316412, 080316413, 080316414, 080316415, 080316416, 080316417, 080316418, 080316419, 080316420, 080316421, 080316422, 080316423, 080316424, 080316425, 080316426, 080316427, 080316428, 080316429, 080316430, 080316431, 080316432, 080316433, 080316434, 080316435, 080316436, 080316437, 080316438, 080316439, 080316440, 080316441, 080316442, 080316443, 080316444, 080316445, 080316446, 080316447, 080316448, 080316449, 080316450, 080316451, 080316452, 080316453, 080316454, 080316455, 080316456, 080316457, 080316458, 080316459, 080316460, 080316461, 080316462, 080316463, 080316464, 080316465, 080316466, 080316467, 080316468, 080316469, 080316470, 080316471, 080316472, 080316473, 080316474, 080316475, 080316476, 080316477, 080316478, 080316479, 080316480, 080316481, 080316482, 080316483, 080316484, 080316485, 080316486, 080316487, 080316488, 080316489, 080316490, 080316491, 080316493, 080316494, 080316495, 080316496, 080316497, 080316498, 080316499, 080316500, 080316501, 080316502, 080316503, 080316504, 080316505, 080316506, 080316507, 080316508, 080316509, 080316510, 080316511, 080316512, 080316513, 080316514, 080316515, 080316516, 080316517, 080316518, 080316519, 080316520, 080316521, 080316522, 080316523, 080316524, 080316525, 080316526, 080316527, 080316528, 080316529, 080316530, 080316531, 080316532, 080316533, 080316534, 080316535, 080316536, 080316537, 080316538, 080316539, 080316540, 080316541, 080316542, 080316543, 080316544, 080316545, 080316546, 080316547, 080316548, 080316549, 080316550, 080316551, 080316552, 080316553, 080316554, 080316555, 080316556, 080316557, 080316558, 080316559, 080316560, 080316561, 080316562, 080316563, 080316564, 080316565, 080316566, 080316567, 080316568, 080316569, 080316570, 080316571, 080316572, 080316573, 080316574, 080316575, 080316576, 080316577, 080316578, 080316579.080316580, 080316581, 080316582, 080316583, 080316584, 080316585, 080316586, 080316587, 080316588, 080316589, 080316590, 080316591, 080316592, 080316593, 080316594, 080316595, 080316596, 080316597, 080316598, 080316599, 080316600, 080316601, 080316602, 080316603, 080316604, 080316605, 080316606, 080316607, 080316608, 080316609, 080316610, 080316611, 080316612, 080316613, 080316614, 080316615, 080316616, 080316617, 080316618, 080316619, 080316620, 080316621, 080316622, 080316623, 080316624, 080316625, 080316626, 080316627, 080316628, 080316629, 080316630, 080316631, 080316632, 080316633, 080316634, 080316635, 080316636, 080316637, 080316638, 080316639, 080316640, 080316641, 080316642, 080316643, 080316644, 080316645, 080316646, 080316647, 080316648, 080316649, 080316650, 080316651, 080316652, 080316653, 080316654, 080316655, 080316656, 080316657, 080316671, 080316672, 080316673, 080316674, 080316675, 080316676, 080316677, 080316678, 080316679, 080316680, 080316681, 080316682, 080316683, 080316684, 080316685, 080316686, 080316687, 080316688, 080316689, 080316690, 080316691, 080316692, 080316693, 080316694, 080316695, 080316696, 080316697, 080316698, 080316699, 080316700, 080316701, 080316702, 080316703, 080316704, 080316705, 080316706, 080316707, 080316708, 080316709, 080316710, 080316711, 080316712, 080316713, 080316714, 080316715, 080316716, 080316717, 080316723, 080316724, 080316725, 080316726, 080316727, 080316728, 080316729, 080316730, 080316731, 080316732, 080316733, 080316734, 080316735, 080316736, 080316737, 080316738, 080316739, 080316740, 080316741, 080316742, 080316743, 080316744, 080316745, 080316746, 080316747, 080316748, 080316749, 080316750, 080316751, 080316752, 080316753, 080316754, 080316755, 080316756, 080316757, 080316758, 080316759, 080316760, 080316761, 080316762, 080316763, 080316764, 080316765, 080316766, 080316767, 080316768, 080316769, 080316770, 080316771, 080316772, 080316773, 080316774, 080316775, 080316776, 080316777, 080316778, 080316815, 080316816, 080316817, 080316818, 080316819, 080316820, 080316821, 080316822, 080316823, 080316824, 080316825, 080316826, 080316827, 080316828, 080316829, 080316830, 080316831, 080316832, 080316833, 080316834, 080316835, 080316836, 080316837, 080316838, 080316839, 080316840, 080316841, 080316842, 080316843, 080316844, 080316845, 080316846, 080316847, 080316848, 080316849, 080316850, 080316851, 080316852, 080316853, 080316854, 080316855, 080316856, 080316857, 080316858, 080316859, 080415001, 080415002, 080415003, 080415004, 080415005, 080415006, 080415007, 080415008, 080415009, 080415010, 080415011, 080415012, 080415013, 080415014, 080415015, 080415016, 080415017, 080415018, 080415019, 080415020, 080415021, 080415022, 080415023, 080415024, 080415025, 080415026, 080415027, 080415028, 080415029, 080415030, 080415031, 080415032, 080415033, 080415034, 080415035, 080415036, 080415037, 080415038, 080415039, 080415040, 080415041, 080415042, 080415043, 080415044, 080415045, 080415046, 080415047, 080415048, 080415049, 080415050, 080415051, 080415052, 080415053, 080415054, 080415055, 080415056, 080415057, 080415058, 080415059, 080415060, 080415061, 080415062, 080415063, 080415064, 080415065, 080415066, 080415067, 080415068, 080415069, 080415070, 080415071, 080415072, 080415073, 080415074, 080415075, 080415076, 080415077, 080415078, 080415079, 080415080, 080415081, 080415082, 080415083, 080415084, 080415085, 080415086, 080415087, 080415088, 080415089, 080415090, 080415091, 080415092, 080415093, 080415094, 080415095, 080415096, 080415097, 080415098, 080415099, 080415100, 080415101, 080415102, 080415103, 080415104, 080415105, 080415106, 080415107, 080415108, 080415109, 080415110, 080415111, 080415112, 080415113, 080415114, 080415115, 080415116, 080415117, 080415118, 080415119, 080415120, 080415121, 080415122, 080415123, 080415124, 080415125, 080415126, 080415127, 080415128, 080415129, 080415130, 080415131, 080415132, 080415133, 080415134, 080415135, 080415136, 080415137, 080415138, 080415139, 080415140, 080415141, 080415142, 080415143, 080415144, 080415145, 080415146, 080415147, 080415148, 080415149, 080415150, 080415151, 080415152, 080415153, 080415154, 080415155, 080415156, 080415157, 080415158, 080415159, 080415160, 080415161, 080415162, 080415163, 080415164, 080415165, 080415166, 080415167, 080415168, 080415169, 080415170, 080415171, 080415172, 080415173, 080415174, 080415175, 080415176, 080415177, 080415178, 080415179, 080415180, 080415181, 080415182, 080415183, 080415184, 080415185, 080415186, 080415187, 080415188, 080415189, 080415190, 080415191, 080415192, 080415193, 080415194, 080415195, 080415196, 080415197, 080415198, 080415199, 080415200, 080415201, 080415202, 080415203, 080415204, 080415205, 080415206, 080415207, 080415208, 080415209, 080415210, 080415211, 080415212, 080415213, 080415214, 080415215, 080415216, 080415217, 080415218, 080415219, 080415220, 080415221, 080415222, 080415223, 080415224, 080415225, 080415226, 080415227, 080415228, 080415229, 080415230, 080415231, 080415232, 080415233, 080415234, 080415235, 080415236, 080415237, 080415238, 080415239, 080415240, 080415241, 080415242, 080415243, 080415244, 080415245, 080415246, 080415247, 080415248, 080415249, 080415250, 080415251, 080415252, 080415253, 080415254, 080415255, 080415256, 080415257, 080415258, 080415259, 080415260, 080415261, 080415262, 080415263, 080415264, 080415265, 080415266, 080415267, 080415268, 080415269, 080415270, 080415271, 080415272, 080415273, 080415274, 080415275, 080415276, 080415277, 080415278, 080415279, 080415280, 080415281, 080415282, 080415283, 080415284, 080415285, 080415286, 080415287, 080415288, 080415289, 080415290, 080415291, 080415292, 080415293, 080415294, 080415295, 080415296, 080415297, 080415298, 080415299, 080415300, 080415301, 080415302, 080415303, 080415304, 080415305, 080415306, 080415307, 080415308, 080415309, 080415310, 080415311, 080415312, 080415313, 080415314, 080415315, 080415316, 080415317, 080415318, 080415319, 080415320, 080415321, 080415322, 080415323, 080415324, 080415325, 080415326, 080415327, 080415328, 080415329, 080415330, 080415331, 080415332, 080415333, 080415334, 080415335, 080415336, 080415337, 080415338, 080415339, 080415340, 080415341, 080415342, 080415343, 080415344, 080415345, 080415346, 080415347, 080415348, 080415349, 080415350, 080415351, 080415352, 080415353, 080415354, 080415355, 080415356, 080415357, 080415358, 080415359, 080415360, 080415361, 080415362, 080415363, 080415364, 080415365, 080415366, 080415367, 080415368, 080415369, 080415370, 080415371, 080415372, 080415373, 080415374, 080415375, 080415376, 080415377, 080415378, 080415379, 080415380, 080415381, 080415382, 080415383, 080415384, 080415385, 080415386, 080415387, 080415388, 080415389, 080415390, 080415391, 080415392, 080415393, 080415394, 080415395, 080415396, 080415397, 080415398, 080415399, 080415400, 080415401, 080415402, 080415403, 080415404, 080415405, 080415406, 080415407, 080415408, 080415409, 080415410, 080415411, 080415412, 080415413, 080415414, 080415415, 080415416, 080415417, 080415418, 080415419, 080415420, 080415421, 080415422, 080415423, 080415424, 080415425, 080415426, 080415427, 080415428, 080415429, 080415430, 080415431, 080415432, 080415433, 080415434, 080415435, 080415436, 080415437, 080415438, 080415439, 080415440, 080415441, 080415442, 080415443, 080415444, 080415445, 080415446, 080415447, 080415448, 080415449, 080415450, 080415451, 080415452, 080415453, 080415454, 080415455, 080415456, 080415457, 080415458, 080415459, 080415460, 080415461, 080415462, 080415463, 080415464, 080415465, 080415466, 080415467, 080415468, 080415469, 080415470, 080415471, 080415472, 080415473, 080415474, 080415475, 080415476, 080415477, 080415478, 080415479, 080415480, 080415481, 080415482, 080415483, 080415484, 080415485, 080415486, 080415487, 080415488, 080415489, 080415490, 080415491, 080415492, 080415493, 080415494, 080415495, 080415496, 080415497, 080415498, 080415499, 080415500, 080415501, 080415502, 080415503, 080415504, 080415505, 080415506, 080415507, 080415508, 080415509, 080415510, 080415511, 080415512, 080415513, 080415514, 080415515, 080415516, 080415517, 080415518, 080415519, 080415520, 080415521, 080415522, 080415523, 080415524, 080415525, 080415526, 080415527, 080415528, 080415529, 080415530, 080415531, 080415532, 080415533, 080415534, 080415535, 080415536, 080415537, 080415538, 080415539, 080415540, 080415541, 080415542, 080415543, 080415544, 080415545, 080415546, 080415547, 080415548, 080415549, 080415550, 080415551, 080415552, 080415553, 080415554, 080415555, 080415556, 080415557, 080415558, 080415559, 080415560, 080415561, 080415562, 080415563, 080415564, 080415565, 080415566, 080415567, 080415568, 080415569, 080415570, 080415571, 080415572, 080415573, 080415574, 080415575, 080415576, 080415577, 080415578, 080415579, 080415580, 080415581, 080415582, 080415583, 080415584, 080415585, 080415586, 080415587, 080415588, 080415589, 080415590, 080415591, 080415592, 080415593, 080415594, 080415595, 080415596, 080415597, 080415598, 080415599, 080415600, 080415601, 080415602, 080415603, 080415604, 080415605, 080415606, 080415607, 080415608, 080415609, 080415610, 080415611, 080415612, 080415613, 