Class 2 Device Recall JOURNEY " UNI TIBIAL BASEPLATE, RIGHT MEDIAL/LEFT LATERAL, |
|
Date Initiated by Firm |
January 23, 2009 |
Date Posted |
March 03, 2010 |
Recall Status1 |
Terminated 3 on March 25, 2011 |
Recall Number |
Z-0934-2010 |
Recall Event ID |
54269 |
510(K)Number |
K052265
|
Product Classification |
Knee implant - Product Code HSX
|
Product |
JOURNEY " UNI TIBIAL BASEPLATE, RIGHT MEDIAL/LEFT LATERAL, A/P 55 MM, M/L 32 MM, REF 71422236, FOR CEMENTED USE ONLY, QTY: (1), STERILE R, SMITH & NEPHEW, INC. MEMPHIS, TN 38116 |
Code Information |
All Lot Numbers |
Recalling Firm/ Manufacturer |
Smith & Nephew Inc 1450 E. Brooks Rd Memphis TN 38116
|
For Additional Information Contact |
David Archer 901-399-5427
|
Manufacturer Reason for Recall |
The firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
|
FDA Determined Cause 2 |
Device Design |
Action |
All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 12/23/2009. All affected Smith & Nephew International Distributors were notified via e-mail and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. For questions, please contact the firm, Leah Easley at 901-399-6137. |
Quantity in Commerce |
838 units |
Distribution |
US and Argentina, Switzerland, Australia, Canada, Belgium, Italy, Spain, Thailand, Austria; PROC, Finland, Germany, Malaysia, Portugal, Netherlands, Singapore, and England |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HSX and Original Applicant = SMITH & NEPHEW, INC.
|
|
|
|