Date Initiated by Firm | August 11, 2009 |
Date Posted | November 10, 2009 |
Recall Status1 |
Terminated 3 on December 22, 2009 |
Recall Number | Z-0164-2010 |
Recall Event ID |
53453 |
510(K)Number | K060631 |
Product Classification |
System, tomography, computed, emission - Product Code KPS
|
Product | Siemens Biograph mCT-S 64, model number 10248672.
Medical Imaging Equipment. |
Code Information |
serial numbers: 1002 and 1003 |
Recalling Firm/ Manufacturer |
Siemens Medical Solutions USA Inc. 810 Innovation Dr Knoxville TN 37932-2562
|
For Additional Information Contact | Maria Ebio 865-218-2532 |
Manufacturer Reason for Recall | In the unlikely event that the device is modified to operate in the tilted position, the telescoping support mechanism will not engage properly. |
FDA Determined Cause 2 | Device Design |
Action | Siemens Medical Solutions USA, Inc. issued an "Urgent Field Correction - Recall" notice to consignees by letter dated September 3, 2009 informing them of the problem and that a Siemens service representative will be in contact shortly to service the affected product.
For further information, contact Siemens at 1-800-888-7436. |
Quantity in Commerce | 2 units |
Distribution | Worldwide Distribution -- United States (IL and OH), Germany, France and Hong Kong. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = KPS
|