080415614, 080415615, 080415616, 080415617, 080415618, 080415619, 080415620, 080415621, 080415622, 080415623, 080415624, 080415625, 080415626, 080415627, 080415628, 080415629, 080415630, 080415631, 080415632, 080415633, 080415634, 080415635, 080415636, 080415637, 080415638, 080415639, 080415640, 080415641, 080415642, 080415643, 080415644, 080415645, 080415646, 080415647, 080415648, 080415649, 080415650, 080415651, 080415652, 080415653, 080415654, 080415655, 080415656, 080415657, 080415658, 080415659, 080415660, 080415661, 080415662, 080415663, 080415664, 080415665, 080415666, 080415667, 080415668, 080415669, 080415670, 080415671, 080415672, 080415673, 080415674, 080415675, 080415676, 080415677, 080415678, 080415679, 080415680, 080415681, 080415682, 080415683, 080415684, 080415685, 080415686, 080415687, 080415688, 080415689, 080415690, 080415691, 080415692, 080415693, 080415694, 080415695, 080415696, 080415697, 080415698, 080415699, 080415700, 080415701, 080415702, 080415703, 080415704, 080415705, 080415706, 080415707, 080415708, 080415709, 080415710, 080415711, 080415712, 080415713, 080415714, 080415715, 080415716, 080415717, 080415718, 080415719, 080415720, 080415721, 080415722, 080415723, 080415724, 080415725, 080415726, 080415727, 080415728, 080415729, 080415730, 080415731, 080415732, 080415733, 080415734, 080415735, 080415736, 080415737, 080415738, 080415739, 080415740, 080415741, 080415742, 080415743, 080415744, 080415745, 080415746, 080415747, 080415748, 080415749, 080415750, 080415751, 080415752, 080415753, 080415754, 080415755, 080415756, 080415757, 080415758, 080415759, 080415760, 080415761, 080415762, 080415763, 080415764, 080415765, 080415766, 080415767, 080415768, 080415769, 080415770, 080415771, 080415772, 080415773, 080415774, 080415775, 080415776, 080415777, 080415778, 080415779, 080415780, 080415781, 080415782, 080415783, 080415784, 080415785, 080415786, 080415787, 080415788, 080415789, 080415790, 080415791, 080415792, 080415793, 080415794, 080415795, 080415796, 080415797, 080415798, 080415799, 080415800, 080415801, 080415802, 080415803, 080415804, 080415805, 080415806, 080415807, 080415808, 080415809, 080415810, 080415811, 080415812, 080415813, 080415814, 080415815, 080415816, 080415817, 080415818, 080415819, 080415820, 080415821, 080415822, 080415823, 080415824, 080415825, 080415826, 080415827, 080415828, 080415829, 080415830, 080415831, 080415832, 080415833, 080415834, 080415835, 080415836, 080415837, 080415838, 080415839, 080415840, 080415841, 080415842, 080415843, 080415844, 080415845, 080415846, 080415847, 080415848, 080415849, 080415850, 080415851, 080415852, 080415853, 080415854, 080415855, 080415856, 080415857, 080415858, 080415859, 080415860, 080415861, 080415862, 080415863, 080415864, 080415865, 080415866, 080415867, 080415868, 080415869, 080415870, 080415871, 080415872, 080415873, 080415874, 080415875, 080415876, 080415877, 080415878, 080415879, 080415880, 080415881, 080415882, 080415883, 080415884, 080415885, 080415886, 080415887, 080415888, 080415889, 080415890, 080415891, 080415892, 080415893, 080415894, 080415895, 080415896, 080415897, 080415898, 080415899, 080415900, 080415901, 080415902, 080415903, 080415904, 080415905, 080415906, 080415907, 080415908, 080415909, 080415910, 080415911, 080415912, 080415913, 080415914, 080415915, 080415916, 080415917, 080415918, 080415919, 080415920, 080415921, 080415922, 080415923, 080415924, 080415925, 080415926, 080415927, 080415928, 080415929, 080415930, 080415931, 080415932, 080415933, 080415934, 080415935, 080415936, 080415937, 080415938, 080415939, 080415940, 080415941, 080415942, 080415943, 080415944, 080415945, 080415946, 080415947, 080415948, 080415949, 080415950, 080415951, 080415952, 080415953, 080415954, 080415955, 080415956, 080415957, 080415958, 080415959, 080415960, 080415961, 080415962, 080415963, 080415964, 080415965, 080415966, 080415967, 080415968, 080415969, 080415970, 080415971, 080415972, 080415973, 080415974, 080415975, 080415976, 080415977, 080415978, 080415979, 080415980, 080415981, 080415982, 080415983, 080415984, 080415985, 080415986, 080415987, 080415988, 080415989, 080415990, 080415991, 080415992, 080415993, 080415994, 080415995, 080415996, 080415997, 080415998, 080415999, 080416000, 080416001, 080416002, 080416003, 080416004, 080416005, 080416006, 080416007, 080416008, 080416009, 080416010, 080416011, 080416012, 080416013, 080416014, 080416015, 080416016, 080416017, 080416018, 080416019, 080416020, 080416021, 080416022, 080416023, 080416024, 080416025, 080416026, 080416027, 080416028, 080416029, 080416030, 080416031, 080416032, 080416033, 080416034, 080416035, 080416036, 080416037, 080416038, 080416039, 080416040, 080416041, 080416042, 080416043, 080416044, 080416045, 080416046, 080416047, 080416048, 080416049, 080416050, 080416051, 080416052, 080416053, 080416054, 080416055, 080416056, 080416057, 080416058, 080416059, 080416060, 080416061, 080416062, 080416063, 080416064, 080416065, 080416066, 080416067, 080416068, 080416069, 080416070, 080416071, 080416072, 080416073, 080416074, 080416075, 080416076, 080416077, 080416078, 080416079, 080416080, 080416081, 080416082, 080416083, 080416084, 080416085, 080416086, 080416087, 080416088, 080416089, 080416090, 080416091, 080416092, 080416093, 080416094, 080416095, 080416096, 080416097, 080416098, 080416099, 080416100, 080416101, 080416102, 080416103, 080416104, 080416105, 080416106, 080416107, 080416108, 080416109, 080416110, 080416111, 080416112, 080416113, 080416114, 080416115, 080416116, 080416117, 080416118, 080416119, 080416120, 080416121, 080416122, 080416123, 080416124, 080416125, 080416126, 080416127, 080416128, 080416129, 080416130, 080416131, 080416132, 080416133, 080416134, 080416135, 080416136, 080416137, 080416138, 080416139, 080416140, 080416141, 080416142, 080416143, 080416144, 080416145, 080416146, 080416147, 080416148, 080416149, 080416150, 080416151, 080416152, 080416153, 080416154, 080416155, 080416156, 080416157, 080416158, 080416159, 080416160, 080416161, 080416162, 080416163, 080416164, 080416165, 080416166, 080416167, 080416168, 080416169, 080416170, 080416171, 080416172, 080416173, 080416174, 080416175, 080416176, 080416177, 080416178, 080416179, 080416180, 080416181, 080416182, 080416183, 080416184, 080416185, 080416186, 080416187, 080416188, 080416189, 080416190, 080416191, 080416192, 080416193, 080416194, 080416195, 080416196, 080416197, 080416198, 080416199, 080416200, 080416201, 080416202, 080416203, 080416204, 080416205, 080416206, 080416207, 080416208, 080416209, 080416210, 080416211, 080416212, 080416213, 080416214, 080416215, 080416216, 080416217, 080416218, 080416219, 080416220, 080416221, 080416222, 080416223, 080416224, 080416225, 080416226, 080416227, 080416228, 080416229, 080416230, 080416231, 080416232, 080416233, 080416234, 080416235, 080416236, 080416237, 080416238, 080416239, 080416240, 080416241, 080416242, 080416243, 080416244, 080416245, 080416291, 080416292, 080416293, 080416294, 080416295, 080416296, 080416297, 080416298, 080416299, 080416300, 080416301, 080416302, 080416303, 080416304, 080416305, 080416306, 080416307, 080416308, 080416309, 080416310, 080416311, 080416312, 080416313, 080416314, 080416315, 080416316, 080416317, 080416318, 080416319, 080416320, 080416321, 080416322, 080416323, 080416324, 080416325, 080416326, 080416327, 080416328, 080416329, 080416330, 080416331, 080416332, 080416333, 080416334, 080416335, 080416336, 080416337, 080416338, 080416339, 080416340, 080416341, 080416342, 080416343, 080416344, 080416345, 080416346, 080416347, 080416348, 080416349, 080416350, 080416351, 080416352, 080416353, 080416354, 080416355, 080416356, 080416357, 080416358, 080416359, 080416360, 080416361, 080416362, 080416363, 080416364, 080416365, 080416366, 080416367, 080416368, 080416369, 080416370, 080416371, 080416372, 080416373, 080416374, 080416375, 080416376, 080416377, 080416378, 080416379, 080416380, 080416381, 080416382, 080416383, 080416384, 080416385, 080416386, 080416387, 080416388, 080416389, 080416390, 080416391, 080416392, 080416393, 080416394, 080416395, 080416396, 080416397, 080416398, 080416399, 080416400, 080416401, 080416402, 080416403, 080416404, 080416405, 080416406, 080416407, 080416408, 080416409, 080416410, 080416411, 080416412, 080416413, 080416414, 080416415, 080416416, 080416417, 080416418, 080416419, 080416420, 080416421, 080416422, 080416423, 080416424, 080416425, 080416426, 080416427, 080416428, 080416429, 080416430, 080416431, 080416432, 080416433, 080416434, 080416435, 080416436, 080416437, 080416438, 080416439, 080416440, 080416441, 080416442, 080416443, 080416444, 080416445, 080416446, 080416447, 080416448, 080416449, 080416450, 080416451, 080416452, 080416453, 080416454, 080416455, 080416456, 080416457, 080416458, 080416459, 080416460, 080416461, 080416462, 080416463, 080416464, 080416465, 080416466, 080416467, 080416468, 080416469, 080416470, 080416471, 080416472, 080416473, 080416474, 080416475, 080416476, 080416477, 080416478, 080416479, 080416480, 080416481, 080416482, 080416483, 080416484, 080416485, 080416486, 080416487, 080416488, 080416489, 080416490, 080416491, 080416492, 080416493, 080416494, 080416495, 080416496, 080416497, 080416498, 080416499, 080416500, 080416501, 080416502, 080416503, 080416504, 080416505, 080416506, 080416507, 080416508, 080416509, 080416510, 080416511, 080416512, 080416513, 080416514, 080416515, 080416516, 080416517, 080416518, 080416519, 080416520, 080416521, 080416522, 080416523, 080416524, 080416525, 080416526, 080416527, 080416528, 080416529, 080416530, 080416531, 080416532, 080416533, 080416534, 080416535, 080416536, 080416537, 080416538, 080416539, 080416540, 080416541, 080416542, 080416543, 080416544, 080416545, 080416546, 080416547, 080416548, 080416549, 080416550, 080416551, 080416552, 080416553, 080416554, 080416555, 080416556, 080416557, 080416558, 080416559, 080416560, 080416561, 080416562, 080416563, 080416564, 080416565, 080416566, 080416567, 080416568, 080416569, 080416570, 080416571, 080416572, 080416573, 080416574, 080416575, 080416576, 080416577, 080416578, 080416579, 080416580, 080416581, 080416582, 080416583, 080416584, 080416585, 080416586, 080416587, 080416588, 080416589, 080416590, 080416591, 080416592, 080416593, 080416594, 080416595, 080416596, 080416597, 080416598, 080416599, 080416600, 080416601, 080416602, 080416603, 080416604, 080416605, 080416606, 080416607, 080416608, 080416609, 080416610, 080416611, 080416612, 080416613, 080416614, 080416615, 080416616, 080416617, 080416618, 080416619, 080416620, 080416621, 080416622, 080416623, 080416624, 080416625, 080416626, 080416627, 080416628, 080416629, 080416630, 080416631, 080416632, 080416633, 080416634, 080416635, 080416636, 080416637, 080416638, 080416639, 080416640, 080416641, 080416642, 080416643, 080416644, 080416645, 080416646, 080416647, 080416648, 080416649, 080416650, 080416651, 080416652, 080416653, 080416654, 080416655, 080416656, 080416657, 080416658, 080416659, 080416660, 080515001, 080515002, 080515003, 080515004, 080515005, 080515006, 080515007, 080515008, 080515009, 080515010, 080515011, 080515012, 080515013, 080515014, 080515015, 080515016, 080515017, 080515018, 080515019, 080515020, 080515021, 080515022, 080515023, 080515024, 080515025, 080515026, 080515027, 080515028, 080515029, 080515030, 080515031, 080515032, 080515033, 080515034, 080515035, 080515036, 080515037, 080515038, 080515039, 080515040, 080515041, 080515042, 080515043, 080515044, 080515045, 080515046, 080515047, 080515048, 080515049, 080515050, 080515051, 080515052, 080515053, 080515054, 080515055, 080515056, 080515057, 080515058, 080515059, 080515060, 080515061, 080515062, 080515063, 080515064, 080515065, 080515066, 080515067, 080515068, 080515069, 080515070, 080515071, 080515072, 080515073, 080515074, 080515075, 080515076, 080515077, 080515078, 080515079, 080515080, 080515081, 080515082, 080515083, 080515084, 080515085, 080515086, 080515087, 080515088, 080515089, 080515090, 080515091, 080515092, 080515093, 080515094, 080515095, 080515096, 080515097, 080515098, 080515099, 080515100, 080515101, 080515102, 080515103, 080515104, 080515105, 080515106, 080515107, 080515108, 080515109, 080515110, 080515111, 080515112, 080515113, 080515114, 080515115, 080515116, 080515117, 080515118, 080515119, 080515120, 080515121, 080515122, 080515123, 080515124, 080515125, 080515126, 080515127, 080515128, 080515129, 080515130, 080515131, 080515132, 080515133, 080515134, 080515135, 080515136, 080515137, 080515138, 080515139, 080515140, 080515141, 080515142, 080515143, 080515144, 080515145, 080515146, 080515147, 080515148, 080515149, 080515150, 080515151, 080515152, 080515153, 080515154, 080515155, 080515156, 080515157, 080515158, 080515159, 080515160, 080515161, 080515162, 080515163, 080515164, 080515165, 080515166, 080515167, 080515168, 080515169, 080515170, 080515171, 080515172, 080515173, 080515174, 080515175, 080515176, 080515177, 080515178, 080515179, 080515180, 080515181, 080515182, 080515183, 080515184, 080515185, 080515186, 080515187, 080515188, 080515189, 080515190, 080515191, 080515192, 080515193, 080515194, 080515195, 080515196, 080515197, 080515198, 080515199, 080515200, 080515201, 080515202, 080515203, 080515204, 080515205, 080515206, 080515207, 080515208, 080515209, 080515210, 080515211, 080515212, 080515213, 080515214, 080515215, 080515216, 080515217, 080515218, 080515219, 080515220, 080515221, 080515222, 080515223, 080515224, 080515225, 080515226, 080515227, 080515228, 080515229, 080515230, 080515231, 080515232, 080515233, 080515234, 080515235, 080515236, 080515237, 080515238, 080515239, 080515240, 080515241, 080515242, 080515243, 080515244, 080515245, 080515246, 080515247, 080515248, 080515249, 080515250, 080515251, 080515252, 080515253, 080515254, 080515255, 080515256, 080515257, 080515258, 080515259, 080515260, 080515261, 080515262, 080515263, 080515264, 080515265, 080515266, 080515267, 080515268, 080515269, 080515270, 080515271, 080515272, 080515273, 080515274, 080515275, 080515276, 080515277, 080515278, 080515279, 080515280, 080515281, 080515282, 080515283, 080515284, 080515285, 080515286, 080515287, 080515288, 080515289, 080515290, 080515291, 080515292, 080515293, 080515294, 080515295, 080515296, 080515297, 080515298, 080515299, 080515300, 080515301, 080515302, 080515303, 080515304, 080515305, 080515306, 080515307, 080515308, 080515309, 080515310, 080515311, 080515312, 080515313, 080515314, 080515315, 080515316, 080515317, 080515318, 080515319, 080515320, 080515321, 080515322, 080515323, 080515324, 080515325, 080515326, 080515327, 080515328, 080515329, 080515330, 080515331, 080515332, 080515333, 080515334, 080515335, 080515336, 080515337, 080515338, 080515339, 080515340, 080515341, 080515342, 080515343, 080515344, 080515345, 080515346, 080515347, 080515348, 080515349, 080515350, 080515351, 080515352, 080515353, 080515354, 080515355, 080515356, 080515357, 080515358, 080515359, 080515360, 080515361, 080515362, 080515363, 080515364, 080515365, 080515366, 080515367, 080515368, 080515369, 080515370, 080515371, 080515372, 080515373, 080515374, 080515375, 080515376, 080515377, 080515378, 080515379, 080515380, 080515381, 080515382, 080515383, 080515384, 080515385, 080515386, 080515387, 080515388, 080515389, 080515390, 080515391, 080515392, 080515393, 080515394, 080515395, 080515396, 080515397, 080515398, 080515399, 080515400, 080515401, 080515402, 080515403, 080515404, 080515405, 080515406, 080515407, 080515408, 080515409, 080515410, 080515411, 080515412, 080515413, 080515414, 080515415, 080515416, 080515417, 080515418, 080515419, 080515420, 080515421, 080515422, 080515423, 080515424, 080515425, 080515426, 080515427, 080515428, 080515429, 080515430, 080515431, 080515432, 080515433, 080515434, 080515435, 080515436, 080515437, 080515438, 080515439, 080515440, 080515441, 080515442, 080515443, 080515444, 080515445, 080515446, 080515447, 080515448, 080515449, 080515450, 080515451, 080515452, 080515453, 080515454, 080515455, 080515456, 080515457, 080515458, 080515459, 080515460, 080515461, 080515462, 080515463, 080515464, 080515465, 080515466, 080515467, 080515468, 080515469, 080515470, 080515471, 080515472, 080515473, 080515474, 080515475, 080515476, 080515477, 080515478, 080515479, 080515480, 080515481, 080515482, 080515483, 080515484, 080515485, 080515486, 080515487, 080515488, 080515489, 080515490, 080515491, 080515492, 080515493, 080515494, 080515495, 080515496, 080515497, 080515498, 080515499, 080515500, 080515501, 080515502, 080515503, 080515504, 080515505, 080515506, 080515507, 080515508, 080515509, 080515510, 080515511, 080515512, 080515513, 080515514, 080515515, 080515516, 080515517, 080515518, 080515519, 080515520, 080515521, 080515522, 080515523, 080515524, 080515525, 080515526, 080515527, 080515528, 080515529, 080515530, 080515531, 080515532, 080515533, 080515534, 080515535, 080515536, 080515537, 080515538, 080515539, 080515540, 080515541, 080515542, 080515543, 080515544, 080515545, 080515546, 080515547, 080515548, 080515549, 080515550, 080515551, 080515552, 080515553, 080515554, 080515555, 080515556, 080515557, 080515558, 080515559, 080515560, 080515561, 080515562, 080515563, 080515564, 080515565, 080515566, 080515567, 080515568, 080515569, 080515570, 080515571, 080515572, 080515573, 080515574, 080515575, 080515576, 080515577, 080515578, 080515579, 080515580, 080515581, 080515582, 080515583, 080515584, 080515585, 080515586, 080515587, 080515588, 080515589, 080515590, 080515591, 080515592, 080515593, 080515594, 080515595, 080515596, 080515597, 080515598, 080515599, 080515600, 080515601, 080515602, 080515603, 080515604, 080515605, 080515606, 080515607, 080515608, 080515609, 080515610, 080515611, 080515612, 080515613, 080515614, 080515615, 080515616, 080515617, 080515618, 080515619, 080515620, 080515621, 080515622, 080515623, 080515624, 080515625, 080515626, 080515627, 080515628, 080515629, 080515630, 080515631, 080515632, 080515633, 080515634, 080515635, 080515636, 080515637, 080515638, 080515639, 080515640, 080515641, 080515642, 080515643, 080515644, 080515645, 080515646, 080515647, 080515648, 080515649, 080515650, 080515651, 080515652, 080515653, 080515654, 080515655, 080515656, 080515657, 080515658, 080515659, 080515660, 080515661, 080515662, 080515663, 080515664, 080515665, 080515666, 080515667, 080515668, 080515669, 080515670, 080515671, 080515672, 080515673, 080515674, 080515675, 080515676, 080515677, 080515678, 080515679, 080515680, 080515681, 080515682, 080515683, 080515684, 080515685, 080515686, 080515687, 080515688, 080515689, 080515690, 080515691, 080515692, 080515693, 080515694, 080515695, 080515696, 080515697, 080515698, 080515699, 080515700, 080515701, 080515702, 080515703, 080515704, 080515705, 080515706, 080515707, 080515708, 080515709, 080515710, 080515711, 080515712, 080515713, 080515714, 080515715, 080515716, 080515717, 080515718, 080515719, 080515720, 080515721, 080515722, 080515723, 080515724, 080515725, 080515726, 080515727, 080515728, 080515729, 080515730, 080515731, 080515732, 080515733, 080515734, 080515735, 080515736, 080515737, 080515738, 080515739, 080515740, 080515741, 080515742, 080515743, 080515744, 080515745, 080515746, 080515747, 080515748, 080515749, 080515750, 080515751, 080515752, 080515753, 080515754, 080515755, 080515756, 080515757, 080515758, 080515759, 080515760, 080515761, 080515762, 080515763, 080515764, 080515765, 080515766, 080515767, 080515768, 080515769, 080515770, 080515771, 080515772, 080515773, 080515774, 080515775, 080515776, 080515777, 080515778, 080515779, 080515780, 080515781, 080515782, 080515783, 080515784, 080515785, 080515786, 080515787, 080515788, 080515789, 080515790, 080515791, 080515792, 080515793, 080515794, 080515795, 080515796, 080515797, 080515798, 080515799, 080515800, 080515801, 080515802, 080515803, 080515804, 080515805, 080515806, 080515807, 080515808, 080515809, 080515810, 080515811, 080515812, 080515813, 080515814, 080515815, 080515816, 080515817, 080515818, 080515819, 080515820, 080515821, 080515822, 080515823, 080515824, 080515825, 080515826, 080515827, 080515828, 080515829, 080515830, 080515831, 080515832, 080515833, 080515834, 080515835, 080515836, 080515837, 080515838, 080515839, 080515840, 080515841, 080515842, 080515843, 080515844, 080515845, 080515846, 080515847, 080515848, 080515849, 080515850, 080515851, 080515852, 080515853, 080515854, 080515855, 080515856, 080515857, 080515858, 080515859, 080515860, 080515861, 080515862, 080515863, 080515864, 080515865, 080515866, 080515867, 080515868, 080515869, 080515870, 080515871, 080515872, 080515873, 080515874, 080515875, 080515876, 080515877, 080515878, 080515879, 080515880, 080515881, 080515882, 080515883, 080515884, 080515885, 080515886, 080515887, 080515888, 080515889, 080515890, 080515891, 080515892, 080515893, 080515894, 080515895, 080515896, 080515897, 080515898, 080515899, 080515900, 080515901, 080515902, 080515903, 080515904, 080515905, 080515906, 080515907, 080515908, 080515909, 080515910, 080515911, 080515912, 080515913, 080515914, 080515915, 080515916, 080515917, 080515918, 080515919, 080515920, 080515921, 080515922, 080515923, 080515924, 080515925, 080515926, 080515927, 080515928, 080515929, 080515930, 080515931, 080515932, 080515933, 080515934, 080515935, 080515936, 080515937, 080515938, 080515939, 080515940, 080515941, 080515942, 080515943, 080515944, 080515945, 080515946, 080515947, 080515948, 080515949, 080515950, 080515951, 080515952, 080515953, 080515954, 080515955, 080515956, 080515957, 080515958, 080515959, 080515960, 080515961, 080515962, 080515963, 080515964, 080515965, 080515966, 080515967, 080515968, 080515969, 080515970, 080515971, 080515972, 080515973, 080515974, 080515975, 080515976, 080515977, 080515978, 080515979, 080515980, 080515981, 080515982, 080515983, 080515984, 080515985, 080515986, 080515987, 080515988, 080515989, 080515990, 080515991, 080515992, 080515993, 080515994, 080515995, 080515996, 080515997, 080515998, 080515999, 080516000, 080516001, 080516002, 080516003, 080516004, 080516005, 080516006, 080516007, 080516008, 080516009, 080516010, 080516011, 080516012, 080516013, 080516014, 080516015, 080516016, 080516017, 080516018, 080516019, 080516020, 080516021, 080516022, 080516023, 080516024, 080516025, 080516026, 080516027, 080516028, 080516029, 080516030, 080516031, 080516032, 080516033, 080516034, 080516035, 080516036, 080516037, 080516038, 080516039, 080516040, 080516041, 080516042, 080516043, 080516044, 080516045, 080516046, 080516047, 080516048, 080516049, 080516050, 080516051, 080516052, 080516053, 080516054, 080516055, 080516056, 080516057, 080516058, 080516059, 080516060, 080516061, 080516062, 080516063, 080516064, 080516065, 080516066, 080516067, 080516068.080516069, 080516070, 080516071, 080516072, 080516073, 080516074, 080516075, 080516076, 080516077, 080516078, 080516079, 080516080, 080516081, 080516082, 080516083, 080516084, 080516085, 080516086, 080516087, 080516088, 080516089, 080516090, 080516091, 080516092, 080516093, 080516094, 080516095, 080516096, 080516097, 080516098, 080516099, 080516100, 080516101, 080516102, 080516103, 080516104, 080516105, 080516106, 080516107, 080516108, 080516109, 080516110, 080516111, 080516112, 080516113, 080516114, 080516115, 080516116, 080516117, 080516118, 080516119, 080516120, 080516121, 080516122, 080516123, 080516124, 080516125, 080516126, 080516127, 080516128, 080516129, 080516130, 080516131, 080516132, 080516133, 080516134, 080516135, 080516136, 080516137, 080516138, 080516139, 080516140, 080516141, 080516142, 080516143, 080516144, 080516145, 080516146, 080516147, 080516148, 080516149, 080516150, 080516151, 080516152, 080516153, 080516154, 080516155, 080516156, 080516157, 080516158, 080516159, 080516160, 080516161, 080516162, 080516163, 080516164, 080516165, 080516166, 080516167, 080516168, 080516169, 080516170, 080516171, 080516172, 080516173, 080516174, 080516175, 080516176, 080516177, 080516178, 080516179, 080516180, 080516181, 080516182, 080516183, 080516184, 080516185, 080516186, 080516187, 080516188, 080516189, 080516190, 080516191, 080516192, 080516193, 080516194, 080516195, 080516196, 080516197, 080516198, 080516199, 080516200, 080516201, 080516202, 080516203, 080516204, 080516205, 080516206, 080516207, 080516208, 080516209, 080516210, 080516211, 080516212, 080516213, 080516214, 080516215, 080516216, 080516217, 080516218, 080516219, 080516220, 080516221, 080516222, 080516223, 080516224, 080516225, 080516226, 080516227, 080516228, 080516229, 080516230, 080516231, 080516232, 080516233, 080516234, 080516235, 080516236, 080516252, 080516253, 080516254, 080516255, 080516256, 080516257, 080516258, 080516259, 080516260, 080516261, 080516262, 080516263, 080516264, 080516265, 080516266, 080516267, 080516268, 080516269, 080516270, 080516271, 080516272, 080516273, 080516274, 080516275, 080516276, 080516277, 080516278, 080516279, 080516280, 080516281, 080516282, 080516283, 080516284, 080516285, 080516286, 080516287, 080516288, 080516289, 080516290, 080516291, 080516292, 080516293, 080516294, 080516295, 080516296, 080516297, 080516298, 080516299, 080516300, 080516301, 080516302, 080516303, 080516304, 080516305, 080516306, 080516307, 080516308, 080516309, 080516310, 080516311, 080516312, 080516313, 080516314, 080516315, 080516316, 080516317, 080516318, 080516319, 080516320, 080516321, 080516322, 080516323, 080516324, 080516325, 080516326, 080516327, 080516328, 080516329, 080516330, 080516331, 080516332, 080516333, 080516334, 080516335, 080516336, 080516337, 080516338, 080516339, 080516340, 080516341, 080516342, 080516343, 080516344, 080516345, 080516346, 080516347, 080516348, 080516349, 080516350, 080516351, 080516352, 080516353, 080516354, 080516355, 080516356, 080516357, 080516358, 080516359, 080516360, 080516361, 080516362, 080516363, 080516364, 080516365, 080516366, 080516367, 080516368, 080516369, 080516370, 080516371, 080516372, 080516373, 080516374, 080516375, 080516376, 080516377, 080516378, 080516379, 080516380, 080516381, 080516382, 080516383, 080516384, 080516385, 080516386, 080516387, 080516388, 080516389, 080516390, 080516391, 080516392, 080516393, 080516394, 080516395, 080516396, 080516397, 080516398, 080516399, 080516400, 080516426, 080516427, 080516428, 080516429, 080516430, 080516431, 080516432, 080516433, 080516434, 080516435, 080516436, 080516437, 080516438, 080516439, 080516440, 080516441, 080516442, 080516443, 080516444, 080516445, 080516446, 080516447, 080516448, 080516449, 080516450, 080516451, 080516452, 080516453, 080516454, 080516455, 080516456, 080516457, 080516458, 080516459, 080516460, 080516461, 080516462, 080516463, 080516464, 080516465, 080516466, 080516467, 080516468, 080516469, 080516470, 080516471, 080516472, 080516473, 080516474, 080516475, 080516476, 080516477, 080516478, 080516479, 080516480, 080516481, 080516482, 080516483, 080516484, 080516485, 080516486, 080516487, 080516488, 080516489, 080516490, 080516491, 080516492, 080516493, 080516494, 080516495, 080516496, 080516497, 080516498, 080516499, 080516500, 080516501, 080516502, 080516503, 080516504, 080516505, 080516506, 080516507, 080516508, 080516509, 080516510, 080516511, 080516512, 080516513, 080516514, 080516515, 080516516, 080516517, 080516518, 080516519, 080516520, 080516521, 080516522, 080516523, 080516524, 080516525, 080516540, 080516541, 080516542, 080516543, 080516544, 080516545, 080516546, 080516547, 080516548, 080516549, 080516550, 080516551, 080516552, 080516553, 080516554, 080516555, 080516556, 080615001, 080615002, 080615003, 080615004, 080615005, 080615006, 080615007, 080615008, 080615009, 080615010, 080615011, 080615012, 080615013, 080615014, 080615015, 080615016, 080615017, 080615018, 080615019, 080615020, 080615021, 080615022, 080615023, 080615024, 080615025, 080615026, 080615027, 080615028, 080615029, 080615030, 080615031, 080615032, 080615033, 080615034, 080615035, 080615036, 080615037, 080615038, 080615039, 080615040, 080615042, 080615043, 080615044, 080615045, 080615046, 080615047, 080615048, 080615049, 080615050, 080615051, 080615052, 080615053, 080615054, 080615055, 080615056, 080615057, 080615058, 080615059, 080615060, 080615061, 080615062, 080615063, 080615064, 080615065, 080615066, 080615067, 080615068, 080615069, 080615070, 080615071, 080615072, 080615073, 080615074, 080615075, 080615076, 080615077, 080615078, 080615079, 080615080, 080615081, 080615082, 080615083, 080615084, 080615085, 080615086, 080615087, 080615088, 080615089, 080615090, 080615091, 080615092, 080615093, 080615094, 080615095, 080615096, 080615097, 080615098, 080615099, 080615100, 080615101, 080615102, 080615103, 080615104, 080615105, 080615106, 080615107, 080615108, 080615109, 080615110, 080615111, 080615112, 080615113, 080615114, 080615115, 080615116, 080615117, 080615118, 080615119, 080615120, 080615121, 080615122, 080615123, 080615124, 080615125, 080615126, 080615127, 080615128, 080615129, 080615130, 080615131, 080615132, 080615133, 080615134, 080615135, 080615136, 080615137, 080615138, 080615139, 080615140, 080615141, 080615142, 080615143, 080615144, 080615145, 080615146, 080615147, 080615148, 080615149, 080615150, 080615151, 080615152, 080615153, 080615154, 080615155, 080615156, 080615157, 080615158, 080615159, 080615160, 080615161, 080615162, 080615163, 080615164, 080615165, 080615166, 080615167, 080615168, 080615169, 080615170, 080615171, 080615172, 080615173, 080615174, 080615175, 080615176, 080615177, 080615178, 080615179, 080615180, 080615181, 080615182, 080615183, 080615184, 080615185, 080615186, 080615187, 080615188, 080615189, 080615190, 080615191, 080615192, 080615193, 080615194, 080615195, 080615196, 080615197, 080615198, 080615199, 080615200, 080615201, 080615202, 080615203, 080615204, 080615205, 080615206, 080615207, 080615208, 080615209, 080615210, 080615211, 080615212, 080615213, 080615214, 080615215, 080615216, 080615217, 080615218, 080615219, 080615220, 080615221, 080615222, 080615223, 080615224, 080615225, 080615226, 080615227, 080615228, 080615229, 080615230, 080615231, 080615232, 080615233, 080615234, 080615235, 080615236, 080615237, 080615238, 080615239, 080615240, 080615241, 080615242, 080615243, 080615244, 080615245, 080615246, 080615247, 080615248, 080615249, 080615250, 080615251, 080615252, 080615253, 080615254, 080615255, 080615256, 080615257, 080615258, 080615259, 080615260, 080615261, 080615262, 080615263, 080615264, 080615265, 080615266, 080615267, 080615268, 080615269, 080615270, 080615271, 080615272, 080615273, 080615274, 080615275, 080615276, 080615277, 080615278, 080615279, 080615280, 080615281, 080615282, 080615283, 080615284, 080615285, 080615286, 080615287, 080615288, 080615289, 080615290, 080615291, 080615292, 080615293, 080615294, 080615295, 080615296, 080615297, 080615298, 080615299, 080615300, 080615301, 080615302, 080615303, 080615304, 080615305, 080615306, 080615307, 080615308, 080615309, 080615310, 080615311, 080615312, 080615313, 080615314, 080615315, 080615316, 080615317, 080615318, 080615319, 080615320, 080615321, 080615322, 080615323, 080615324, 080615325, 080615326, 080615327, 080615328, 080615329, 080615330, 080615331, 080615332, 080615333, 080615334, 080615335, 080615336, 080615337, 080615338, 080615339, 080615340, 080615341, 080615342, 080615343, 080615344, 080615345, 080615346, 080615347, 080615348, 080615349, 080615350, 080615351, 080615352, 080615353, 080615354, 080615355, 080615356, 080615357, 080615358, 080615359, 080615360, 080615361, 080615362, 080615363, 080615364, 080615365, 080615366, 080615367, 080615368, 080615369, 080615370, 080615371, 080615372, 080615373, 080615374, 080615375, 080615376, 080615377, 080615378, 080615379, 080615380, 080615381, 080615382, 080615383, 080615384, 080615385, 080615386, 080615387, 080615388, 080615389, 080615390, 080615391, 080615392, 080615393, 080615394, 080615395, 080615396, 080615397, 080615398, 080615399, 080615400, 080615401, 080615402, 080615403, 080615404, 080615405, 080615406, 080615407, 080615408, 080615409, 080615410, 080615411, 080615412, 080615413, 080615414, 080615415, 080615416, 080615417, 080615418, 080615419, 080615420, 080615421, 080615422, 080615423, 080615424, 080615425, 080615426, 080615427, 080615428, 080615429, 080615430, 080615431, 080615432, 080615433, 080615434, 080615435, 080615436, 080615437, 080615438, 080615439, 080615440, 080615441, 080615442, 080615443, 080615444, 080615445, 080615446, 080615447, 080615448, 080615449, 080615450, 080615451, 080615452, 080615453, 080615454, 080615455, 080615456, 080615457, 080615458, 080615459, 080615460, 080615461, 080615462, 080615463, 080615464, 080615465, 080615466, 080615467, 080615468, 080615469, 080615470, 080615471, 080615472, 080615473, 080615474, 080615475, 080615476, 080615477, 080615478, 080615479, 080615480, 080615481, 080615482, 080615483, 080615484, 080615485, 080615486, 080615487, 080615488, 080615489, 080615490, 080615491, 080615492, 080615493, 080615494, 080615495, 080615496, 080615497, 080615498, 080615499, 080615500, 080615501, 080615502, 080615503, 080615504, 080615505, 080615506, 080615507, 080615508, 080615509, 080615510, 080615511, 080615512, 080615513, 080615514, 080615515, 080615516, 080615517, 080615518, 080615519, 080615520, 080615521, 080615522, 080615523, 080615524, 080615525, 080615526, 080615527, 080615528, 080615529, 080615530, 080615531, 080615532, 080615533, 080615534, 080615535, 080615536, 080615537, 080615538, 080615539, 080615540, 080615541, 080615542, 080615543, 080615544, 080615545, 080615546, 080615547, 080615548, 080615549, 080615550, 080615551, 080615552, 080615553, 080615554, 080615555, 080615556, 080615557, 080615558, 080615559, 080615560, 080615561, 080615562, 080615563, 080615564, 080615565, 080615566, 080615567, 080615568, 080615569, 080615570, 080615571, 080615572, 080615573, 080615574, 080615575, 080615576, 080615577, 080615578, 080615579, 080615594, 080615595, 080615596, 080615597, 080615598, 080615599, 080615600, 080615601, 080615602, 080615603, 080615604, 080615605, 080615606, 080615607, 080615608, 080615609, 080615610, 080615611, 080615612, 080615613, 080615614, 080615615, 080615616, 080615617, 080615618, 080615619, 080615620, 080615621, 080615622, 080615623, 080615624, 080615625, 080615626, 080615627, 080615628, 080615629, 080615630, 080615631, 080615632, 080615633, 080615634, 080615635, 080615636, 080615637, 080615638, 080615639, 080615640, 080615641, 080615642, 080615643, 080615644, 080615645, 080615646, 080615647, 080615648, 080615649, 080615650, 080615651, 080615652, 080615653, 080615654, 080615655, 080615656, 080615657, 080615658, 080615659, 080615660, 080615661, 080615662, 080615663, 080615664, 080615665, 080615666, 080615667, 080615668, 080615669, 080615670, 080615671, 080615672, 080615673, 080615674, 080615675, 080615676, 080615677, 080615678, 080615679, 080615680, 080615681, 080615682, 080615683, 080615684, 080615685, 080615686, 080615687, 080615688, 080615689, 080615690, 080615691, 080615692, 080615693, 080615694, 080615695, 080615696, 080615697, 080615698, 080615699, 080615700, 080615701, 080615702, 080615703, 080615704, 080615705, 080615706, 080615707, 080615708, 080615709, 080615710, 080615711, 080615712, 080615713, 080615714, 080615715, 080615716, 080615717, 080615718, 080615719, 080615720, 080615721, 080615722, 080615723, 080615724, 080615725, 080615726, 080615727, 080615728, 080615729, 080615730, 080615731, 080615732, 080615733, 080615734, 080615735, 080615736, 080615737, 080615738, 080615739, 080615740, 080615741, 080615742, 080615743, 080615744, 080615745, 080615746, 080615747, 080615748, 080615749, 080615750, 080615751, 080615752, 080615770, 080615771, 080615772, 080615773, 080615774, 080615775, 080615776, 080615777, 080615778, 080615779, 080615780, 080615781, 080615782, 080615783, 080615784, 080615785, 080615786, 080615787, 080615788, 080615789, 080615790, 080615791, 080615792, 080615793, 080615794, 080615795, 080615796, 080615797, 080615798, 080615799, 080615800, 080615801, 080615802, 080615803, 080615804, 080615805, 080615806, 080615807, 080615808, 080615809, 080615810, 080615811, 080615812, 080615813, 080615814, 080615815, 080615816, 080615817, 080615818, 080615819, 080615820, 080615821, 080615822, 080615823, 080615824, 080615825, 080615826, 080615827, 080615828, 080615829, 080615830, 080615831, 080615832, 080615833, 080615834, 080615835, 080615836, 080615837, 080615838, 080615839, 080615840, 080615841, 080615842, 080615843, 080615844, 080615845, 080615846, 080615847, 080615848, 080615849, 080615850, 080615851, 080615852, 080615853, 080615854, 080615855, 080615856, 080615857, 080615858, 080615859, 080615860, 080615861, 080615862, 080615863, 080615864, 080615865, 080615866, 080615867, 080615868, 080615869, 080615870, 080615871, 080615872, 080615873, 080615874, 080615875, 080615876, 080615877, 080615878, 080615879, 080615880, 080615881, 080615882, 080615883, 080615884, 080615885, 080615886, 080615887, 080615888, 080615889, 080615890, 080615891, 080615892, 080615893, 080615894, 080615895, 080615896, 080615897, 080615898, 080615899, 080615900, 080615901, 080615902, 080615903, 080615904, 080615905, 080615906, 080615907, 080615908, 080615909, 080615910, 080615911, 080615912, 080615913, 080615914, 080615915, 080615916, 080615917, 080615918, 080615919, 080615920, 080615921, 080615922, 080615923, 080615924, 080615925, 080615926, 080615927, 080615928, 080615929, 080615930, 080615931, 080615932, 080615933, 080615934, 080615935, 080615936, 080615937, 080615938, 080615939, 080615940, 080615941, 080615942, 080615943, 080615944, 080615945, 080615946, 080615947, 080615948, 080615949, 080615950, 080615951, 080615952, 080615953, 080615954, 080615955, 080615956, 080615957, 080615958, 080615959, 080615960, 080615961, 080615962, 080615963, 080615964, 080615965, 080615966, 080615967, 080615968, 080615969, 080615970, 080615971, 080615972, 080615973, 080615974, 080615975, 080615976, 080615977, 080615978, 080615979, 080615980, 080615981, 080615982, 080615983, 080615984, 080615985, 080615986, 080615987, 080615988, 080615989, 080615990, 080615991, 080615992, 080615993, 080615994, 080615995, 080615996, 080615997, 080615998, 080615999, 080616000, 080616001, 080616002, 080616003, 080616004, 080616005, 080616006, 080616007, 080616008, 080616009, 080616010, 080616011, 080616012, 080616013, 080616014, 080616015, 080616016, 080616017, 080616018, 080616019, 080616020, 080616021, 080616022, 080616023, 080616024, 080616025, 080616026, 080616027, 080616028, 080616029, 080616030, 080616031, 080616032, 080616033, 080616034, 080616035, 080616036, 080616037, 080616038, 080616039, 080616040, 080616041, 080616042, 080616043, 080616044, 080616045, 080616046, 080616047, 080616048, 080616049, 080616050, 080616051, 080616052, 080616053, 080616054, 080616055, 080616056, 080616057, 080616058, 080616059, 080616060, 080616061, 080616062, 080616063, 080616064, 080616065, 080616066, 080616067, 080616068, 080616069, 080616070, 080616071, 080616072, 080616074, 080616075, 080616076, 080616077, 080616078, 080616079, 080616080, 080616081, 080616082, 080616083, 080616084, 080616085, 080616086, 080616087, 080616088, 080616089, 080616090, 080616091, 080616092, 080616093, 080616094, 080616095, 080616096, 080616097, 080616098, 080616099, 080616100, 080616101, 080616102, 080616103, 080616104, 080616105, 080616106, 080616107, 080616108, 080616109, 080616110, 080616111, 080616112, 080616113, 080616114, 080616115, 080616116, 080616117, 080616118, 080616119, 080616120, 080616123, 080616124, 080616125, 080616126, 080616127, 080616128, 080616129, 080616130, 080616132, 080616133, 080616134, 080616135, 080616136, 080616139, 080616140, 080616141, 080616142, 080616143, 080616144, 080616145, 080616146, 080616147, 080616148, 080616149, 080616150, 080616151, 080616152, 080616153, 080616154, 080616155, 080616208, 080616209, 080616210, 080616211, 080616212, 080616213, 080616214, 080616215, 080616216, 080616217, 080616218, 080616219, 080616220, 080616221, 080616222, 080616223, 080616224, 080616225, 080616226, 080616227, 080616228, 080616229, 080616230, 080616231, 080616232, 080616233, 080616234, 080616235, 080616236, 080616237, 080616238, 080616239, 080616240, 080616241, 080616242, 080616243, 080616244, 080616245, 080616246, 080616247, 080616248, 080616249, 080616250, 080616251, 080616252, 080616253, 080616254, 080616255, 080616256, 080616257, 080616258, 080616259, 080616260, 080616261, 080616262, 080616263, 080616264, 080616265, 080616266, 080616267, 080616268, 080616269, 080616270, 080616271, 080616272, 080616273, 080616274, 080616275, 080616276, 080616277, 080616278, 080616279, 080616280, 080616281, 080616282, 080616283, 080616284, 080616285, 080616286, 080616287, 080616288, 080616289, 080616290, 080616291, 080616292, 080616293, 080616294, 080616295, 080616296, 080616297, 080616298, 080616299, 080616300, 080616301, 080616302, 080616303, 080616304, 080616305, 080616306, 080616307, 080616308, 080616309, 080616310, 080616311, 080616312, 080616313, 080616314, 080616315, 080616316, 080616317, 080616318, 080616319, 080616320, 080616321, 080616322, 080616323, 080616324, 080616325, 080616326, 080616327, 080616328, 080616329, 080616330, 080616331, 080616332, 080616333, 080616334, 080616335, 080616336, 080616337, 080616338, 080616339, 080616340, 080616341, 080616342, 080616343, 080616344, 080616345, 080616346, 080616347, 080616348, 080616349, 080616350, 080616351, 080616352, 080616353, 080616354, 080616355, 080616356, 080616357, 080616358, 080616359, 080616360, 080616361, 080616362, 080616363, 080616364, 080616365, 080616366, 080616367, 080616368, 080616369, 080616370, 080616371, 080616372, 080616373, 080616374, 080616375, 080616376, 080616377, 080616378, 080616379, 080616380, 080616381, 080616382, 080616383, 080616384, 080616385, 080616386, 080616387, 080616388, 080616389, 080616390, 080616391, 080616392, 080616393, 080616394, 080616395, 080616396, 080616397, 080616398, 080616399, 080616400, 080616401, 080616402, 080616403, 080616404, 080616405, 080616406, 080616407, 080616408, 080616409, 080616410, 080616411, 080616412, 080616413, 080616414, 080616415, 080616416, 080616417, 080616418, 080616419, 080616420, 080616421, 080616422, 080616423, 080616424, 080616425, 080616426, 080616427, 080616428, 080616446, 080616447, 080616448, 080616449, 080616450, 080616451, 080616452, 080616453, 080616455, 080616456, 080616458, 080616459, 080616460, 080616461, 080616462, 080616463, 080616465, 080616466, 080616467, 080616468, 080616469, 080616470, 080616472, 080616473, 080616474, 080616475, 080616476, 080616477, 080616478, 080616479, 080616480, 080616481, 080616482, 080616483, 080616484, 080616485, 080616486, 080616487, 080616488, 080616489, 080616490, 080616491, 080616492, 080616493, 080616494, 080616495, 080616496, 080616497, 080616498, 080616499, 080616500, 080616501, 080616502, 080616503, 080616504, 080616505, 080616506, 080616507, 080616508, 080616509, 080616510, 080616511, 080616512, 080616513, 080616514, 080616515, 080616516, 080616517, 080616518, 080616519, 080616520, 080616521, 080616522, 080616523, 080616524, 080616525, 080616526, 080616527, 080616528, 080616529, 080616530, 080616531, 080616532, 080616533, 080616534, 080616535, 080616536, 080616537, 080616538, 080616539, 080616540, 080616541, 080616542, 080616543, 080616544, 080616545, 080616546, 080616547, 080616548, 080616549, 080616550, 080616551, 080616552, 080616553, 080616554, 080616555, 080616556, 080616557, 080616558, 080616559, 080616560, 080616561, 080616562, 080616563, 080616564, 080616565, 080616566, 080616567, 080616568, 080616569, 080616570, 080616571, 080616572, 080616573, 080616574, 080616575, 080616576, 080616577, 080616578, 080616579, 080616580, 080616581, 080616582, 080616583, 080616584, 080616585, 080616586, 080616587, 080616588, 080616589, 080616590, 080616591, 080616592, 080616593, 080616594, 080616595, 080616596, 080616597, 080616598, 080616599, 080616600, 080616601, 080715001, 080715002, 080715003, 080715004, 080715005, 080715006, 080715007, 080715008, 080715009, 080715010, 080715011, 080715012, 080715013, 080715014, 080715015, 080715016, 080715017, 080715018, 080715019, 080715020, 080715021, 080715022, 080715023, 080715024, 080715025, 080715026, 080715027, 080715028, 080715029, 080715030, 080715031, 080715032, 080715033, 080715034, 080715035, 080715036, 080715037, 080715038, 080715039, 080715040, 080715041, 080715042, 080715043, 080715044, 080715045, 080715046, 080715047, 080715048, 080715049, 080715050, 080715051, 080715052, 080715053, 080715054, 080715055, 080715056, 080715057, 080715058, 080715059, 080715060, 080715061, 080715062, 080715063, 080715064, 080715065, 080715066, 080715067, 080715068, 080715069, 080715070, 080715071, 080715072, 080715073, 080715074, 080715076, 080715077, 080715078, 080715079, 080715080, 080715081, 080715082, 080715083, 080715084, 080715085, 080715086, 080715087, 080715088, 080715089, 080715090, 080715091, 080715092, 080715093, 080715094, 080715095, 080715096, 080715097, 080715098, 080715099, 080715100, 080715101, 080715102, 080715103, 080715104, 080715105, 080715106, 080715107, 080715108, 080715109, 080715110, 080715111, 080715112, 080715113, 080715114, 080715115, 080715116, 080715117, 080715118, 080715119, 080715120, 080715121, 080715122, 080715123, 080715124, 080715125, 080715126, 080715127, 080715128, 080715129, 080715130, 080715131, 080715132, 080715133, 080715134, 080715135, 080715136, 080715137, 080715138, 080715139, 080715140, 080715141, 080715142, 080715143, 080715144, 080715145, 080715146, 080715147, 080715148, 080715149, 080715150, 080715151, 080715152, 080715153, 080715154, 080715155, 080715156, 080715157, 080715158, 080715159, 080715160, 080715161, 080715162, 080715163, 080715164, 080715165, 080715166, 080715167, 080715168, 080715169, 080715170, 080715171, 080715172, 080715173, 080715174, 080715175, 080715176, 080715177, 080715178, 080715179, 080715180
080715181, 080715182, 080715183, 080715184, 080715185, 080715186, 080715187, 080715188, 080715189, 080715190, 080715191, 080715192, 080715193, 080715194, 080715195, 080715196, 080715197, 080715198, 080715199, 080715200, 080715201, 080715202, 080715203, 080715204, 080715205, 080715206, 080715207, 080715208, 080715209, 080715210, 080715211, 080715212, 080715213, 080715214, 080715215, 080715216, 080715217, 080715218, 080715219, 080715220, 080715221, 080715222, 080715223, 080715224, 080715225, 080715226, 080715227, 080715228, 080715229, 080715230, 080715231, 080715232, 080715233, 080715234, 080715235, 080715236, 080715237, 080715238, 080715239, 080715240, 080715241, 080715242, 080715243, 080715244, 080715245, 080715246, 080715247, 080715248, 080715249, 080715250, 080715251, 080715252, 080715253, 080715254, 080715255, 080715256, 080715257, 080715258, 080715259, 080715260, 080715261, 080715262, 080715263, 080715264, 080715265, 080715266, 080715267, 080715268, 080715269, 080715270, 080715271, 080715272, 080715273, 080715274, 080715275, 080715276, 080715277, 080715278, 080715279, 080715280, 080715281, 080715282, 080715283, 080715284, 080715285, 080715286, 080715287, 080715288, 080715289, 080715290, 080715291, 080715292, 080715293, 080715294, 080715295, 080715296, 080715297, 080715298, 080715299, 080715300, 080715301, 080715302, 080715303, 080715304, 080715305, 080715306, 080715307, 080715308, 080715309, 080715332, 080715333, 080715334, 080715335, 080715336, 080715337, 080715338, 080715339, 080715340, 080715341, 080715342, 080715343, 080715344, 080715345, 080715346, 080715347, 080715348, 080715349, 080715350, 080715351, 080715352, 080715353, 080715354, 080715355, 080715356, 080715357, 080715358, 080715359, 080715360, 080715361, 080715362, 080715363, 080715364, 080715365, 080715366, 080715367, 080715368, 080715369, 080715370, 080715371, 080715372, 080715373, 080715374, 080715375, 080715376, 080715377, 080715378, 080715379, 080715380, 080715381, 080715382, 080715383, 080715384, 080715385, 080715386, 080715387, 080715388, 080715389, 080715390, 080715391, 080715392, 080715393, 080715394, 080715395, 080715422, 080715423, 080715424, 080715425, 080715426, 080715427, 080715428, 080715429, 080715430, 080715431, 080715432, 080715433, 080715434, 080715435, 080715436, 080715437, 080715438, 080715439, 080715440, 080715441, 080715442, 080715443, 080715444, 080715445, 080715446, 080715447, 080715448, 080715449, 080715450, 080715451, 080715452, 080715453, 080715454, 080715455, 080715456, 080715457, 080715458, 080715459, 080715460, 080715461, 080715462, 080715463, 080715464, 080715465, 080715466, 080715467, 080715468, 080715469, 080715470, 080715471, 080715472, 080715473, 080715474, 080715475, 080715476, 080715477, 080715478, 080715479, 080715480, 080715481, 080715482, 080715483, 080715484, 080715485, 080715486, 080715487, 080715488, 080715489, 080715490, 080715491, 080715492, 080715493, 080715494, 080715495, 080715496, 080715497, 080715498, 080715499, 080715500, 080715501, 080715502, 080715503, 080715504, 080715505, 080715506, 080715507, 080715508, 080715512, 080715513, 080715514, 080715515, 080715516, 080715517, 080715518, 080715519, 080715520, 080715521, 080715522, 080715523, 080715524, 080715525, 080715526, 080715527, 080715528, 080715529, 080715530, 080715531, 080715532, 080715533, 080715534, 080715535, 080715536, 080715537, 080715538, 080715539, 080715540, 080715541, 080715542, 080715543, 080715544, 080715545, 080715546, 080715547, 080715548, 080715549, 080715550, 080715551, 080715552, 080715553, 080715554, 080715555, 080715556, 080715557, 080715558, 080715559, 080715560, 080715561, 080715562, 080715563, 080715564, 080715565, 080715566, 080715567, 080715568, 080715569, 080715570, 080715571, 080715572, 080715573, 080715574, 080715575, 080715576, 080715577, 080715578, 080715579, 080715580, 080715581, 080715582, 080715583, 080715584, 080715585, 080715586, 080715587, 080715588, 080715589, 080715590, 080715591, 080715592, 080715593, 080715594, 080715595, 080715596, 080715597, 080715598, 080715599, 080715600, 080715601, 080715602, 080715603, 080715604, 080715605, 080715606, 080715607, 080715612, 080715613, 080715614, 080715615, 080715616, 080715617, 080715618, 080715619, 080715620, 080715621, 080715622, 080715623, 080715624, 080715625, 080715626, 080715627, 080715628, 080715629, 080715630, 080715631, 080715632, 080715633, 080715634, 080715635, 080715636, 080715637, 080715638, 080715639, 080715640, 080715641, 080715642, 080715643, 080715644, 080715645, 080715646, 080715647, 080715648, 080715649, 080715650, 080715651, 080715652, 080715653, 080715654, 080715655, 080715656, 080715657, 080715658, 080715659, 080715660, 080715661, 080715662, 080715663, 080715664, 080715665, 080715666, 080715667, 080715668, 080715669, 080715670, 080715671, 080715672, 080715673, 080715674, 080715675, 080715676, 080715677, 080715678, 080715679, 080715680, 080715681, 080715682, 080715683, 080715684, 080715685, 080715686, 080715687, 080715688, 080715689, 080715690, 080715691, 080715692, 080715693, 080715694, 080715695, 080715696, 080715697, 080715698, 080715699, 080715700, 080715701, 080715702, 080715703, 080715704, 080715705, 080715706, 080715707, 080715708, 080715709, 080715710, 080715711, 080715712, 080715713, 080715714, 080715715, 080715716, 080715717, 080715718, 080715719, 080715720, 080715721, 080715722, 080715723, 080715724, 080715725, 080715726, 080715727, 080715728, 080715729, 080715730, 080715731, 080715732, 080715733, 080715734, 080715735, 080715736, 080715737, 080715738, 080715739, 080715740, 080715741, 080715742, 080715743, 080715744, 080715745, 080715746, 080715747, 080715748, 080715749, 080715750, 080715751, 080715752, 080715753, 080715754, 080715755, 080715756, 080715757, 080715758, 080715759, 080715760, 080715761, 080715762, 080715763, 080715764, 080715765, 080715766, 080715767, 080715768, 080715769, 080715770, 080715771, 080715772, 080715773, 080715774, 080715775, 080715776, 080715777, 080715778, 080715779, 080715780, 080715781, 080715782, 080715783, 080715784, 080715785, 080715794, 080715795, 080715796, 080715797, 080715798, 080715799, 080715800, 080715801, 080715802, 080715803, 080715804, 080715805, 080715806, 080715807, 080715808, 080715809, 080715810, 080715811, 080715812, 080715813, 080715814, 080715815, 080715816, 080715817, 080715818, 080715819, 080715820, 080715821, 080715822, 080715823, 080715824, 080715825, 080715826, 080715827, 080715828, 080715829, 080715830, 080715831, 080715832, 080715833, 080715834, 080715835, 080715836, 080715837, 080715838, 080715839, 080715840, 080715841, 080715842, 080715843, 080715844, 080715845, 080715846, 080715847, 080715848, 080715849, 080715850, 080715851, 080715852, 080715853, 080715854, 080715855, 080715856, 080715857, 080715858, 080715859, 080715860, 080715861, 080715862, 080715863, 080715864, 080715865, 080715866, 080715867, 080715868, 080715869, 080715870, 080715871, 080715872, 080715873, 080715874, 080715875, 080715876, 080715877, 080715878, 080715879, 080715880, 080715881, 080715882, 080715883, 080715884, 080715885, 080715886, 080715887, 080715888, 080715889, 080715890, 080715891, 080715892, 080715893, 080715894, 080715895, 080715896, 080715897, 080715898, 080715899, 080715900, 080715901, 080715902, 080715903, 080715904, 080715905, 080715906, 080715907, 080715908, 080715909, 080715910, 080715911, 080715912, 080715913, 080715914, 080716041, 080716042, 080716043, 080716044, 080716045, 080716046, 080716047, 080716048, 080716049, 080716050, 080716051, 080716052, 080716053, 080716054, 080716055, 080716056, 080716057, 080716058, 080716059, 080716060, 080716061, 080716062, 080716063, 080716064, 080716065, 080716066, 080716067, 080716068, 080716069, 080716070, 080716071, 080716072, 080716073, 080716074, 080716075, 080716076, 080716077, 080716078, 080716079, 080716080, 080716081, 080716082, 080716083, 080716084, 080716085, 080716086, 080716087, 080716088, 080716089, 080716090, 080716093, 080716094, 080716095, 080716096, 080716097, 080716098, 080716099, 080716100, 080716101, 080716102, 080716103, 080716104, 080716105, 080716106, 080716107, 080716108, 080716109, 080716110, 080716111, 080716112, 080716113, 080716114, 080716115, 080716116, 080716117, 080716118, 080716119, 080716120, 080716121, 080716122, 080716123, 080716124, 080716125, 080716126, 080716127, 080716128, 080716129, 080716130, 080716131, 080716132, 080716133, 080716134, 080716135, 080716136, 080716137, 080716138, 080716139, 080716140, 080716141, 080716142, 080716145, 080716146, 080716147, 080716148, 080716149, 080716150, 080716151, 080716152, 080716153, 080716154, 080716155, 080716156, 080716157, 080716158, 080716159, 080716160, 080716161, 080716162, 080716163, 080716164, 080716165, 080716166, 080716167, 080716168, 080716169, 080716170, 080716171, 080716172, 080716173, 080716174, 080716175, 080716176, 080716177, 080716178, 080716179, 080716180, 080716181, 080716182, 080716183, 080716184, 080716185, 080716186, 080716187, 080716188, 080716189, 080716190, 080716191, 080716192, 080716193, 080716194, 080716195, 080716196, 080716197, 080716198, 080716199, 080716200, 080716201, 080716202, 080716203, 080716204, 080716205, 080716206, 080716207, 080716208, 080716209, 080716210, 080716211, 080716212, 080716213, 080716214, 080716215, 080716216, 080716217, 080716218, 080716219, 080716220, 080716221, 080716222, 080716223, 080716224, 080716225, 080716226, 080716227, 080716228, 080716229, 080716230, 080716231, 080716232, 080716233, 080716234, 080716235, 080716236, 080716237, 080716238, 080716239, 080716240, 080716241, 080716242, 080716243, 080716244, 080716245, 080716246, 080716247, 080716248, 080716249, 080716250, 080716251, 080716252, 080716253, 080716254, 080716255, 080716256, 080716257, 080716258, 080716259, 080716260, 080716261, 080716262, 080716263, 080716264, 080716265, 080716266, 080716267, 080716268, 080716269, 080716270, 080716271, 080716272, 080716273, 080716274, 080716275, 080716276, 080716277, 080716278, 080716279, 080716280, 080716281, 080716282, 080716283, 080716284, 080716285, 080716286, 080716287, 080716288, 080716289, 080716290, 080716291, 080716292, 080716293, 080716294, 080716307, 080716308, 080716309, 080716310, 080716311, 080716312, 080716313, 080716314, 080716315, 080716316, 080716317, 080716318, 080716319, 080716320, 080716321, 080716322, 080716323, 080716324, 080716325, 080716326, 080716327, 080716328, 080716329, 080716330, 080716331, 080716332, 080716333, 080716334, 080716335, 080716336, 080716337, 080716338, 080716339, 080716340, 080716341, 080716342, 080716343, 080716344, 080716345, 080716346, 080716347, 080716348, 080716349, 080716350, 080716351, 080716352, 080716353, 080716354, 080716355, 080716356, 080716357, 080716358, 080716359, 080716360, 080716361, 080716362, 080716363, 080716364, 080716365, 080716366, 080716367, 080716368, 080716369, 080716370, 080716371, 080716372, 080716373, 080716374, 080716375, 080716376, 080716377, 080716378, 080716379, 080716380, 080716381, 080716382, 080716383, 080716384, 080716385, 080716386, 080716387, 080716388, 080716389, 080716390, 080716391, 080716392, 080716393, 080716394, 080716395, 080716396, 080815002, 080815003, 080815004, 080815005, 080815006, 080815007, 080815008, 080815009, 080815010, 080815011, 080815012, 080815013, 080815014, 080815015, 080815016, 080815017, 080815018, 080815019, 080815020, 080815021, 080815022, 080815023, 080815024, 080815025, 080815026, 080815027, 080815028, 080815029, 080815030, 080815031, 080815032, 080815033, 080815034, 080815035, 080815036, 080815037, 080815038, 080815039, 080815040, 080815041, 080815042, 080815043, 080815044, 080815045, 080815046, 080815047, 080815048, 080815049, 080815050, 080815051, 080815052, 080815053, 080815054, 080815055, 080815056, 080815057, 080815058, 080815059, 080815060, 080815061, 080815062, 080815063, 080815064, 080815065, 080815066, 080815067, 080815068, 080815069, 080815070, 080815071, 080815072, 080815073, 080815074, 080815075, 080815076, 080815077, 080815078, 080815079, 080815080, 080815081, 080815082, 080815083, 080815084, 080815085, 080815086, 080815087, 080815088, 080815089, 080815090, 080815091, 080815092, 080815093, 080815094, 080815095, 080815096, 080815097, 080815098, 080815099, 080815100, 080815101, 080815102, 080815103, 080815104, 080815105, 080815106, 080815107, 080815108, 080815109, 080815110, 080815111, 080815112, 080815113, 080815114, 080815115, 080815116, 080815117, 080815118, 080815119, 080815120, 080815121, 080815122, 080815123, 080815124, 080815125, 080815126, 080815127, 080815128, 080815129, 080815130, 080815131, 080815132, 080815133, 080815134, 080815135, 080815136, 080815137, 080815138, 080815139, 080815140, 080815141, 080815142, 080815143, 080815144, 080815145, 080815146, 080815147, 080815148, 080815149, 080815150, 080815151, 080815152, 080815153, 080815154, 080815155, 080815156, 080815157, 080815158, 080815159, 080815160, 080815161, 080815162, 080815163, 080815164, 080815165, 080815166, 080815167, 080815168, 080815169, 080815170, 080815171, 080815172, 080815173, 080815174, 080815175, 080815176, 080815177, 080815178, 080815179, 080815180, 080815181, 080815182, 080815183, 080815184, 080815185, 080815186, 080815187, 080815188, 080815189, 080815190, 080815191, 080815192, 080815193, 080815194, 080815195, 080815196, 080815197, 080815198, 080815199, 080815200, 080815201, 080815202, 080815203, 080815204, 080815205, 080815207, 080815208, 080815209, 080815210, 080815211, 080815212, 080815213, 080815214, 080815215, 080815216, 080815217, 080815218, 080815219, 080815220, 080815221, 080815222, 080815223, 080815224, 080815225, 080815226, 080815227, 080815228, 080815229, 080815230, 080815231, 080815232, 080815233, 080815234, 080815235, 080815236, 080815237, 080815238, 080815239, 080815240, 080815241, 080815242, 080815243, 080815244, 080815245, 080815246, 080815247, 080815248, 080815249, 080815250, 080815251, 080815252, 080815253, 080815254, 080815255, 080815256, 080815257, 080815258, 080815259, 080815260, 080815261, 080815262, 080815263, 080815264, 080815265, 080815266, 080815267, 080815268, 080815269, 080815270, 080815271, 080815272, 080815273, 080815274, 080815275, 080815276, 080815277, 080815278, 080815279, 080815280, 080815281, 080815282, 080815283, 080815284, 080815285, 080815286, 080815287, 080815288, 080815289, 080815290, 080815291, 080815292, 080815293, 080815294, 080815295, 080815296, 080815297, 080815298, 080815299, 080815300, 080815301, 080815302, 080815303, 080815304, 080815305, 080815306, 080815307, 080815308, 080815309, 080815310, 080815311, 080815312, 080815313, 080815314, 080815315, 080815316, 080815317, 080815318, 080815319, 080815320, 080815321, 080815322, 080815323, 080815324, 080815325, 080815326, 080815327, 080815328, 080815329, 080815330, 080815331, 080815332, 080815333, 080815334, 080815335, 080815336, 080815337, 080815338, 080815339, 080815340, 080815341, 080815342, 080815343, 080815344, 080815345, 080815346, 080815347, 080815348, 080815349, 080815350, 080815351, 080815352, 080815353, 080815354, 080815355, 080815356, 080815357, 080815358, 080815359, 080815360, 080815361, 080815362, 080815363, 080815364, 080815365, 080815366, 080815367, 080815368, 080815369, 080815370, 080815371, 080815372, 080815373, 080815374, 080815375, 080815376, 080815377, 080815378, 080815379, 080815380, 080815381, 080815382, 080815383, 080815384, 080815385, 080815386, 080815387, 080815388, 080815389, 080815390, 080815391, 080815392, 080815393, 080815394, 080815395, 080815396, 080815397, 080815398, 080815399, 080815400, 080815401, 080815402, 080815403, 080815404, 080815405, 080815406, 080815407, 080815408, 080815409, 080815410, 080815411, 080815412, 080815413, 080815414, 080815415, 080815416, 080815417, 080815418, 080815419, 080815420, 080815421, 080815422, 080815423, 080815424, 080815425, 080815426, 080815427, 080815428, 080815429, 080815430, 080815431, 080815432, 080815433, 080815434, 080815435, 080815436, 080815437, 080815438, 080815439, 080815440, 080815441, 080815442, 080815443, 080815444, 080815445, 080815446, 080815447, 080815448, 080815449, 080815450, 080815451, 080815452, 080815453, 080815454, 080815455, 080815456, 080815457, 080815458, 080815459, 080815460, 080815461, 080815462, 080815463, 080815464, 080815465, 080815466, 080815467, 080815468, 080815469, 080815470, 080815471, 080815472, 080815473, 080815474, 080815475, 080815476, 080815477, 080815478, 080815479, 080815480, 080815481, 080815482, 080815483, 080815484, 080815485, 080815486, 080815487, 080815488, 080815489, 080815490, 080815491, 080815492, 080815493, 080815494, 080815495, 080815496, 080815497, 080815498, 080815499, 080815500, 080815501, 080815502, 080815503, 080815504, 080815505, 080815506, 080815507, 080815508, 080815509, 080815510, 080815511, 080815512, 080815513, 080815514, 080815515, 080815516, 080815517, 080815518, 080815519, 080815520, 080815521, 080815522, 080815523, 080815524, 080815525, 080815526, 080815527, 080815528, 080815529, 080815530, 080815531, 080815532, 080815533, 080815534, 080815535, 080815536, 080815537, 080815538, 080815539, 080815540, 080815541, 080815542, 080815543, 080815544, 080815545, 080815546, 080815547, 080815548, 080815549, 080815550, 080815551, 080815552, 080815553, 080815554, 080815555, 080815556, 080815557, 080815558, 080815559, 080815560, 080815561, 080815562, 080815563, 080815564, 080815565, 080815566, 080815567, 080815568, 080815569, 080815570, 080815571, 080815572, 080815573, 080815574, 080815575, 080815576, 080815577, 080815578, 080815579, 080815580, 080815581, 080815582, 080815583, 080815584, 080815585, 080815586, 080815587, 080815588, 080815589, 080815590, 080815591, 080815592, 080815593, 080815594, 080815595, 080815596, 080815597, 080815598, 080815599, 080815600, 080815601, 080815602, 080815603, 080815604, 080815605, 080815606, 080815607, 080815608, 080815609, 080815610, 080815611, 080815612, 080815613, 080815614, 080815615, 080815616, 080815617, 080815618, 080815619, 080815620, 080815621, 080815622, 080815623, 080815624, 080815625, 080815626, 080815627, 080815628, 080815629, 080815630, 080815631, 080815633, 080815634, 080815635, 080815636, 080815637, 080815638, 080815639, 080815640, 080815641, 080815642, 080815643, 080815644, 080815645, 080815646, 080815647, 080815648, 080815649, 080815650, 080815651, 080815652, 080815653, 080815654, 080815655, 080815656, 080815657, 080815658, 080815659, 080815660, 080815661, 080815662, 080815663, 080815664, 080815665, 080815666, 080815667, 080815668, 080815669, 080815670, 080815671, 080815672, 080815673, 080815674, 080815675, 080815676, 080815677, 080815678, 080815679, 080815680, 080815681, 080815682, 080815683, 080815684, 080815685, 080815686, 080815687, 080815688, 080815689, 080815690, 080815691, 080815692, 080815693, 080815694, 080815695, 080815696, 080815697, 080815698, 080815699, 080815700, 080815701, 080815702, 080815703, 080815704, 080815705, 080815706, 080815707, 080815708, 080815709, 080815710, 080815711, 080815712, 080815713, 080815714, 080815715, 080815716, 080815717, 080815718, 080815719, 080815720, 080815721, 080815722, 080815723, 080815724, 080815725, 080815726, 080815727, 080815728, 080815729, 080815730, 080815731, 080815732, 080815733, 080815734, 080815735, 080815736, 080815737, 080815738, 080815739, 080815740, 080815741, 080815742, 080815743, 080815744, 080815745, 080815746, 080815747, 080815748, 080815749, 080815750, 080815751, 080815752, 080815753, 080815754, 080815755, 080815756, 080815757, 080815758, 080815759, 080815760, 080815761, 080815762, 080815763, 080815764, 080815765, 080815766, 080815767, 080815768, 080815769, 080815770, 080815771, 080815772, 080815773, 080815774, 080815775, 080815776, 080815777, 080815778, 080815779, 080815780, 080815781, 080815782, 080815783, 080815784, 080815785, 080815786, 080815787, 080815788, 080815789, 080815790, 080815791, 080815792, 080815793, 080815794, 080815795, 080815796, 080815797, 080815798, 080815799, 080815800, 080815801, 080815802, 080815803, 080815804, 080815805, 080815806, 080815807, 080815808, 080815809, 080815810, 080815811, 080815812, 080815813, 080815814, 080815815, 080815816, 080815817, 080815818, 080815819, 080815820, 080815821, 080815822, 080815823, 080815824, 080815825, 080815826, 080815827, 080815828, 080815829, 080815830, 080815831, 080815832, 080815833, 080815835, 080815836, 080815837, 080815838, 080815839, 080815840, 080815841, 080815842, 080815843, 080815845, 080815846, 080815847, 080815848, 080815849, 080815850, 080815851, 080815852, 080815853, 080815854, 080815855, 080815856, 080815857, 080815858, 080815859, 080815860, 080815861, 080815862, 080815863, 080815864, 080815865, 080815866, 080815867, 080815868, 080815869, 080815870, 080815871, 080815872, 080815873, 080815874, 080815875, 080815876, 080815877, 080815878, 080815879, 080815880, 080815881, 080815882, 080815883, 080815884, 080815885, 080815886, 080815887, 080815888, 080815889, 080815890, 080815891, 080815892, 080815893, 080815894, 080815895, 080815896, 080815897, 080815898, 080815899, 080815900, 080815901, 080815902, 080815903, 080815904, 080815905, 080815906, 080815907, 080815908, 080815909, 080815910, 080815911, 080815912, 080815913, 080815914, 080815915, 080815916, 080815917, 080815918, 080815919, 080815920, 080815921, 080815922, 080815923, 080815924, 080815925, 080815926, 080815927, 080815928, 080815929, 080815930, 080815931, 080815932, 080815933, 080815934, 080815935, 080815936, 080815937, 080815938, 080815939, 080815940, 080815941, 080815942, 080815943, 080815944, 080815945, 080815946, 080815947, 080815948, 080815949, 080815950, 080815951, 080815952, 080815953, 080815954, 080815955, 080815956, 080815957, 080815958, 080815959, 080815960, 080815961, 080815962, 080815963, 080815964, 080815965, 080815966, 080815967, 080815968, 080815969, 080815970, 080815971, 080815972, 080815973, 080815974, 080815975, 080815976, 080815977, 080815978, 080815979, 080815980, 080815981, 080815982, 080815983, 080815984, 080815985, 080815986, 080815987, 080815988, 080815989, 080815990, 080815991, 080815992, 080815993, 080815994, 080815995, 080815996, 080815997, 080815998, 080815999, 080816000, 080816001, 080816002, 080816003, 080816004, 080816005, 080816006, 080816007, 080816008, 080816009, 080816010, 080816011, 080816012, 080816013, 080816014, 080816015, 080816016, 080816017, 080816018, 080816019, 080816020, 080816021, 080816022, 080816023, 080816024, 080816025, 080816026, 080816027, 080816028, 080816029, 080816030, 080816031, 080816032, 080816033, 080816034, 080816035, 080816036, 080816037, 080816038, 080816039, 080816040, 080816041, 080816042, 080816043, 080816044, 080816045, 080816046, 080816047, 080816048, 080816049, 080816050, 080816051, 080816052, 080816053, 080816054, 080816055, 080816056, 080816057, 080816058, 080816059, 080816060, 080816061, 080816062, 080816063, 080816064, 080816065, 080816066, 080816067, 080816068, 080816069, 080816070, 080816071, 080816072, 080816073, 080816074, 080816075, 080816076, 080816077, 080816078, 080816079, 080816080, 080816081, 080816082, 080816083, 080816084, 080816085, 080816086, 080816087, 080816088, 080816089, 080816090, 080816091, 080816092, 080816093, 080816094, 080816095, 080816096, 080816097, 080816098, 080816099, 080816100, 080816101, 080816102, 080816103, 080816104, 080816105, 080816106, 080816107, 080816108, 080816109, 080816110, 080816111, 080816112, 080816113, 080816114, 080816115, 080816116, 080816117, 080816118, 080816119, 080816120, 080816121, 080816122, 080816123, 080816124, 080816125, 080816126, 080816127, 080816128, 080816129, 080816130, 080816131, 080816132, 080816133, 080816134, 080816135, 080816136, 080816137, 080816138, 080816139, 080816140, 080816141, 080816142, 080816143, 080816144, 080816145, 080816146, 080816147, 080816148, 080816149, 080816150, 080816151, 080816152, 080816153, 080816154, 080816155, 080816156, 080816157, 080816158, 080816159, 080816160, 080816161, 080816162, 080816163, 080816164, 080816165, 080816166, 080816167, 080816168, 080816169, 080816170, 080816171, 080816172, 080816173, 080816174, 080816176, 080816177, 080816178, 080816179, 080816180, 080816181, 080816182, 080816183, 080816184, 080816185, 080816186, 080816187, 080816188, 080816189, 080816190, 080816191, 080816192, 080816193, 080816194, 080816195, 080816196, 080816197, 080816198, 080816199, 080816200, 080816201, 080816202, 080816203, 080816204, 080816205, 080816206, 080816207, 080816208, 080816209, 080816210, 080816211, 080816212, 080816213, 080816214, 080816215, 080816216, 080816217, 080816218, 080816219, 080816220, 080816221, 080816222, 080816223, 080816224, 080816225, 080816226, 080816227. 080816228, 080816229, 080816230, 080816231, 080816232, 080816233, 080816234, 080816235, 080816236, 080816237, 080816238, 080816239, 080816240, 080816241, 080816242, 080816243, 080816244, 080816245, 080816246, 080816247, 080816248, 080816249, 080816250, 080816251, 080816252, 080816253, 080816254, 080816255, 080816256, 080816257, 080816258, 080816273, 080816274, 080816275, 080816276, 080816277, 080816278, 080816279, 080816280, 080816281, 080816282, 080816283, 080816284, 080816285, 080816286, 080816287, 080816288, 080816289, 080816290, 080816291, 080816292, 080816293, 080816294, 080816295, 080816296, 080816297, 080816298, 080816299, 080816300, 080816301, 080816302, 080816303, 080816304, 080816305, 080816306, 080816307, 080816308, 080816309, 080816310, 080816311, 080816312, 080816314, 080816315, 080816316, 080816317, 080816318, 080816319, 080816320, 080816321, 080816322, 080816323, 080816324, 080816325, 080816326, 080816327, 080816328, 080816329, 080816330, 080816331, 080816332, 080816333, 080816334, 080816335, 080816336, 080816337, 080816338, 080816339, 080816340, 080816341, 080816342, 080816343, 080816344, 080816345, 080816346, 080816347, 080816348, 080816349, 080816350, 080816351, 080816352, 080816353, 080816354, 080816355, 080816356, 080816357, 080816358, 080816359, 080816360, 080816361, 080816362, 080816363, 080816364, 080816365, 080816366, 080816367, 080816368, 080816391, 080816392, 080816393, 080816394, 080816395, 080816396, 080816397, 080816398, 080816399, 080816400, 080816401, 080816402, 080816403, 080816404, 080816405, 080816406, 080816407, 080816408, 080816409, 080816410, 080816411, 080816412, 080816413, 080816414, 080816415, 080816416, 080816417, 080816418, 080816419, 080816420, 080816421, 080816422, 080816423, 080816424, 080816425, 080816426, 080816427, 080816428, 080816429, 080816430, 080816431, 080816432, 080816433, 080816434, 080816435, 080816436, 080816437, 080816438, 080816439, 080816440, 080816441, 080816442, 080816443, 080816444, 080816445, 080816446, 080816447, 080816448, 080816449, 080816450, 080816451, 080816452, 080816453, 080816454, 080816455, 080816456, 080816457, 080816458, 080816459, 080816460, 080816461, 080816462, 080816463, 080816464, 080816465, 080816466, 080816467, 080816468, 080816469, 080816470, 080816471, 080816472, 080816473, 080816474, 080816475, 080816476, 080816477, 080816478, 080816479, 080816480, 080816481, 080816482, 080816483, 080816484, 080816485, 080816486, 080816487, 080816488, 080816489, 080816490, 080816491, 080816492, 080816493, 080816494, 080816495, 080816496, 080816497, 080816498, 080816499, 080816501, 080816502, 080816503, 080816504, 080816505, 080816506, 080816507, 080816508, 080816509, 080816510, 080816511, 080816512, 080816513, 080816514, 080816515, 080816516, 080816517, 080816518, 080816519, 080816520, 080816521, 080816522, 080816523, 080816524, 080816525, 080816526, 080816527, 080816528, 080816529, 080816530, 080816531, 080816532, 080816533, 080816534, 080816535, 080816536, 080816537, 080816538, 080816539, 080816540, 080816541, 080816542, 080816543, 080816544, 080816545, 080816546, 080816547, 080816548, 080816549, 080816550, 080816551, 080816552, 080816553, 080816554, 080816555, 080816556, 080816557, 080816558, 080816559, 080816560, 080816561, 080816562, 080816563, 080816564, 080816565, 080816566, 080816567, 080816568, 080816569, 080816570, 080816571, 080816572, 080816573, 080816574, 080816575, 080816576, 080816577, 080816578, 080816579, 080816580, 080816581, 080816582, 080816583, 080816584, 080816585, 080816586, 080816587, 080816588, 080816589, 080816590, 080816591, 080816592, 080816593, 080816594, 080816595, 080816596, 080816597, 080816598, 080816599, 080816600, 080816601, 080816602, 080816603, 080816604, 080816605, 080816606, 080816607, 080816608, 080816609, 080816610, 080816611, 080816612, 080816613, 080816614, 080816615, 080816616, 080816617, 080816618, 080816619, 080816620, 080816621, 080816622, 080816623, 080816624, 080816625, 080816626, 080816627, 080816628, 080816629, 080816630, 080816631, 080816632, 080816633, 080816634, 080816635, 080816636, 080816637, 080816638, 080816639, 080816640, 080816641, 080816642, 080816643, 080816644, 080816645, 080816646, 080816647, 080816648, 080816649, 080816650, 080816651, 080816652, 080816653, 080816654, 080816655, 080816656, 080816657, 080816675, 080816676, 080816677, 080816678, 080816679, 080816680, 080816681, 080816682, 080816683, 080816684, 080816685, 080816686, 080816687, 080816688, 080816689, 080816690, 080816691, 080816692, 080816693, 080816694, 080816695, 080816696, 080816697, 080816698, 080816699, 080816700, 080816701, 080816702, 080816708, 080816709, 080816710, 080816711, 080816712, 080816713, 080816714, 080816715, 080816716, 080816717, 080816718, 080816719, 080816720, 080816721, 080816722, 080816723, 080816724, 080816725, 080816726, 080816727, 080816728, 080816729, 080816730, 080816731, 080816732, 080816733, 080816734, 080816735, 080816736, 080816737, 080816738, 080816739, 080816740, 080816741, 080816742, 080816767, 080816768, 080816769, 080816770, 080816771, 080816772, 080816773, 080816809, 080816810, 080816811, 080816812, 080816813, 080816814, 080816815, 080816816, 080816817, 080816818, 080816819, 080816820, 080816821, 080816822, 080816823, 080816824, 080816825, 080816826, 080816827, 080816828, 080816829, 080816830, 080816831, 080816832, 080816833, 080816834, 080816835, 080816836, 080816837, 080816838, 080816839, 080816840, 080915161, 080915162, 080915163, 080915164, 080915165, 080915166, 080915167, 080915168, 080915169, 080915170, 080915171, 080915172, 080915173, 080915174, 080915175, 080915176, 080915177, 080915178, 080915179, 080915180, 080915181, 080915182 and 080915183. |
Recalling Firm/ Manufacturer |
Stryker Medical Div. of Stryker Corporation 3800 E. Centre Ave. Portage MI 49002
|
For Additional Information Contact | 269-324-6527 |
Manufacturer Reason for Recall | The backrest of the bed may not stay in an upright position (may drift down when left raised). |
FDA Determined Cause 2 | Device Design |
Action | Stryker Medical issued an "Urgent - Medical Device Correction" letter dated August 10, 2009 informing customers to locate the affected product listed and verify whether it is functioning properly. In the case the device is not properly functioning, remove from service and contact the firm as soon as possible at 1-800-STRYKER, option 3. Customers are also asked to return an enclosed return receipt post card and also forward the notification to users who have loaned or purchased the device. A representative from Stryker field service will be contacting you to perform device corrections shortly.
For further information, contact Stryker Medical at 1-269-324-6527. |
Quantity in Commerce | 12,599 total products |
Distribution | Worldwide Distribution -- United States, Brazil, Canada, Chile, China, India, Korea, Latin America, Mexico, Singapore and Switzerland. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|